DMM4040 Tektronix, DMM4040 Datasheet - Page 11

Digital Multimeters 6.5 DIGIT (35 PPM) PRECISION METER USB

DMM4040

Manufacturer Part Number
DMM4040
Description
Digital Multimeters 6.5 DIGIT (35 PPM) PRECISION METER USB
Manufacturer
Tektronix
Datasheet

Specifications of DMM4040

Accuracy
0.0035 %
True Rms
Yes
Resistance Range
10 Ohms to 1 GOhms
Frequency
47 Hz to 440 Hz
Equipment Type
Digital Multimeter
Display Count
6.5
Current, Rating
100 μA to 10 A, 100 pA Resolution
Dimensions
3.46 × 8.56 × 11.7 (88 × 217 × 297) In. (mm)
Display Type
Vacuum Fluorescent Display, dot matrix
Frequency Range
47 to 440 Hz
Includes
National Instrument's LabVIEW SignalExpress™ TE Limited Ed.
Power Consumption
28 VA peak (12 W average)
Resolution
6.5 Digit Resolution
Temperature, Operating
0 to 55 °C
Temperature, Storage
–40 to 70 °C
Voltage, Rating
100 mV to 1000 V, 100 nV Resolution
Warranty
3 Years
Weight
8.0 (3.6) Lb. (kg)
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
AC Current
The following AC current specifications are for sinusoidal signals with amplitudes
greater than 5% of range. For inputs from 1% to 5% of range, add an additional
error of 0.1% of range.
Input Characteristics
*
*
*
Characteristic
Input Protection
Measurement Method
AC Filter Bandwidth
Maximum Crest Factor
Additional Crest Factor
Errors (<100 Hz)
Range
100 μA
1 mA
10 mA
100 mA
400 mA*
1 A*
3 A*
10 A*
5
8
9
400 mA available in software version 1.0.700.18 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1 minute off; maximum crest factor 3:1 at 400 mA.
Part of 10 A range.
Available at front-panel connectors only.
Slow
Medium
Fast
8
5, 8
8
9
Description
Tool-accessible 11 A / 1000 V and 440 mA / 1000 V
fuses, limits of 400 mA continuous 550 mA for 2 minutes
on, 1 minute off
AC-coupled true RMS, DC-coupled to the fuse and shunt
(no blocking capacitor)
3 Hz to 10 kHz
20 Hz to 10 kHz
200 Hz to 10 kHz
5:1 at Full Scale
Crest factor 1-2, 0.05% of full scale
Crest factor 2-3, 0.2% of full scale
Crest factor 3-4, 0.4% of full scale
Crest factor 4-5, 0.5% of full scale
Only applies to non-sinusoid signals
Resolution
1.000000 mA
10.00000 mA
100.0000 mA
100.0000 μA
400.000 mA
1.000000 A
10.00000 A
3.00000 A
4½ Digits
100 nA
100 μA
100 μA
10 nA
10 μA
1 mA
1 mA
1 μA
Resolution
5½ Digits
100 nA
100 μA
100 μA
10 nA
10 μA
10 μA
1 nA
1 μA
Digital Multimeters — Tektronix DMM4050 and DMM4040
6½ Digits
100 pA
100 nA
10 nA
10 μA
10 μA
1 nA
1 μA
1 μA
Shunt Resistance
www.tektronix.com/dmm4050_4040
(Ohms)
0.01 Ω
0.01 Ω
0.01 Ω
100 Ω
100 Ω
1 Ω
1 Ω
1 Ω
Burden Voltage
<0.015 V
<0.025 V
<0.15 V
<0.25 V
<0.50 V
<0.05 V
<0.05 V
<0.5 V
11