34905A AGILENT TECHNOLOGIES, 34905A Datasheet

Computers, Data Switches

34905A

Manufacturer Part Number
34905A
Description
Computers, Data Switches
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 34905A

Frequency Range
45Hz To 66Hz
Bandwidth
2GHz
Isolation
-35dB
Insertion Loss
-3dB
Lead Free Status / RoHS Status
na
data acquisition systems
at a fraction of the cost
Agilent performance
of other standalone
Agilent 34970A
Data Acquisition/
Switch Unit Family
Product Overview
34970A
34972A
• 3-slot mainframe with built-in 6½ digit DMM and
• Measures and converts 11 different input signals:
• Gigabit LAN, USB, GPIB or RS-232 IO options for easy
• Graphical Web interface for point and click monitor and
• USB flash drive support to copy/log data in standalone
• Includes BenchLink Data Logger Software to configure
8 optional switch and control plug-in modules
temperature with thermocouples, RTDs and thermistors;
dc/ac volts; 2- and 4-wire resistance; frequency and
period; dc/ac current
connectivity to your PC
control (34972A)
applications (34972A)
and control tests, display results and collect data for
further analysis

Related parts for 34905A

34905A Summary of contents

Page 1

... Product Overview 34970A 34972A • 3-slot mainframe with built-in 6½ digit DMM and 8 optional switch and control plug-in modules • Measures and converts 11 different input signals: temperature with thermocouples, RTDs and thermistors; dc/ac volts; 2- and 4-wire resistance; frequency and period; dc/ac current • ...

Page 2

... Spec interpretation guide 16 Accuracy specifications 18 System measurement rates 19 System specifications 19 Agilent BenchLink Data Logger Software 20 Modules specifications 21 Multiplexer selection guide (34901A, 34902A, and 34908A) 23 Actuator module (34903A) 23 Matrix module (34904A Multiplexer modules (34905A, 34906A) 25 Multifunction module (34907A) 26 Rack mounting and dimensions 27 Ordering information ...

Page 3

... Features Price and performance beyond compare Go ahead and compare the Agilent Technologies 34970A and 34972A Data Acquisition/Switch Units with other data acquisition systems currently available. You’ll find it hard to come up with a system that offers the powerful measure- ment performance, flexibility, connectivity options and ease of use of the 34970A/34972A— ...

Page 4

... Specify per channel measurement configuration • Define and execute switch scans • Open, close, or monitor switch positions • Monitor measurement readings • View and save data • Send SCPI commands and view IO command log • ...

Page 5

... Buy only what you need—and add more modules later as your application grows. 34970A and 34972A are compatible The 34972A LXI Data Acquisition/Switch Unit is an LXI version of the 34970A. It replaces the GPIB and RS-232 interfaces with modern connectivity for a direct connection to your PC or Laptop. The 8 plug-in modules can be used with either unit— ...

Page 6

... HI/LO alarm limits on each input channel, plus 4 TTL alarm outputs • Offers matrix crosspoints or 120 single-ended channels • 8 switch and control plug-in modules to choose from • Agilent BenchLink Data Logger 3 Agilent VEE, IVI-C, IVI-COM and National Instruments LabVIEW drivers included ...

Page 7

... On the other hand, you had the flexibility and higher performance of a modular data acquisition system. The Agilent 34970A/34972A Data Acquisition/Switch Unit gives you the best of both worlds: a simple user interface with low per-channel cost, modular flexibility, standard connectivity and impressive measurement performance. Whether you’ ...

Page 8

An easy-to-use data logger for monitoring and characterization applications Data loggers are used to monitor multiple signals (tempera- ture, voltage, etc.) over extended periods of time to identify irregularities. Example applications include environmental chamber monitoring, component inspection, benchtop test- ing, ...

Page 9

And the 34970A/34972A measures and converts 11 different input signals: • temperature with thermocouples, RTDs, and thermistors • dc and ac volts • 2- and 4-wire resistance • frequency and period • dc and ac current What’s more, each channel ...

Page 10

... The DMM is installed inside the chassis rather than in one of the slots, leaving all three mainframe slots free for switch and control modules. You can choose from eight different modules (see page 13) to get the precise function- ality you need now—while giving you flexibility for future expansion ...

Page 11

... Up to 120 single-ended measurements or 96 matrix crosspoints in a 3½” high, half-rack instrument • Eight switch and control modules include low- frequency and RF multiplexers, matrix and actuation switches, digital input and output, analog output, and event recording • ...

Page 12

... DMM. What you end up with is the lowest cost switch unit on the market. It’s an ideal solution for routing test signals to and from your DUT and assorted instruments, including external DMMs, scopes, counters, and power supplies. ...

