GSC25BG SL Power Electronics Manufacture of Condor/Ault Brands, GSC25BG Datasheet

no-image

GSC25BG

Manufacturer Part Number
GSC25BG
Description
POWER SUPPLY SWITCHER 20W TRIPLE
Manufacturer
SL Power Electronics Manufacture of Condor/Ault Brands
Series
GSC25r
Datasheet

Specifications of GSC25BG

Voltage - Output
5.1V, ±15V
Number Of Outputs
3
Power (watts)
25W
Applications
Commercial
Power Supply Type
Switching (Open Frame)
Voltage - Input
90 ~ 264VAC
Mounting Type
Chassis Mount
1st Output
5.1 VDC @ 2.5A
2nd Output
15 VDC @ 1.5A
3rd Output
-15 VDC @ 200mA
Size / Dimension
4.25" L x 2.5" W x 0.86" H (108mm x 63.5mm x 21.8mm)
Power (watts) - Rated
20W
Operating Temperature
0°C ~ 50°C
Efficiency
70%
Approvals
CE, CSA, EN, TUV, UL
Load Regulation
1%, 4%
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
4th Output
-
• Industry’s smallest 25 W switcher
• 2.50 x 4.25 x 0.86”
• Conducted EMI exceeds FCC Class B and
• Overvoltage protection standard
• Commercial Approved to UL1950,
• Medical Approved to UL2601-1, IEC60601-1
• RoHS Compliant (G suffix)
CISPR 22 Class B (Commercial models) and
CISPR 11 Class B (Medical models)
CSA-C22.2 No.950, EN60950
EN60601-1 and CSA-C22.2 No. 601-1
Ac Input
90-264 Vac, 47-63 Hz single phase.
Input Current
Maximum input current at minimum 120 Vac, 60 Hz with full rated output
load is 0.6 A.
Hold-up Time
15 ms minimum from loss of ac input at full load, nominal line (120 Vac).
Output Power
Normal continuous output power is 25 W, 28 W peak for 60 sec. maximum
duration, 10% duty cycle. Factory set to begin power limiting at approxi-
mately 30 W.
Output Regulation
Regulation from initial setpoint measured by changing load from 5% load to
50% load or 50% load to full load in either direction. Initial setpoint tolerance
is measured at 50% load. A minimum load of 5% of the output current on the
+5.1 V output (125 mA) is required to maintain proper regulation.
Overload Protection
Fully protected against short circuit and output overload. Short circuit protec-
tion is cycling type power limit
Output Noise
0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential mode. Measured with
scope probe directly across output terminals of the power supply with load
terminated with 0.1 uF capacitor.
Transient Response
Main output: 750 μs typical response time for return to within 0.5% of final
value for a 50% load step within the regulation limits of minimum and maxi-
mum load, ∆l/∆t <0.2 A/μs. Maximum voltage deviation is 3.5%. Startup/shut-
down overshoot less than 2%.
Overvoltage Protection
Built in with firing point set per ratings table. OVP firing reduces voltage to
less than 50% of nominal voltage in 50 ms.
Voltage Adjustment
Factory set with fixed resistors to maximize reliability.
Efficiency
70% minimum at full rated load, nominal input voltage.
Input Protection
Internal ac fuse provided on all units.
SPECIFICATIONS
All specifications are typical at nominal input, full load at 25
GLOBAL PERFORMANCE SWITCHERS
Features:
marked to LVD
GSC25 Commercial/GSM25 Medical
O
C unless otherwise stated
25 Watt Global Performance SWitcherS
Inrush Current
Inrush is limited by internal thermistor. The inrush at 230 Vac, averaged over the
first ac half-cycle under cold start conditions will not exceed 32 A.
Temperature Coefficient
0.03%/°C typical on all outputs.
Environmental
Designed for 0 to 50°C operation at full rated output power; derate output current
and total output power by 2.5% per °C between 50 - 70°C. See Environmental and
Packaging Specifications on next page.
EMI/EMC Compliance
All models include built-in EMI filtering to meet the following emissions require-
ments:
EMI SPECIFICATIONS
Conducted Emissions GSC25
Conducted Emissions GSM25
Static Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Commercial Safety:
Condor D.C. Power Supplies, Inc. declares under our sole responsibility that all GSC
models are in conformity with the applicable requirements of EN60950 following
the provisions of the Low Voltage Directive 73/23/EEC. All GSC models are ap-
proved to UL1950, CSA-C22.2 No.950, EN60950.
GSM25 Medical Model Earth Leakage Current
Leakage current measured in the Gnd wire connection when measured per
UL2601-1 or IEC60601-1 is as follows:
Medical Safety
Condor D.C. Power Supplies, Inc. declares under our sole responsibility that all
GSM models are in conformity with the applicable requirements of UL2601-1
Patient Care Equipment, CSA-C22.2 No.601.1, IEC60601-1, EN 60601-1.
GSM25
GSM25
Model
Leakage
Normal
50 μA
94 μA
Leakage
150 μA
78 μA
Fault
COMPLIANCE LEVEL
EN55022 Class B; FCC Class B
EN55011 Class B; FCC Class B
EN61000-4-2, 6 kV contact, 8 kV air
EN61000-4-3, 3 V/meter
EN61000-4-4, 2 kV, 5 kHz
EN61000-4-5, 1 kV diff., 2 kV com.
132 Vca/60 Hz
264 Vca/50 Hz
Voltage
D C P o w e r S u p p l i e s
Test
IEC60601-1
UL2601-1
Method
Test

Related parts for GSC25BG

GSC25BG Summary of contents

Page 1

GLOBAL PERFORMANCE SWITCHERS Features: • Industry’s smallest 25 W switcher • 2.50 x 4.25 x 0.86” • Conducted EMI exceeds FCC Class B and CISPR 22 Class B (Commercial models) and CISPR 11 Class B (Medical models) • Overvoltage protection ...

Page 2

Commercial Medical Model Model Output No. GSC25A GSM25A GSC25B GSM25B GSC25D GSM25D maintain proper regulation on output 2, +5.1 V current must be at least 1⁄4 and not greater ...