LPT62-M Emerson Network Power, LPT62-M Datasheet - Page 2

no-image

LPT62-M

Manufacturer Part Number
LPT62-M
Description
PWR SUPPLY 60W 5/12/12V TRPL MED
Manufacturer
Emerson Network Power
Series
LPT60r
Datasheet

Specifications of LPT62-M

Voltage - Output
5V, ±12V
Number Of Outputs
3
Power (watts)
60W
Applications
Medical
Power Supply Type
Switching (Open Frame)
Voltage - Input
85 ~ 264VAC
Mounting Type
Chassis Mount
1st Output
5 VDC @ 7A
2nd Output
12 VDC @ 3A
3rd Output
-12 VDC @ 700mA
Size / Dimension
5" L x 3" W x 1.69" H (127mm x 76.2mm x 42.9mm)
Power (watts) - Rated
60W
Operating Temperature
0°C ~ 50°C
Efficiency
70%
Approvals
CE, CSA, UL, VDE
Line Regulation
±2%, ±5%
Load Regulation
±2%, ±5%
Output Power Rating
35 W
Input Voltage
85 VAC to 264 VAC
Output Voltage (channel 1)
5 V
Output Current (channel 1)
7 A
Output Voltage (channel 2)
12 V
Output Current (channel 2)
3 A
Output Voltage (channel 3)
- 12 V
Output Current (channel 3)
0.7 A
Mounting Style
Chassis
Dimensions
5 in L x 3 in W x 1.65 in H
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
4th Output
-
Lead Free Status / Rohs Status
Lead free / RoHS Compliant

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
LPT62-M
Manufacturer:
ASTEC
Quantity:
1 000
Environmental Specifications
Operating temperature:
Electromagnetic
susceptibility:
Humidity:
Vibration:
Storage temperature:
Temperature coefficient:
MTBF demonstrated:
Ordering Information
LPT62-M
LPT63-M
1. Peak current lasting <30 seconds with a maximum 10% duty cycle.
2 . At 25°C including initial tolerance, line voltage, load currents and output voltages adjusted to factory settings.
3 . Peak-to-peak with 20 MHz bandwidth and 10μF in parallel with a 0.1μF capacitor at rated line voltage and load ranges.
4. Minimum Loads are required.
Number
Model
Voltage
Output
+5V
+12V
-12V
+5V
+15V
-15V
Minimum
0.7A
0.3A
0A
0.7A
0.3A
0A
Load
0° to 50°C ambient derate each output at 2.5% per degree
from 50° to 70°C, -20°C startup
Designed to meet IEC 801, -2, -3, -4, -5, -6, Level 3
Operating; non-condensing 5% to 95%
Three orthogonal axes, sweep at 1 oct/min, 5 min. dwell
at four major resonances 0.75G peak 5 Hz to 500 Hz,
operational
-40° to 85°C
±.04% per °C
>550,000 hours at full load and 25°C ambient conditions
with Convnection
Maximum Load
cooling
7A
3A
0.7A
7A
2.8A
0.7A
Maximum Load
with 30CFM
Forced Air
8A
3.5A
1A
8A
3.3A
1A
Peak Load
10A
6A
1.5A
10A
4A
1.5A
1
Regulation
±2%
±5%
±5%
±2%
±5%
±5%
2
Ripple P/P (PARD)
50mV
120mV
120mV
50mV
150mV
150mV
3
Embedded Power for
Business-Critical Continuity
LPT60-M Series
Rev. 06.15.07
2 of 3

Related parts for LPT62-M