Power Supply, AC/DC, Medical, 110W, 24V, 4.6A, 100-240 VAC, Switching, Open Frame
Manufacturer
SL Power ( Ault / Condor )
Type
Medicalr
Datasheet

Specifications of MINT1110A2408K01

Brand/series
MINT1110A Series
Configuration
Open-Frame
Mounting Type
Panel
Number Of Outputs
1
Power, Output
110 W
Primary Type
AC-DC
Voltage, Input
110-240 VAC
Voltage, Output
24 VDC
Universal Input
90 - 264 VAC
Lead Free Status / Rohs Status
RoHS Compliant part
MINT1110A
110W Single Output Series
Medical – Universal Input Power Supply
110W Convection Cooled AC-DC Power Supply
Universal Input 90-264 VAC
Less than 1.27” Height
Rugged EMC Compliant Design
3” x 5” Package
Approved to EN/IEC/UL60601-1 for Medical
Chassis & Cover Optional
AC Input
100-240VAC +/- 10%, 47-63 Hz single phase.
Input Current
115VAC: 1.4A.
Inrush Current
The inrush at 264VAC, at a cold start, will not exceed 40A.
Input Fuses
F1, F2: 2.5A, 250VAC fuses provided on all models.
Power Factor
Complies with EN61000-3-2.
Efficiency
87% typical.
Output Power
90 - 110W continuous.
Ripple and Noise
1% pk-pk max., 20MHz BW, differential mode. Measured with noise
probe directly across output terminals.
Output Voltage
See chart.
Voltage Adjustability
Adjustable potentiometer, capable of ±5% change from nominal.
Minimum Load
Not required.
Total Regulation
See chart.
Transient Response
500µs max., 50% load step, typical.
Specifications
SL Power Electronics Corp
______________________________________________________________________________________________________________________________
Compliant (LVD, RoHS)
6050 King Drive Ventura, CA 93003
Phone 805.486.4565 • Fax 858.712.2040
All Specifications are typical at nominal input, full load at 25
Turn On Time
Less than 2 seconds under all rated load conditions.
Hold-up Time
16 ms at 110W, nominal input (120VAC).
Overvoltage Protection
See chart for levels, latching type
Overload Protection
Hiccup Mode
Short Circuit Protection
Hiccup Mode
Overtemperature Protection
Provided on all models.
Medical Safety Standards
EN/IEC/UL 60601-1, CSA22.2, No. 601.1
Isolation
Input-Output: 4000VAC, Input-Ground: 1500VAC, Output-Ground:
500VDC
Earth Leakage Current
<130µA at 264Vac, 60Hz.
Operating Temperature
0° to 70°C (derate from full rated power above 40°C)
Output Power Derating
Derate Output Power by 2.5% per °C above 40°C, up to 70°C max.
Storage Temperature
-40 to +85°C.
Relative Humidity
5% to 95%, non-condensing
2 Year Warranty
Email: info@slpower.com
°
C unless otherwise stated.
www.slpower.com
1 of 3

Related parts for MINT1110A2408K01

MINT1110A2408K01 Summary of contents

Page 1

MINT1110A 110W Single Output Series Medical – Universal Input Power Supply 110W Convection Cooled AC-DC Power Supply • Universal Input 90-264 VAC • Less than 1.27” Height • Rugged EMC Compliant Design • 3” x 5” Package • Approved to ...

Page 2

... MINT1110A 110W Single Output Series Medical – Universal Input Power Supply Model Number Volts (V) MINT1110A1208K01 12V MINT1110A1508K01 15V MINT1110A1808K01 18V MINT1110A2408K01 24V Specifications (continued) Vibration 2 Operating: 0.003 g /Hz, 1.5 g overall, 3 axes, 10 min. / axis rms 2 Non-Operating: 0.026 g /Hz, 5.0 g overall, 3 axes, 1 hr. / axis ...

Page 3

MINT1110A 110W Single Output Series Medical – Universal Input Power Supply Outline Drawing INPUT CONNECTOR: J1 LANDWIN 3361P03E3TSN02 PIN 1) AC LINE PIN 2) AC NEUTRAL OUTPUT CONNECTOR: J2 LANDWIN P/N 3361P08E3T PIN 1) +Vout PIN 5) -Vout PIN 2) ...

Related keywords