LPS44-M Emerson Network Power, LPS44-M Datasheet

no-image

LPS44-M

Manufacturer Part Number
LPS44-M
Description
PWR SUPPLY 40W 15VOUT SGL MED
Manufacturer
Emerson Network Power
Series
LPS40r
Datasheet

Specifications of LPS44-M

Voltage - Output
15V
Number Of Outputs
1
Power (watts)
40W
Applications
Medical
Power Supply Type
Switching (Open Frame)
Voltage - Input
85 ~ 264VAC
Mounting Type
Chassis Mount
1st Output
15 VDC @ 2.6A
Size / Dimension
5" L x 3" W x 1.14" H (127mm x 76.2mm x 29mm)
Power (watts) - Rated
40W
Operating Temperature
0°C ~ 50°C
Efficiency
70%
Approvals
CE, CSA, UL, VDE
Line Regulation
±2%
Load Regulation
±2%
Output Power Rating
39 W
Input Voltage
85 VAC to 264 VAC
Output Voltage (channel 1)
15 V
Output Current (channel 1)
2.6 A
Mounting Style
Chassis
Dimensions
5 in L x 3 in W x 1.14 in H
Rohs Compliant
Yes
Power Supply Output Type
Fixed
No. Of Outputs
1
Output Voltage
15V
Output Current
2.6A
Power Rating
40W
3rd Output
-
2nd Output
-
4th Output
-
Lead Free Status / Rohs Status
Lead free / RoHS Compliant

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
LPS44-M
Manufacturer:
ARTESYN
Quantity:
1 000
Special Features
• Medical safety approvals
• Universal input
• 3” x 5” footprint
• Less than 1U high
• Remote sense on main output
• Low output ripple
• Adjustable 5V output
• Overvoltage protection
• Overload protection
• 110 KHz switching frequency
• LPX40 enclosure kit available
Safety
(IEC601)
LPS40-M Series
40 Watts Medical
Total Power:
Input Voltage:
# of Outputs:
VDE 0750/EN60601-1
UL
CSA CSA 22.2 No.601.1
CE Mark (LVD)
LR109492
UL2601 E182560
40-55 Watts
85-264 VAC
120-300 VDC
Single
Electrical Specifications
Input
Input range:
Frequency:
Inrush current:
Input current:
Efficiency:
EMI filter:
Safety ground
leakage current:
Output
Maximum power:
Adjustment range:
Hold-up time:
Overload protection:
Overvoltage protection: 5V output; 5.7 to 6.7 VDC. Other outputs 10% to 25% above nominal
Remote sense:
85 - 264 VAC; 120 - 300 VDC
47 - 440 Hz
< 18A peak @ 115 VAC; <36A peak @ 230 VAC, cold start @ 25 °C
1A max. (RMS) @ 115 VAC
70% typical at full load
FCC Class B conducted; CISPR 22 Class A conducted;
EN55022 Class A conducted; VDE 0878 PT3 Class A conducted
< 75µA @ 50/60 Hz, 264 VAC input
40 W for convection; 55 W with 30CFM forced air
-5, +10% minimum
20ms @ 40 W load, 115 VAC nominal line
Short circuit protection on all outputs
Case overload protected @ 110 - 145% above peak rating
output
Compensates for 0.5 V lead drop minimum.
Will operate without remote sense connected.
Reverse connection protected.
Embedded Power for
Business-Critical Continuity
LPS40-M Series
Rev. 9.8.09_28
 of 3

Related parts for LPS44-M

LPS44-M Summary of contents

Page 1

LPS40-M Series 40 Watts Medical Total Power: 40-55 Watts Input Voltage: 85-264 VAC 120-300 VDC # of Outputs: Single Special Features • Medical safety approvals • Universal input • 3” x 5” footprint • Less than 1U high • Remote sense on main output • Low output ripple • Adjustable 5V output • Overvoltage protection • Overload protection • 110 KHz switching frequency ...

Page 2

... MTBF demonstrated: > 550,000 hours at full load and 25 °C ambient conditions Ordering Information Output Minimum Model Number Voltage Load LPS42 LPS43 LPS44 LPS45 Peak current lasting < 30 seconds with a maximum 10% duty cycle °C including initial tolerance, line voltage, load currents and output voltages adjusted to factory settings. 3. Peak-to-peak with 20 MHz bandwidth and 10 µF in parallel with a 0.1 µF capacitor at rated line voltage and load ranges. Power Derating Curve 55W 30 CFM Forced Air Cooling 40W 30W Output ...

Page 3

... Mechanical Drawing 0.22" 4.55" (5.6) (115.6) GND M1 3.00" (76.2) 2.55" (64. 1.14" (29) 0.197" Pin Assignments Connector LPS42-M LPS43-M LPS44-M LPS45-M SK1-1 Neutral Neutral Neutral Neutral SK1-3 Line Line Line Line SK2-1 +5V +12V +15V +24V SK2- 2 +5V +12V +15V +24V SK2- 3 +5V +12V +15V +24V ...