MWC100015A-12A ICC/Elpac Power Syst, MWC100015A-12A Datasheet

no-image

MWC100015A-12A

Manufacturer Part Number
MWC100015A-12A
Description
POWER SUP MEDICAL 100W 15V 6.7A
Manufacturer
ICC/Elpac Power Syst
Series
MWC100r
Datasheet

Specifications of MWC100015A-12A

Form
Desktop (3 Prong without Cord)
Voltage - Output Per Connector
15VDC
Current - Output
6.6A
Cord Plug
5 Pin DIN
Voltage - Input
85 ~ 264VAC
Power - Output
100W
Transformer Type
Switching
Efficiency
87%
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Polarization
-
Other names
EPS433
A
1-888-357-2280
saleselpac@iccus.com
www.iccus.com
+44.1383.432920
saleseurope@iccus.com
TECHNICAL SUPPOrT
wOrLDwIDE
EUrOPE
Notes
**Ripple and noise measured with 20MHz bandwidth; 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor.
Input Voltage
Input Frequency
Input Current
Inrush Current
Power Factor
Zero Load
Power Consumption
Touch Current/
Leakage Current
Patient Leakage Current
Input
*visit www.iccus.com for complete details
5-YEAr LIMITED wArrANTY*
85 – 264VAC
100 – 240VAC Nominal
47 – 63Hz
<1.5A rms
<37A at 230VAC cold start
>0.97
<0.5W
<150µA @ 132VAC @ 60Hz
<250µA @ 264VAC @ 60Hz
<50µA @ 132VAC @ 60Hz
<80µA @ 264VAC @ 60Hz
MWC100 Series
Medical Switch Mode Power Supply
Output Voltage
Total Regulation
Minimum Load
Start-Up Delay
Hold-Up Time
Ripple & Noise
Over Voltage Protection
Over Temperature Protection
Over Current Protection
Short Circuit Protection
Output
High Efficiency: Level V
Power Factor Correction
Lifetime Expectation >5 years
Hold-Up Time >48ms at full load
EISA Compliant
CEC Compliant
Floating Output
Medical Approval - EN60601-1 Class I
See Table
+/-5%
No minimum load required
<1s
>48ms at any
input voltage
<1% pk-pk **
110 – 135%
Active - Recoverable; plus
Passive - Non Recoverable
120 – 180%
Shutdown, auto-restart
(hiccup mode)
100 watts

Related parts for MWC100015A-12A

MWC100015A-12A Summary of contents

Page 1

TECHNICAL SUPPOrT wOrLDwIDE 1-888-357-2280 saleselpac@iccus.com www.iccus.com EUrOPE +44.1383.432920 saleseurope@iccus.com Input Input Voltage Input Frequency Input Current Inrush Current Power Factor Zero Load Power Consumption Touch Current/ Leakage Current Patient Leakage Current Notes *visit www.iccus.com for complete details **Ripple and noise ...

Page 2

... Model Number Output Voltage 1 MWC100012A-12A 12.0V MWC100015A-11A 15.0V MWC100018A-11A 18.0V MWC100024A-11A 24.0V Notes 1) All models ship standard with US version input cable. 2) Maximum peak load (120W) lasting 500ms with a maximum 10% duty cycle. 3) Includes initial setting, line regulation, load regulation, and thermal drift. ...