... Reading storage with internal memory Agilent E1412A 6.5-Digit High-Accuracy Multimeter C-Size Data Sheet Description The Agilent Technologies E1412A 6.5-Digit Multimeter is a C-size, 1-slot, message-based VXI module identical in electrical design to the E1312A, differing only in size. It delivers the widest func- tionality in Agilent’s DMM line. ...
Page 2
DC Specifi cations Specifi cations are for 1-hour warm- integration time of 100 PLCs. DC Summary: DC voltage: 300 V max. Voltage accuracy (dc): ± 0.0019% DC Accuracy ± reading + % of range): DC voltage: ...
Page 3
DC Specifi cations (continued) Specifi cations are for 1-hour warm- integration time of 100 PLCs. DC Voltage Characteristic s: Measurement method: A/D linearity: 2/4-wire Ω: Input resistance: 0 ranges: 100 V, 300 V ...
Page 4
DC Specifi cations (continued) Specifi cations are for 1-hour warm- integration time of 100 PLCs. DC:DC Ratio: Measurement method: Input HI-LO/Reference HI-LO Apply “Reference HI-LO” signal to Ohms 4-Wire Sense terminals. Input HI to Input LO: 100 mV ...
Page 5
AC Specifi cations AC Summary AC voltage: 300 V max. Voltage accuracy (ac): ± 0.07% AC Accuracy ± reading + % of range): True RMS AC Voltage 3 : Range 2 Frequency 100.0000 mV 3 Hz-5 Hz 100.0000 ...
Page 6
AC Specifi cations (continued) Additional AC Specifi cations Low Frequency Errors (% of reading) Frequency AC Filter Slow Medium 10 Hz- 0.74 20 Hz- 0.22 40 Hz-100 Hz 0 0.06 100 Hz-200 Hz 0 0.01 200 ...
Page 7
AC Specifi cations (continued) AC Operating Characteristics: Maximum reading rates 0.01 step additional error. Additional settling delay required when input dc level varies. Function Digits ACV and ACI 6.5 ACV and ACI 6.5 ACV and ACI 6.5 ACV ...
Page 8
Frequency and Period Specifi cations Frequency and Period Accuracy (% of reading): Specifi cations are for 1-hour warm-up at 6.5 digits. 2 Function Range Frequency Frequency, Period 100 mV to 300 V 3 Hz-5 Hz Frequency, Period 100 mV to ...
Page 9
Frequency and Period Specifi cations Operating Characteristics: Speeds are for 4.5 digits, Delay 0, and Fast ac fi lter. Function Frequency, Period Frequency, Period Frequency, Period Systems Speeds: Confi guration rates: Autorange time: Max. internal trigger rate: Max. external trigger ...
Page 10
... - -5 Cooling Slot: Watts/slot: 9.40 ∆ 0.05 Air Flow liter/s: 0.80 10 Ordering Information Description Product No. 6.5-Digit Multimeter, E1412A High Accuracy ANSI Z540 E1412A A6J Compliant Calibration 3 year return to E1412A W01 Agilent to 1 year OnSite warranty ...
Page 11
... Switzerland (French) 41 (21) 8113811(Opt 2) Switzerland (German) 0800 (Opt 1) United Kingdom 44 (0) 118 9276201 Other European Countries: www.agilent.com/find/contactus Revised: October 24, 2007 Product specifications and descriptions in this document subject to change without notice. © Agilent Technologies, Inc. 2008 Printed in USA, January 31, 2008 5965-5563E ...