GLM75-24 SL Power ( Ault / Condor ), GLM75-24 Datasheet

no-image

GLM75-24

Manufacturer Part Number
GLM75-24
Description
Power Supply; 75 W; Single; 24 V; 3.1 A (Max.); 90 to 264 VAC; 2.3 A (Max.);
Manufacturer
SL Power ( Ault / Condor )
Type
Medicalr
Datasheet

Specifications of GLM75-24

Brand/series
GLM75 Series
Configuration
Open-Frame
Current, Input
2.3 A (Max.)
Current, Inrush
Inrush at 240 Vac, averaged over the first ac half-cycle under cold start conditions will not exceed 37 A.
Current, Leakage
70 μA @ 264 VAC
Current, Output
3.1 A (Max.)
Dimensions
5.75 in. L x 3.40 in. D x 1.563 in. H
Mounting Type
Panel
Noise, Output
0.5% rms, 1% pk-pk, 20 MHz bandwidth,
Number Of Outputs
1
Operation
Switching
Output
24 VDC @ 3.1 A
Output Type
Single
Overvoltage Protection
Standard on all models.
Power Rating
75 W
Power, Output
75 W
Primary Type
AC-DC
Regulation, Total
1
Standards
UL Listed, CSA Certified, CE Marked
Voltage, Input
90 to 264 VAC
Voltage, Output
24 VDC
40
SPECIFICATIONS:
Ac Input
90-264 Vac, 47-63 Hz single phase.
Input Current
Maximum input current at 120 Vac, 60 Hz with full rated
output load not to exceed 2.3 A.
Output Power
Normal continuous output power is 75 W for unrestricted
natural convection cooling; 110 W with 26 cfm airflow.
Output Regulation
Regulation measured by changing from 5% to 50% load
or 50% load to full load in either direction.
Overload Protection
Factory set to begin power limiting at approximately 120
W (GLC 75-5 is set at approximately 100 W). Fully
protected against short circuit and output overload. Short
circuit protection is cycling type power limit.
Output Noise
0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential
mode. Measured with noise probe directly across output
terminals of the power supply.
Transient Response
Main Output: 500 µs typical response time for return to
within 0.5% of final value for a 50% load step change, i/
shutdown overshoot less than 3%.
Overvoltage Protection
Standard on all models.
Voltage Adjust
Factory set on standard unit; however, potentiometer
adjusts voltage ±5% minimum. Consult factory for
application assistance.
Efficiency
72-85% depending on model.
Input Protection
Internal ac fuse provided. Designed to blow only if a
catastrophic failure occurs in the unit—fuse does not blow
on overload or short circuit.
GLC75
75 Watt Single Output Global Performance Switchers
t<0.2 A/µs. Maximum voltage deviation is 3.5%. Startup/
Commercial/GLM75 Medical
Inrush Current
Inrush limited by internal thermistors. Inrush at 240 Vac,
averaged over the first ac half-cycle under cold start
conditions will not exceed 37 A.
Temperature Coefficient
0.03%/°C typical on all outputs.
Power Fail
A standard TTL or CMOS compatible output goes low
(< 0.5 V) 5ms before output voltage drops more than 4%
below nominal voltage upon loss of ac power. Signal is
factory set to trip on 84 to 94 Vac brown-out depending
upon incoming line impedance and distortion. Other
settings are available through adjustment of built-in
potentiometer (consult factory for assistance). Output will
stay low for 20 ms minimum.
EMI/EMC Compliance
All models include built-in EMI filtering to meet the
following emissions requirements:
EMI SPECIFICATIONS
Conducted Emissions GLC75 EN55022 Class B; FCC Class B
Conducted Emissions GLM75 EN55011 Class B; FCC Class B
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Line Frequency Harmonics
Commercial Safety
All GLC models are approved to UL1950, CSA22.2 No.
234 Level 3, IEC950 and EN60950. Consult factory for
approval status.
Medical Leakage Current
70 µA 264 V @ 50 Hz (normal conditions).
Medical Safety
All GLM models are approved to UL2601-1, CSA-C22.2
No. 601.1, IEC601-1 and EN60601-1. Consult factory for
approval status.
FEATURES:
• Cost-effective single-output power source
• 3.4" x 5.75" x 1.56" (meets 1U applications)
• Universal input 90-264 Vac
• 2-year warranty
• Complies with EN61000-3-2 Class A
• Also available in multiple output versions
• Conducted EMI exceeds FCC Class B and CISPR 22
• Commercial Approved to UL1950, CSA22.2 No. 234
• Medical Approved to UL2601-1, IEC601-1 and
Class B (Commercial models) and CISPR 11 Class B
(Medical models)
and IEC950, EN60950
CSA22.2 No. 601
marked to LVD
EN61000-4-5, 1 kV diff., 2 kV com.
EN61000-3-2 Class A
COMPLIANCE LEVEL
EN61000-4-2, 6 kV contact, 8 kV air
EN61000-4-4, 2 kV, 5 kHz
EN61000-4-3, 3 V/meter

Related parts for GLM75-24

GLM75-24 Summary of contents

Page 1

... EMI/EMC Compliance All models include built-in EMI filtering to meet the following emissions requirements: EMI SPECIFICATIONS COMPLIANCE LEVEL Conducted Emissions GLC75 EN55022 Class B; FCC Class B Conducted Emissions GLM75 EN55011 Class B; FCC Class B Static Discharge EN61000-4- contact air RF Field Susceptibility EN61000-4-3, 3 V/meter Fast Transients/Bursts ...

Page 2

... GLC75 Commercial/GLM75 Medical Rating with unrestricted convection cooling. Total power not to exceed Rating with 26 cfm forced-air cooling. Total power not to exceed 110 W. GLC75/GLM75 MECHANICAL SPECIFICATIONS Watt Single Output Units should be allowed to warm up/operate under non-condensing conditions before application of power Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz, 3 ...