NFS110-7902PJ ARTESYN, NFS110-7902PJ Datasheet

no-image

NFS110-7902PJ

Manufacturer Part Number
NFS110-7902PJ
Description
AC/DC Power Supply Quad-OUT 5.1V/24V/12V/-12V 8A/3.5A/4.5A/0.5A 110W 16-Pin
Manufacturer
ARTESYN
Datasheet

Specifications of NFS110-7902PJ

Input Voltage
90 to 253 V
Output Voltage
5.1|24|12|-12 V
Maximum Output Current
8|3.5|4.5|0.5 A
Number Of Outputs
4
Maximum Output Power
110 W
Efficiency
70(Typ) %
Input Frequency
47 to 440 Hz
Operating Temperature
0 to 70 °C
NFS110 Medical
Series
Single and
quad output
Special Features
• 7.0 x 4.25 x 1.8 inch package
• Medical, dental and
• Overvoltage and short circuit
• 110 W with 20 CFM
• UL, VDE and CSA safety
• EN60601-1 and UL2601
• Available RoHS compliant
• 2 year warranty
Safety
VDE0805/EN60601-1/
IEC601/IEC1010
File No. 10401-3336-1049
Licence No. 2874
UL2601 File No. E147937
CSA C22.2 No. 125
File No. LR41062C
Total Power:
Input Voltage: 90 - 253 Vac
# of Outputs:
laboratory applications
protection
approvals
medical approvals
80 - 110 W
127 - 357 Vdc
Single, quad
Electrical Specifications
Output
Input
Input voltage range
Input frequency range
Input surge current
Safety ground leakage current 132 Vac
All specifications are typical at nominal input, full load at
25
Voltage adjustability
Line regulation
Overshoot/undershoot
Temperature coefficient
Overvoltage protection
Output power limit
Short circuit protection
°
C unless otherwise stated
110 Vac. 50 Hz
230 Vac. 50 Hz
264 Vac
+5.1 V o/p on multi’s
5.1 V single output
12 V single output
15 V single output
24 V single output
LL to HL, FL
All outputs on all units
At turn-on no lead
All outputs
Multi o/p 5.1 V only
5.1 V single
12 V single
15 V single
24 V single
Primary power
limited
90-253 Vac
127-357 Vdc
47-440 Hz
17 A
35 A
50 μA
100 μA
±3.0%
±3.0%
12-14 V
15-18 V
24-30 V
±0.1% max.
0%
±0.02%/°C
6.25 V ±0.75 V
6.25 V ±0.75 V
15.75 V ±1.0 V
22 V ±1.5 V
33 V ±2.5 V
Pin max. 160 W
Pout min. 110 W
Burst mode operation
Embedded Power for
Business-Critical Continuity
NFS110 Series
Rev. 06.10.08
1 of 5

Related parts for NFS110-7902PJ

NFS110-7902PJ Summary of contents

Page 1

... Vac Safety ground leakage current 132 Vac 264 Vac All specifications are typical at nominal input, full load at ° unless otherwise stated Embedded Power for Business-Critical Continuity Rev. 06.10.08 NFS110 Series ±3.0% ±3.0% 12-14 V 15-18 V 24-30 V ±0.1% max. 0% ±0.02%/°C 6.25 V ±0.75 V 6.25 V ± ...

Page 2

... CSA C22.2 No. 125 550 g (19.4 oz) 600 g (21.2 oz) 125,000 hours min. 0° +70 °C -40 °C to +85 °C Derate 2 W/°C 110 W Derate 2.75 W/° 95% RH 10,000 feet max. 40,000 feet max. 2.4 G rms peak Embedded Power for Business-Critical Continuity Rev. 06.10.08 NFS110 Series ...

Page 3

... Total regulation is defined at the static output regulation at 25 °C, including initial tolerance, line voltage within stated limits and output voltages adjusted to their factory settings. Also for NFS110-7902PJ, for 24 V output stated ± regulation ² ...

Page 4

... F A standard L-bracket and cover is available for mounting which contains all screws, connectors and necessary mounting hardware. The kit is available, order part number “NFS110CJ” . OPTIONAL POWER FAIL DETECT TIMING DIAGRAM AC INPUT 4.75V 5V OUTPUT ...

Page 5

... Embedded Power for Business-Critical Continuity Rev. 06.10.08 Americas NFS110 Series 5810 Van Allen Way Carlsbad, CA 92008 USA Telephone: +1 (760) 930 4600 Facsimile: +1 (760) 930 0698 Europe (UK) Waterfront Business Park Merry Hill, Dudley West Midlands, DY5 1LX United Kingdom Telephone: +44 (0) 1384 842 211 ...