EA-IF-C2 EA ELEKTRO-AUTOMATIK, EA-IF-C2 Datasheet - Page 69

no-image

EA-IF-C2

Manufacturer Part Number
EA-IF-C2
Description
INTERFACE CAN (EA-PSI/BCI 800 R)
Manufacturer
EA ELEKTRO-AUTOMATIK
Datasheet

Specifications of EA-IF-C2

Accessory Type
Interface Card
Svhc
No SVHC (18-Jun-2010)
Applications
Engineering Laboratory And Complex Industrial Application
Approval Bodies
CE / EN
Rohs Compliant
Yes
For Use With
EA Elektro-Automatik PSU
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
About the interface cards
© 009, Elektro-Automatik GmbH & Co. KG
Commands to control the output/input
Activates/deactivates the power output/input. A logical assignment for OUTP resp. INP is used. It means, that using INP
with a power supply, which has an OUTPut, will be ignored. An electronic load will react the same way, but vice versa.
OUTPut[:STATe]?
OUTPut[:STATe]
INPut[:STATe]?
INPut[:STATe]
Examples:
OUTPON
INP1
Measurement commands
Used to read the latest measured results. You are required to pay attention to the averaging settings (see „Averaging“) when
using the measurement commands.
MEASure
Examples:
MEAS:CURR?
MEAS:ARR?
Set value commands
Set values can also be read back by attaching a question mark to the command. For the meaning of A, B and A/B mode at
the electronic loads please also read the user manual of those devices!
I. Voltage set value/ Overvoltage threshold
(Specification according to „1999 SCPI Command reference“:19 Source Subsystem)
For electronic loads applies:
• Specific commands only for electronic loads are supported since firmware 3.01 and up.
• the HIGH set value always has to be greater than the LOW set value, else an error is generated. The commands HIGH
and LOW are only valid for Level A/B operation and will generate an error in different level control modes.
• Query and setting of set values are always dedicated to the currently selected „Level Control“ and the preselected „Mode“.
That is, if Level A is active, the set value sent with VOLT is set for Level A etc., but only if mode CV is preselected. Else the
set value is ignored and an error is generated. The control mode has to be selected before the device is set into remote
operation mode. The other set values, not belonging to the preselected mode, can not be set anymore and have to be
preset. Thus, for continuous remote control of the device, we recommend to set the option „Keep set values“ in the setup
menu to „no“, hence the set values are reset to default values when switching „Mode“.
[:SCALar]
:VOLTage[:DC]?
:CURRent[:DC]?
:POWer[:DC]?
:[ARRay]?
-><B0>
<B0>
->B0
<B0>
-><NRf>Unit
-><NRf>Unit
-><NRf>Unit
-><NRf>Unit, <NRf>Unit … Query: Actual voltage value, Actual current value, Actual power value..
Measures the actual current resp. returns its average value.
Returns a device depending number of actual values. For PSI/EL: U, I, P
Switches the power output on, but does not reset alarms or warnings and also does not
quit them. It means, if an alarm is persistent the command can‘t be executed.
Ditto, but for the input (of an electronic load)
Queries the state of the power supply output
Switches the power supply output on/off
Queries the state of the electronic load input
Switches the electronic load input on/off
Query: Actual voltage value
Query: Actual current value
Query: Actual power value
EN
69

Related parts for EA-IF-C2