GPM41-12 SL Power ( Ault / Condor ), GPM41-12 Datasheet

no-image

GPM41-12

Manufacturer Part Number
GPM41-12
Description
Power Supply; 41 W; Single Output; 12 V; 3.3 A; 85 to 264 VAC;
Manufacturer
SL Power ( Ault / Condor )
Type
Medicalr
Datasheet

Specifications of GPM41-12

Brand/series
GPM41 Series
Configuration
Open-Frame
Current, Input
1.3 A @ 120 VAC, 60 Hz
Current, Inrush
The inrush at 240 Vac, averaged over the first ac half-cycle under cold start conditions will not exceed 37 A.
Current, Leakage
60 uA (Commercial), 35 uA (Medical)
Current, Output
3.3 A
Dimensions
4 in. L x 2 in. W x 1.5 in. H
Hold Time
20 ms minimum from loss of ac input at full load, nominal line (115 Vac).
Mounting Type
Panel
Noise, Output
0.5% rms, 1% pk-pk, 20 MHz bandwidth,
Number Of Outputs
1
Operation
Switching
Output
12 VDC @ 3.3 A
Output Type
Single Output
Overvoltage Protection
14 ± 1.0 V
Power Rating
41 W
Power, Output
41 W
Primary Type
AC-DC
Standards
UL Listed, CSA Certified, CE Marked and IEC Approved
Voltage, Input
85 to 264 VAC
Voltage, Output
12 VDC
28
GPC41
41 Watt Global Performance Switchers
SPECIFICATIONS:
Ac Input
85-264 Vac, 47-63 Hz single phase.
Input Current
Maximum input current at 120 Vac, 60 Hz with full rated
output load: 1.3 A
Hold-Up Time
20 ms minimum from loss of ac input at full load, nominal
line (115 Vac).
Output Power
40 W continuous, 50 W peak. Peak ratings are for 60 s
maximum duration, 10% duty cycle. During peak load
condition, output regulation may exceed total regulation
limits.
Overload Protection
Fully protected against short circuit and output overload.
Short circuit protection is cycling type power limit.
Recovery after fault is automatic. Factory set to begin
power limiting at approximately 55 W. See output ratings
chart for additional notes or conditions.
Overvoltage Protection
See output ratings chart.
Efficiency
69-77% at full rated load, nominal input voltage, depend-
ing on model.
Turn-on Time
Less than 1 s at 120 Vac, 25°C (inversely proportional to
input voltage).
Input Protection
Requires external ac line fusing. A 2.5 A, 250 V rated fuse
with I
Inrush Current
Inrush is limited by internal thermistor. Inrush at 240 Vac
under cold start conditions will not exceed 34 A.
Temperature Coefficient
0.03%/°C typical on all outputs.
Environmental
Designed for 0 to 50°C operation at full rated output
power; derate output current and total output power by
2.5% per °C above 50°C. See Environmental and
Packaging Specifications on next page.
2
t > 20 is recommended.
Commercial/GPM41 Medical
FEATURES:
• Ultra small (2.00" x 4.00" x 1.50")
• Wide-range ac input 85-264 Vac
• 2-year warranty
• Conducted EMI exceeds FCC Class B and CISPR 22
• Commercial Approved to UL1950, IEC950, EN60950
• Medical Approved to UL2601-1, IEC601-1
Output Noise
0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential
mode. Measured with noise probe directly across output
terminals of the power supply.
Transient Response
Main output—500 µs typical response time for return to
within 0.5% of final value for a 50% load step change. i/
shutdown overshoot less than 3%.
EMI/EMC Compliance
All models include built-in EMI filtering to meet the
following emissions requirements:
EMI SPECIFICATIONS
Conducted Emissions GPC41 EN55022 Class B; FCC Class B
Conducted Emissions GPM41 EN55011 Class B; FCC Class B
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Commercial Leakage Current
60 µA single fault; 60 Hz input.
Commercial Safety
Approved to UL1950, CSA22.2 No. 234 Level 3, IEC950
and EN60950. UL file #E135803 commercial; CSA
#LR46516 all models. The output is intended for safety
earthed Signal Output and Intermediate Circuits only. All
dc outputs are SELV under normal and single fault
conditions.
Medical Leakage Current
35 µA , 254 Vac 10 Hz input.
Medical Safety
Approved to UL2601, CSA22.2 No. 601 Level 3, IEC601.
UL file E116994; CSA #LR46516. The output(s) are
intended for safety earthed Signal Output and Intermedi-
ate Circuits only. The output(s) are not acceptable for
patient connection without additional isolation. All dc
outputs are SELV under normal and single fault condi-
tions.
t<0.2 A/µs. Maximum voltage deviation is 3.5%. Startup/
Class B (Commercial models) and CISPR 11 Class B
(Medical models)
and CSA22.2-234 L3
and CSA22.2 No. 601
marked to LVD
EN61000-4-5, 1 kV diff., 2 kV com.
EN61000-4-2, 6 kV contact, 8 kV air
EN61000-4-4, 2 kV, 5 kH
COMPLIANCE LEVEL
EN61000-4-3, 3 V/meter

Related parts for GPM41-12

GPM41-12 Summary of contents

Page 1

... EMI/EMC Compliance All models include built-in EMI filtering to meet the following emissions requirements: EMI SPECIFICATIONS COMPLIANCE LEVEL Conducted Emissions GPC41 EN55022 Class B; FCC Class B Conducted Emissions GPM41 EN55011 Class B; FCC Class B Static Discharge EN61000-4- contact air RF Field Susceptibility EN61000-4-3, 3 V/meter Fast Transients/Bursts ...

Page 2

... GPC41 Commercial/GPM41 Medical GPC41/GPM41 MECHANICAL SPECIFICATIONS Watt Single Output Units should be allowed to warm up/operate under non-condensing conditions before application of power Random vibration—10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis non-operating. C. Shock testing—half-sinusoidal, 10 ± duration, ± direction, 3 orthogonal axes, ...