FTPL Fluke, FTPL Datasheet

Test Probes FUSED TEST PROBES, LEADS , PAIR RED/BLK

FTPL

Manufacturer Part Number
FTPL
Description
Test Probes FUSED TEST PROBES, LEADS , PAIR RED/BLK
Manufacturer
Fluke
Datasheet

Specifications of FTPL

Color
Red, Black
Description/function
Have built-in replacement fuses for additional measurement protection
Equipment Type
Fused Test Probes with Leads
Features
2 mm threaded probe tips, removable 4 mm lantern style spring contacts
Voltage Rating
1000 V
Current, Operating
A 10
Insulation Material
GS38 Insulated IC Caps
Standards
Meets ESA Guidelines, and CAT III 1000 V Rated
Temperature, Operating, Range
–20°C to +50°C (–2°F to +120°F)
Voltage, Operating
1000 V
Operating Temperature
–20°C to 50°C
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
3460786
Features
Fluke FTP SureGrip
additional measurement protection. If the probe fuse does blow, these
fused test probes will still allow the meter to indicate live voltage, * help-
ing to prevent a possible misreading of no voltage present. These modu-
lar probes can mount onto any standard modular test leads. Rated to
CAT III 1000 V, these probes also have sheathed tips for additional safety.
Specifications
*Voltage readings with blown fuse are approximate and vary with the meter impedance.
Before and after each measurement, verify presence of blown fuse by checking continuity of
the fused test probe. See explanation below under “DMM Input Impedance.” Current mode is
non-functional with blown fuse.
DMM input impedance
When the fuse is blown, there is a 1 Meg parallel resistor that will allow the DMM
to take a measurement. However, the input impedance of the DMM will have a
significant effect on the reading error. As examples, see the following table:
Test Probes and FTPL Fused
Actual Voltage DMM Input Impedance
480 V @ 60 Hz
480 V @ 60 Hz
480 V @ 60 Hz
Built-in fuses for additional measurement protection
Probes allow indication of live voltage even with blown fuse*
2 mm threaded probe tips include removable 4 mm lantern-style
spring contacts and removable GS38 insulated IC caps for probing
closely spaced leads
CAT III 1000 V, CAT IV 600 V, 10 A rating
FTPL Fused Test Probe and Lead Set includes Fluke TL224 test leads
Probes allow indication of live voltage even with blown fuse for added
security*
Built-in 1000 V, 11 A fuses with 20 kA interrupt rating
2 mm threaded probe tips include removable 4 mm lantern-style
spring contacts and removable GS38 insulated IC caps for probing
closely spaced leads
Operating temperature -20 °C to 50 °C (-5 °F to 120 °F)
CAT III 1000 V, CAT IV 600 V, 10 A rating, Pollution Degree 2
Recommended for use with Fluke modular TL224 SureGrip™ silicone
test leads and compatible with most modular test leads
Replace fuse by unscrewing probe handle at finger guard. Use only
Fluke fuse part number 803293.
Current mode is non-functional with blown fuse
FTPL Fused Test Probe and Lead Set includes Fluke TL224 test leads
Fluke SureGrip
10 M ohms || 100 pf
1 M ohms || 100 pf
3 k ohms || 100 pf
Test Probes with Leads
fused test probes have build-in replaceable fuses for
DMM reading (V) if
FTP Fuse Blown
226.3
417.9
1.3
Error
52.8
99.7
12.9
(%)
FTP Fused
Technical Data
Fluke. Keeping your world
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2009 Fluke Corporation. Specifications subject to
change without notice. Printed in U.S.A.
4/2009 3471734 D-EN-N Rev A
Modification of this document is not permitted
without written permission from Fluke Corporation.
FTPL Fused Test Probes with Leads