NIPXI-2530 NI [National Instruments Corporation], NIPXI-2530 Datasheet

no-image

NIPXI-2530

Manufacturer Part Number
NIPXI-2530
Description
61⁄2-Digit Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter
Manufacturer
NI [National Instruments Corporation]
Datasheet
6
Isolated Digitizer, and LCR Meter
NI PXI-4072
• Multifunction device
• 20 built-in measurements
• 10- to 23-bit flexible resolution
• Built-in self-calibration
• ±300 VDC/V
• 425 V
Calibration
• Gain, temperature, and offset
• 2-year external calibration cycle
Overview
The National Instruments PXI-4072 FlexDMM, based on the NI 4070
architecture, combines the functionality found in three common test
instruments – a digital multimeter, an LCR meter, and a digitizer. This
functionality provides engineers with the 20 most common test functions
in a compact 3U PXI module, including voltage, current, capacitance,
inductance, temperature, and resistance. Integrating these measurements
in a PXI module reduces test system size and cost, increases throughput,
and shortens test development time, making the NI PXI 4072 an
excellent fit for use in automated tests on both the production floor and
in an R&D environment.
to 5 S/s at 7 digits. As a digitizer, the FlexDMM can acquire both AC-
and DC-coupled voltage and current waveforms up to ±300 V and ±1 A
input at a maximum sampling rate of 1.8 MS/s. As an LCR meter, the
FlexDMM delivers 0.25 percent basic accuracy for both inductance and
capacitance at measurement rates up to 40 S/s. Using the intuitive
NI-DMM instrument driver, you can quickly change from function to
function in a single application programming interface (API).
1
• 6
• 1.8 MS/s isolated digitizer
• LCR meter (inductance,
self-calibration
The PXI-4072 offers a DC reading rate from 10 kS/s at 4
2
capacitance, and resistance)
-Digit Digital Multimeter, 1.8 MS/s
1
2
-digit digital multimeter
p
AC measurements
rms
isolation
Operating Systems
• Windows Vista/XP/2000
• Linux
Recommended Software
• LabVIEW
• LabVIEW Real-Time Module
• LabWindows
• LabVIEW SignalExpress
Driver Software (included)
• NI-DMM driver
• DMM Express VI for LabVIEW
• DMM Soft Front Panel
®
1
/CVI
2
digits
Table 1. PXI-4072 Key Specifications
Mode
Capacitance
Inductance
Ordering Information
NI PXI-4072............................................................................778270-01
Includes the P-1 Probe Set, NI-DMM, DMM Express VI, and DMM Soft Front Panel.
Recommended Switching and Accessories
NI PXI-2527
NI PXI-2530
P-1 Probe Set (standard probe) ............................................761000-01
P-2 Probe Set (additional probe) ..........................................184698-01
P-3 Probe Set (banana plug to bare wire) ............................185692-01
10 A current shunt, CSM-10A ..............................................777488-02
BUY NOW!
For complete product specifications, pricing, and accessory information,
call 800 813 3693 (U.S.) or go to ni.com/dmm.
300 V multiplexer switch ................................................ 778572-27
128x1 multiplexer switch..................................................778660-01
300 pF, 1 nF, 10 nF, 100 nF, 1 µF, 10 µF
100 µF, 1000 µF, 10,000 µF
100 mH, 1 H, 5 H
10 µH, 100 µH
1 mH, 10 mH
Ranges
Reading Rate (S/s)
20
40
20
3
3

Related parts for NIPXI-2530

NIPXI-2530 Summary of contents

Page 1

Digital Multimeter, 1.8 MS/s 2 Isolated Digitizer, and LCR Meter NI PXI-4072 • Multifunction device Operating Systems • ⁄ -digit digital multimeter • Windows Vista/XP/2000 2 • 1.8 MS/s isolated digitizer • Linux • ...

Page 2

Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter 2 Specifications Specifications subject to change without notice. For the most current and complete specifications, visit ni.com/modularinstruments. DC Functions DC Voltage ±(ppm of reading + ppm of ...

Page 3

Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter 2 AC Current 2-Year ±(% of reading + % of range) Range (rms) Peak Current Resolution 10 mA ± 100 mA ±200 mA 100 ...

Page 4

Digital Multimeter, 1.8 MS/s Isolated Digitizer, and LCR Meter 2 DC Voltage ±(ppm of reading + ppm of range) Range Input Impedance Flatness Error (20 kHz) 100 mV >10 GΩ,1 MΩ -0. >10 ...

Page 5

NI Services and Support NI has the services and support to meet your needs around the globe and through the application life cycle – from planning and development through deployment and ongoing maintenance. We offer services and service levels to ...

Related keywords