POWER SUP MED 12VOUT 12.5A 150W

CSS150-12

Manufacturer Part NumberCSS150-12
DescriptionPOWER SUP MED 12VOUT 12.5A 150W
ManufacturerTDK Corporation
SeriesCSS150
TypeMedical
CSS150-12 datasheet
 


Specifications of CSS150-12

Power Supply TypeSwitching (Open Frame)Voltage - Output12V
Number Of Outputs1Power (watts)150W
ApplicationsMedicalVoltage - Input90 ~ 264VAC
Mounting TypeChassis Mount1st Output12 VDC @ 12.5A
Size / Dimension5" L x 3" W x 1.3" H (127mm x 76.2mm x 33mm)Power (watts) - Rated150W
Operating Temperature0°C ~ 70°CEfficiency89%
ApprovalsCE, EN, IEC, ULLine Regulation±3%
Load Regulation±3%Power Supply Output TypeFixed
No. Of Outputs1Output Voltage12V
Output Current12.5APower Rating150W
Input Voltage90V AC To 264V AC / 120V DC To 180V DCBrand/seriesCSS150 Series
ConfigurationOpen FrameOperationSwitching
Output12 V @ 8.3 APower, Output150 W
Primary TypeAC-DCSpecial FeaturesDula Input Fuses;Global Operation
Voltage, Input90 to 264 VACVoltage, Output12 V
Safety ApprovalsUL/EN/IEC60601-1 and CE MarkIte Safety ApprovalsUL/EN60950-1
Lead Free Status / RoHS StatusLead free / RoHS Compliant3rd Output-
2nd Output-4th Output-
Other names285-1973  
1
Page 1
2
Page 2
Page 1/2

Download datasheet (652Kb)Embed
Next
F e a t u r e s a n d B e n e f i t s
Feature
Industry standard footprint
Wide Range AC Input
Dual input fuses
S p e c i f i c a t i o n s
ITEMS
Input Voltage range
V
Inrush Current
A
Input Current (115/230VAC)
A
Leakage Current
µA
Hold Up Time (Typ)
ms
Temperature Coefficient
-
Voltage Accuracy
%
Adjustment Range
%
Minimum Load
A
Total Regulation
%
Ripple & Noise
%
Overcurrent Protection
-
Overvoltage Protection
V
Remote On/Off
-
Efficiency
%
Off load power draw
W
Operating Temperature
-
Storage Temperature
°C
Humidity (non condensing)
-
Cooling
-
Withstand Voltage
-
Isolation Resistance
-
Vibration (non operating)
-
Shock
-
Safety Agency Approvals
-
Conducted & Radiated EMI
-
Immunity
-
MTBF
-
Weight (Typ)
g
Size (WxLxH)
mm
Warranty
yrs
(1) Derate linearly to 90W convection rating from 90 to 100VAC input.
3055 Del Sol Blvd • San Diego, CA 92154 • 1-800-526-2324
CSS150 Series
3 x 5" 150W AC-DC Medical Power Supplies
Wide Range AC Input
Low profile, Industry Standard Footprint
Global Safety Agency Compliance
K e y M a r k e t S e g m e n t s & A p p l i c a t i o n s
Medical
Gaming
Test and Measurement
Benefit
Availability to second source
Global Operation
Easier system compliance
90 - 264VAC (47 - 400Hz)(1) or 120 - 180VDC
<60A maximum at 264VAC input, 25°C ambient cold start
<250µA 264VAC 63Hz
16ms at 115VAC input
±0.05%/°C
2% peak to peak
110 - 150%
110 - 150% of nominal (Cycle input power to reset)
Unit on: Floating or low <1.2V, Unit off: Apply 5V with respect to 0V
89% typical
0 to +70°C derate linearly to 50% load from 50 to 70°C
-20 to +85°C
20 - 95% RH
Convection or forced air (2.5m/s)
Input to Ground 1.5kVAC, Input to Output 4kVAC, Output to Ground 500VDC for 1 min.
>100M at 25C & 70%RH, Output to Ground 500VDC
19.6m/s
(10~55Hz:2G Constant, X,Y,Z 60min each.)
2
< 196.1 m/s
UL60601-1, EN60601-1, IEC60601-1,CE Mark, UL60950-1, EN60950-1
EN55011-B, FCC Class B
EN60601-1-2
100,000 (MIL-217F-HDBK)
3 x 5 x 1.3" (including underside components)
CSS150
1.7 / 0.85
±1%
±3%
None
±3%
<0.5W
(20G)
2
340g
Two Years
25

CSS150-12 Summary of contents

  • Page 1

    ... Warranty yrs (1) Derate linearly to 90W convection rating from 90 to 100VAC input. 3055 Del Sol Blvd • San Diego, CA 92154 • 1-800-526-2324 CSS150 Series 3 x 5" 150W AC-DC Medical Power Supplies Wide Range AC Input Low profile, Industry Standard Footprint Global Safety Agency Compliance & ...

  • Page 2

    ... Output Model Voltage (V) CSS150-12 12V 11.64 - 12.36V CSS150-15 15V 14.55 - 15.45V CSS150-24 24V 23.28 - 24.72V CSS150-36 36V 34.92 - 37.08V CSS150-48 48V 46.60 - 49.44V Other Medical Products 40W pcb mount medical CSS 40 - 500W medical NV175 175 - 200W 1-4 outputs NV300 300W 1-4 outputs For Additional Information, please visit us ...