MSA150018A ICC/Elpac Power Syst, MSA150018A Datasheet

PWR SUPPLY MED SW 150W 18V 8.3A

MSA150018A

Manufacturer Part Number
MSA150018A
Description
PWR SUPPLY MED SW 150W 18V 8.3A
Manufacturer
ICC/Elpac Power Syst
Series
MSA 150r
Datasheet

Specifications of MSA150018A

Voltage - Output
18V
Number Of Outputs
1
Power (watts)
150W
Applications
Medical
Power Supply Type
Switching (Open Frame)
Voltage - Input
85 ~ 264VAC
Mounting Type
Chassis Mount
1st Output
18 VDC @ 8.3A
Size / Dimension
5" L x 3" W x 1.22" H (127mm x 76.2mm x 30.9mm)
Power (watts) - Rated
150W
Operating Temperature
0°C ~ 70°C
Efficiency
91%
Approvals
CAN/CSA, CB, CE, cTUVus, UL
Line Regulation
±5%
Load Regulation
±5%
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
3rd Output
-
2nd Output
-
4th Output
-
Other names
EPS443
A
1-888-357-2280
saleselpac@iccus.com
www.iccus.com
+44.1383.432920
saleseurope@iccus.com
TECHNICAL SUPPOrT
WOrLDWIDE
EUrOPE
Notes
**Ripple and noise measured with 20MHz bandwidth; 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor.
Input Voltage
Input Frequency
Input Current
Inrush Current
Power Factor
Zero Load
Power Consumption
Earth Leakage Current
(Typical)
Patient Leakage Current
(Typical)
Input
*visit www.iccus.com for complete details
5-YEAr LIMITED WArrANTY*
85 – 264VAC
100 – 240VAC Nominal
47 – 63Hz
<2A rms
<37A at 230VAC cold start
>0.97
<0.5W
<150µA @ 132VAC @ 60Hz
<300µA @ 264VAC @ 60Hz
<50µA @ 132VAC @ 60Hz
<75µA @ 264VAC @ 60Hz
Medical Switch Mode Power Supply
MSA150 Series
Output Voltage
Total Regulation
Minimum Load
Start-Up Delay
Hold-Up Time
Ripple & Noise
Over Voltage Protection
Over Temperature Protection
Over Current Protection
Short Circuit Protection
Output
High Efficiency: Level V
Up to 180W with Forced Air
High Power Density 9.8W/in³
Lifetime Expectation >5 years
Hold-Up Time >25ms at full load
EISA & CEC Compliant
Floating Output
Medical Approval - EN60601-1 Class I
See Table
+/-5%
No minimum load required
<1.5s
>25ms at any
input voltage
<1% pk-pk **
110 – 135%
Active - Recoverable; plus
Passive - Non Recoverable
120 – 180%
Shutdown, auto-restart
(hiccup mode)
150 Watts

Related parts for MSA150018A

MSA150018A Summary of contents

Page 1

TECHNICAL SUPPOrT WOrLDWIDE 1-888-357-2280 saleselpac@iccus.com www.iccus.com EUrOPE +44.1383.432920 saleseurope@iccus.com Input Input Voltage Input Frequency Input Current Inrush Current Power Factor Zero Load Power Consumption Earth Leakage Current (Typical) Patient Leakage Current (Typical) Notes *visit www.iccus.com for complete details **Ripple and ...

Page 2

... Model Number Output Voltage MSA150012A 12.0V MSA150015A 15.0V MSA150018A 18.0V MSA150024A 24.0V Notes 1) With convection cooling. Peak load (180W) lasting up to 500ms with a maximum 10% duty cycle. 2) Sustained output current with minimum 100 LFM 3) Includes initial setting, line regulation, load regulation, and thermal drift. ...