EA-IF-R2 EA ELEKTRO-AUTOMATIK, EA-IF-R2 Datasheet - Page 58

no-image

EA-IF-R2

Manufacturer Part Number
EA-IF-R2
Description
INTERFACE RS232 (EA-PSI/BCI 800 R)
Manufacturer
EA ELEKTRO-AUTOMATIK
Datasheet

Specifications of EA-IF-R2

Svhc
No SVHC (18-Jun-2010)
Accessory Type
Interface Card
Applications
Engineering Laboratory And Complex Industrial Application
Approval Bodies
CE / EN
Rohs Compliant
Yes
For Use With
EA Elektro-Automatik PSU
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
About the interface cards
© 009, Elektro-Automatik GmbH & Co. KG
4.1 RS232 cards IF-R1 and IF-R2
The RS3 interface card links the power supply with a
controlling unit (PC) via its serial port, also called COM port.
The settings of this serial connection have to be configured
on both ends to the same values. At the power supply this is
done in the setup menu. A 1:1 cable has to be used.
The card type 1 (IF-R1) features an additional serial interface
which is used to link multiple power supplies in order to build
the System Link Mode. More information about it in section
„6. The System Link Mode (only PSI9000)“.
4.1.1 Configuring the RS232 card
The interface card is configured in the setup menu.
It is absolutely necessary to choose and set a unique device
address, also called
linked unit. Only then a unit can be identified and controlled
correctly. This address is used to access a device.
Activate the menu with
Here you set the desired device node and you also get an
overview which cards are currently installed. By selecting
a card with
you enter the configuration menu for that particular card.
Each card has to be configured individually. You can now
setup the parameters:
The selected baud rate has to be determined in dependency
of the used cable length. At 15m a maximum of 9.6 kBd is
strongly recommended. 1kBd = 1000Bd.
4.2 USB cards IF-U1 and IF-U2
The USB interface works similiar to the RS3 card, but
it is more comfortable when connecting multiple units to a
PC via an USB hub. You can connect and control up to any
number (limited by the USB specs) of units with only one
PC and one USB port.
The card type 1 (IF-U1) features an additional serial interface
which is used to link multiple power supplies in order to build
the System Link Mode.
Never connect any of these ports to an Ethernet hub
or switch or Ethernet port of a PC!
= {1..30}
= {9.6 kBd, 19. kBd, 38.4 kBd, 57.6 kBd}
Slot A : { IF-… }
Slot B : { IF-… }
Slot {A|B}: IF-R1 +
M
device node
Baud rate
+
Communication +
Choose one of up to 30 device nodes
depends on the equipped card
depends on the equipped card
„device
node“, for every connected or
Default:
Default:
57.6 kBd
1
More information about it in section „4.5 System Link
Mode“.
4.2.1 Configuring the USB card
The interface card is configured in the setup menu.
It is absolutely necessary to choose and set a unique device
address, also called
linked unit. Only then a unit can be identified and controlled
correctly. The address is used to access a unit.
Activate the menu with
Here you set the desired device node and you also get an
overview which cards are currently installed. A further con-
figuration of the USB card is not required.
4.3 CAN cards IF-C1 and IF-U2
CAN Standard: V.0 part A
Cable length: depending on the baud rate
Specialty: Gateway to RS3 card or USB card
The communication over the CAN bus is specifically desi-
gned to suit the needs of test applications and systems, like
for instance in the automotive industry. A subsequent imple-
mentation into existing systems and the modification of a
related software application is possible and unproblematic.
The networking of CAN devices provides the advantage of
a faster communication and a fail-safe bus topology. The
driver chip on the CAN card can support up to 110
nodes
units). The LabView VIs resp. the communication protocol
can handle up to 30 units per address segment (RID). Thus
it is theoretically possible to set up a bus system of up to 110
units, which will operate with at least 4 address segments.
The address segments are relocatable, so that the one or
multiple devices can be implemented into an existing CAN
bus without the need to reconfigure the whole system.
Never connect any of these ports to an Ethernet hub
or switch or Ethernet port of a PC!
= {1..30}
Slot A : { IF-… }
Slot B : { IF-… }
M
device node
(the term device node is used for addresses of CAN
+
Communication +
Choose one of up to 30 device nodes
depending on the equipped card
depending on the equipped card
„device
node“, for every connected or
Default:
1
device
EN
58

Related parts for EA-IF-R2