HIH-4000-003 Honeywell, HIH-4000-003 Datasheet - Page 2

Board Mount Humidity Sensors SENSING ACCURACY3.5% lead pitch SIP

HIH-4000-003

Manufacturer Part Number
HIH-4000-003
Description
Board Mount Humidity Sensors SENSING ACCURACY3.5% lead pitch SIP
Manufacturer
Honeywell
Datasheet

Specifications of HIH-4000-003

Rh Range
0 % to 100 %
Rh Accuracy
+/- 3.5 %
Package / Case
SIP
Output Type
Analog
Supply Voltage (max)
5.8 V
Supply Current
500 uA
Maximum Operating Temperature
+ 85 C
Minimum Operating Temperature
- 40 C
Mounting Style
Through Hole
Lead Free Status / RoHS Status
Lead free / RoHS Compliant

Available stocks

Company
Part Number
Manufacturer
Quantity
Price
Part Number:
HIH-4000-003
Manufacturer:
HONEYWELL
Quantity:
1 000
HIH-4000 Series
TABLE 1. PERFORMANCE SPECIFICATIONS (At 5 Vdc supply and 25 ºC [77 ºF] unless otherwise noted.)
FACTORY CALIBRATION DATA
HIH-4000 Sensors may be ordered with a calibration and data
printout. See Table 2 and the order guide on the back page.
TABLE 2. EXAMPLE DATA PRINTOUT
2
Parameter
Interchangeability (first order curve)
Accuracy (best fit straight line)
Hysterisis
Repeatability
Settling time
Response time (1/e in slow moving air)
Stability (at 50% RH)
Voltage supply
Current supply
Voltage output (1
Temperature compensation
Output voltage temperature, coefficient at
50% RH, 5 V
Operating temperature
Operating humidity
Storage temperature
Storage humidity
Specific Notes:
1. Can only be achieved with the supplied slope and offset.
2. Device is calibrated at 5 Vdc and 25 ºC.
3. Non-condensing environment.
Model
Channel
Wafer
MRP
Calculated values at 5 V
Linear output for 3.5% RH
accuracy at 25 °C
Ratiometric response for
0% RH to 100% RH
V
V
Zero offset
Slope
RH
V
0% RH to 59% RH
60% RH to 100% RH
For HIH-4000-003 and HIH-4000-004 catalog listings only.
OUT
OUT
OUT
at 0% RH
at 75.3% RH
www.honeywell.com/sensing
st
order curve fit)
HIH-4000-003
92
030996M
337313
0.826 V
3.198 V
0.826 V
31.483 mV/%RH
(V
(V
V
SUPPLY
OUT
OUT
- zero offset)/slope
- 0.826)/0.0315
(0.1652 to 0.7952)
V
True RH = (Sensor RH)/(1.0546 – 0.00216T), T in ºC
OUT
Minimum
-40[-40]
-50[-58]
=(V
-3.5
-8
-5
4
0
SUPPLY
)(0.0062(sensor RH) + 0.16), typical at 25 ºC
See Figure 1.
See Figure 1.
See Figure 2.
General Notes:
• Sensor is ratiometric to supply voltage.
• Extended exposure to >90% RH causes a reversible shift
• Sensor is light sensitive. For best performance, shield
Typical
±0.5
200
1.2
15
-4
of 3% RH.
sensor from bright light.
3
Maximum
125[257]
85[185]
+3.5
500
100
5.8
70
5
8
mV/ºC
% RH
% RH
% RH
% RH
% RH
% RH
ºC[ºF]
% RH
ºC[ºF]
% RH
Unit
Vdc
ms
μA
s
Specific Note
1
2
3
3

Related parts for HIH-4000-003