DMM4050 Tektronix, DMM4050 Datasheet - Page 5

Digital Multimeters 6.5 DIGIT (24 PPM) PRECISION METER USB

DMM4050

Manufacturer Part Number
DMM4050
Description
Digital Multimeters 6.5 DIGIT (24 PPM) PRECISION METER USB
Manufacturer
Tektronix
Datasheet

Specifications of DMM4050

Accuracy
0.0024 %
True Rms
Yes
Resistance Range
10 Ohms to 1 GOhms
Frequency
47 Hz to 440 Hz
Equipment Type
Digital Multimeter
Display Count
6.5
Current, Rating
100 μA to 10 A, 100 pA Resolution
Dimensions
3.46 × 8.56 × 11.7 (88 × 217 × 297) In. (mm)
Display Type
Vacuum Fluorescent Display, dot matrix
Frequency Range
47 to 440 Hz
Includes
National Instrument's LabVIEW SignalExpress™ TE Limited Ed.
Power Consumption
28 VA peak (12 W average)
Resolution
6.5 Digit Resolution
Temperature, Operating
0 to 55 °C
Temperature, Storage
–40 to 70 °C
Voltage, Rating
100 mV to 1000 V, 100 nV Resolution
Warranty
3 Years
Weight
8.0 (3.6) Lb. (kg)
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Electrical Specifications
DC Voltage Specifications
Accuracy specifications are valid for 6½ digit resolution mode after at least a 1 hour
warm-up with Auto Zero enabled.
24-hour specifications are relative to calibration standards and assume a controlled
electromagnetic environment per EN 61326-1:2000-11.
Input Characteristics
*
Characteristic
Maximum Input
Common Mode
Rejection
Normal Mode Rejection
Measurement Method
A/D Linearity
Input Bias Current
Auto-zero Off Operation Following instrument warm-up at calibration temperature
Analog Filter
DC Ratio
Settling Considerations
Range
100 mV
1 V
10 V
100 V
1000 V
2
Inputs beyond ±14 V are clamped through 200 kΩ typical. 10 MΩ is default input impedance.
Description
1000 V on any range
140 dB at 50 or 60 Hz ±0.1% (1 kΩ unbalance)
60 dB for NPLC of 1 or greater with analog filter off and
power line frequency ±0.1%
100 dB for NPLC of 1 or greater with analog filter on and
power line frequency ±0.1%
Multiramp A/D
0.0002% of measurement + 0.0001% of range
<30 pA at 25 °C
±1 °C and less than 10 minutes, add error: 0.0002%
range additional error +5 μV
When using the analog filter, specifications are relative
to within one hour of using the ZERO function for that
range and NPLC setting
Accuracy is ±(Input accuracy + Reference accuracy),
where Input accuracy = DC Voltage accuracy for the HI
to LO Input (in ppm of the Input voltage), and Reference
accuracy = DC Voltage accuracy for the HI to LO
(Sense) Reference (in ppm of the Reference voltage)
Measurement settling times are affected by source
impedance, cable dielectric characteristics, and input
signal changes
100.0000 mV
Resolution
1,000.000 V
1.000000 V
10.00000 V
100.0000 V
4½ Digits
100 mV
100 μV
10 mV
10 μV
1 mV
Digital Multimeters — Tektronix DMM4050 and DMM4040
Resolution
5½ Digits
100 μV
10 mV
10 μV
1 mV
1 μV
6½ Digits
100 nV
100 μV
10 μV
1 mV
1 μV
www.tektronix.com/dmm4050_4040
Input Impedance
10 MΩ or >10 GΩ*
10 MΩ or >10 GΩ*
10 MΩ or >10 GΩ*
10 MΩ ±1%
10 MΩ ±1%
2
2
2
5