IR205 EXTECH [Extech Instruments Corporation.], IR205 Datasheet

no-image

IR205

Manufacturer Part Number
IR205
Description
Pocket InfraRed Thermometer
Manufacturer
EXTECH [Extech Instruments Corporation.]
Datasheet
Pocket InfraRed Thermometer
Introduction
Congratulations on your purchase of Extech’s IR201 Pocket InfraRed
Thermometer. This device allows the user to take non-contact temperature
measurements simply by pointing the meter and pressing a button. Features
include: Laser pointer, Max/Min, Adjustable emissivity, Lock ON and HI
audible alarm. This meter is shipped fully tested and calibrated and, with
proper use, will provide years of reliable service.
IR Measurements
Taking Measurements
Laser Pointer
Set Feature
Display
Controls
1.
2.
3.
4.
5.
6.
7.
Model IR205
User's Manual
Point the meter toward the object being measured, then press and hold
the MEASURE button. The meter will turn ON and take readings (SCAN).
Read the measured temperature on the LCD. When the MEASURE
button is released, the last measurement is HELD in the display for
approximately 8 seconds before the meter shuts down.
While pressing the MEASURE button, Press the LASER button to turn
the laser pointer ON or OFF.
While in the MEAS mode;
Press the SET button to gain access to the following features in
sequence.
Release the MEAS button when the feature needed is selected
Press the Laser button to change the selection or value
Press the MEAS button to confirm the selection.
#1: Max/Min:
#2: °C/°F:
#3: Є ▲:
#4: Є ▼:
#5:
#6: Hi ▲:
#7: HI ▼:
#8:
IR Sensor
Laser Pointer
MEASURE Button
Battery Compartment (rear)
LASER button
SET button
Display
:
Displays only the maximum or minimum value
during a measurement scan.
Select measurement units
Increase emissivity value.
Decrease emissivity value
Lock On/Off (Overrides auto-power off)
Increase High Alarm value
Decrease High Alarm value
Audible Alarm On/Off:
Specifications
Measurement considerations:
Distance to spot diagram
Battery Replacement
A flashing display indicates that the AAA batteries which power the meter
have weakened. Open the battery compartment on the rear of the meter,
replace the batteries and close the compartment.
Warranty
EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in
parts and workmanship for one year from date of shipment (a six month limited warranty
applies on sensors and cables). If it should become necessary to return the instrument for
service during or beyond the warranty period, contact the Customer Service Department at
(781) 890-7440 for authorization. A Return Authorization (RA) number must be issued
before any product is returned to Extech. The sender is responsible for shipping charges,
freight, insurance and proper packaging to prevent damage in transit. This warranty does
not apply to defects resulting from action of the user such as misuse, improper wiring,
operation outside of specification, improper maintenance or repair, or unauthorized
modification. Extech specifically disclaims any implied warranties or merchantability or
fitness for a specific purpose and will not be liable for any direct, indirect, incidental or
consequential damages. Extech's total liability is limited to repair or replacement of the
product. The warranty set forth above is inclusive and no other warranty, whether written or
oral, is expressed or implied.
Display
Power off
Laser power
Operating Temperature
Operating Humidity
Power Supply
Weight
Ranges(Resolution)
Accuracy
Recommended measuring
distance
Emissivity settings
Distance to Target Ratio
Size
If the surface of an object to be measured is covered with frost, dust or
other materials, surface cleaning must be performed before accurate
measurements can be made.
If the surface of an object to be measured is highly reflective with a very
low emissivity value, apply masking tape or apply flat black paint to the
surface to raise to improve the accuracy of the measurement.
The object being measured should be larger than the spot size as
illustrated in the distance to spot diagram.
-31 to 518°F/ -35 to 270°C (1°F/1°C)
± (2.5% of reading + 3°F or 2°C)
(-4° to 518°F / -20° to 270°C):
± 10°F or 5°C
(-31° to -4°F / -35° to -20°C)
2” to 36” / 5cm to 91cm
0.1 to 1.00 adjustable
6:1
3 ½ -digit (2000 count) LCD display
Automatic shutoff after approx. 8 seconds
Laser power less than 1mW (red)
32 to 122°F (0 to 50°C)
Max. 80% RH
2 x AAA battery
2.7 oz. / 76g
3.7 x 2 x 1" (94 x 51 x 25 mm)
6”(15cm)
12”(30cm)
1”(2cm)
18”(48cm)
2”(5cm)
IR205 V1.0 3/06
3”(8cm)

Related parts for IR205

Related keywords