Page 13

... Multiplexer (4-wire selectable) 34903A SPDT/form Actuator/GP Switch 34904A Matrix 2-wire armature 34905A Common low Dual Mux 50Ω (unterminated) 34906A Common low Dual Mux 75Ω (unterminated) 34907A Two 8-bit digital I/O ports Multifunction Module ...

Page 14

... The 34970A/34972A uses our proprietary relay mainte- nance system to help you to predict relay end-of-life and avoid costly production line downtime. It automatically counts every individual switch closure and stores it in nonvolatile memory on each module. You can query the total number of cycles on any individual channel so you can ...

Page 15

... etc.)—not the full scale value (1 etc.). • DMM measurement accuracies include all switching errors. Switching errors are also listed separately in the module specifications section. Temperature measurement accuracies include ITS-90 conversion errors. The thermo- couple accuracies include the reference junction error as well. • ...

Page 16

... Includes measurement error, switching error, and transducer conversion error [3] Range DC voltage 100.0000 mV 1.000000 V 10.00000 V 100.0000 V 300.000 V [4] True RMS AC voltage All ranges from 100.0000 mV to 100.0000 V 300.0000 V [6] Resistance 100.0000 Ω 1.000000 kΩ ...

Page 17

Measurement characteristics DC voltage Measurement Method Continuously Integrating Multi-slope III A-D converter A/D linearity 0.0002% of reading + 0.0001 % of range Input resistance 100 mV ranges Selectable 10 MΩ or > 10,000 MΩ 100 ...

Page 18

Single channel reading rates to I/O or internal memory Single channel ASCII readings Single channel while changing scale (eg MEAS dcV 10/MEAS dcV 1) Single channel while changing function (eg MEAS dcV/MEAS Ohms) Scanning measurement rates to I/O or internal ...

Page 19

... Add, delete, size, and configure real time Strip chart with markers and alarm indication, bar and scatter charts, Histogram with statistics, Bar meter, and Data table Sliders, switches, buttons, and LED lights Start/Stop scanning on alarm condition Control 34903A relay state or 34907A digital output on alarm ...

Page 20

... On-module screw terminals accept wire sizes from 16 gage to 22 gage. Twenty-gage wire is recommended for high channel count applications. The 34905A and 34906A RF Multiplexers use SMB connectors. A standard set of (10) BNC-to-SMB adapter cables is provided with each RF module for convenient BNC connections. ...

Page 21

... General Purpose Multiplexer • 60 ch/s scanning • Two- and four-wire scanning • Built-in thermocouple reference junction • 300 V switching The Agilent 34901A is the most versatile multiplexer for general purpose scanning. It combines dense, multifunction switching with 60-channel/second scan rates to address a broad spectrum of data acquisition applications ...

Page 22

... Use this module for high-throughput automated test applications as well as high-speed data logging and monitoring tasks. Sixteen two-wire inputs switch up to 300 V. Two- and four-wire channels may be mixed on the same module. User provided shunt resistors are required for current measurements. ...

Page 23

... Combine it with matrix and multiplexer modules to build custom switch systems. Its 300 contacts can handle enough for many power line switching applica- tions. 34904A 4x8 Two-wire Matrix Switch • ...

Page 24

... Dual 4-Channel RF Multiplexers • 2 GHz bandwidth • BNC to SMB adapter cables included The Agilent 34905A and 34906A RF multiplexers offer broadband switching capabilities for high frequency and pulsed signals. Use them to route test signals between your device under test and your signal generator, oscillo- scope, spectrum analyzer, or other instrumentation ...

Page 25

... Use the digital outputs with an external power supply to control microwave switches and attenuators, solenoids, power relays, indicators, and more. Use the digital inputs to sense limit switch and digital bus status. There are no complex handshake modes; reads and writes are initiated either from the front panel or the bus. ...

Page 26

Rack mounting and dimensions To rack mount a single instrument, order adapter kit 5063-9240 (Option 1CM). To rack mount two instruments side-by-side, order lock-link kit 5061-9694 and fl ange kit 5063-9212. To install one or two instruments in a sliding ...

Page 27

... Italy: Italian manual Modules 34901A 20-Channel armature multiplexer 34902A 16-Channel reed multiplexer 34903A 20-Channel actuator/general purpose switch 34904A Two-wire matrix switch 34905A Dual 4-Channel RF multiplexer, 50 Ohms 34906A Dual 4-Channel RF multiplexer, 75 Ohms 34907A Multifunction module 34908A 40-Channel single-ended multiplexer Accessories 34830A Benchlink Data Logger Pro Software optional ...

Page 28

... Sweden 0200- Switzerland 0800 United Kingdom 44 (0) 118 9276201 Other European Countries: www.agilent.com/find/contactus Revised: October 1, 2009 Product specifications and descriptions in this document subject to change without notice. © Agilent Technologies, Inc. 1998, 2000, 2004, 2005, 2010 Printed in USA, March 8, 2010 5965-5290EN ...