00043 Simpson, 00043 Datasheet - Page 11

no-image

00043

Manufacturer Part Number
00043
Description
Test Probes PROBE SET THREADED W
Manufacturer
Simpson
Datasheet

Specifications of 00043

Lead Free Status / RoHS Status
Lead free / RoHS Compliant
from 50V to 1000V only.
The Instrument is equipped with AC and DC measuring ranges up to 1000V. This is
advantageous for testing for the absence of voltage and the discharge of capacitive
devices under test.
The 501 features a quick testing signal lamp within the test probe that illuminates the
work area. The signal lamp can also be used for quick Pass/Fail evaluations of
insulation resistances. Minimum values for insulation resistances are present if the
lamp lights up.
Comes equipped with built-in test probes, an alligator clip, a leatherette carrying case
and batteries.
• Five Test Voltages: 50, 100, 250, 500, and 1000 VDC
• Voltage Measurement to 1000 V AC/DC
• Test Probe Signal Lamp Illuminates Work Area and Indicates Pass/Fail of
• Battery Powered with Battery Test and Battery Charge Indicator
• Rugged Tester Design for Industrial Applications with Leatherette Carrying Case
Measuring Ranges
INSULATION RESISTANCE
DIRECT AND ALTERNATING CURRENT
The model 501 can be used for general purpose insulation tests that require voltages
ACCESSORIES
501 Insulation Tester, 1000 V
200K …20M
400K …40M
200K …10M
400K …20M
Ordering Information
1M …100M
2M …200M
4M …400M
100K …5M
20K …1M
40K …2M
Measuring
Insulation Resistance
Measuring
0 to 1000V
Range
0…100K
0…250K
0…500K
Range
AC/DC
0…50K
0…1M
0…4
Open-Circuit Voltage U
0 or 40 to 200Hz
Frequency
1200V
120V
300V
600V
60V
9V
Resistance
o
Catalog No.
Internal
900K
Internal Resistance R
12501
200K
300K
400K
600K
200K
800K
150K
100K
30K
10K
40K
60K
20K
80K
50K
1200V AC/DC
Maximam
Continuous
Overload
i
Lights for Resistance
Lamp in Test Prod
(1) Switching Accuracy
(1) Referring to scale length
Intrinsic Error (1)
> 100K
> 200K
> 500K
> 1M
> 2M
2.5%
1)
9
Specifications
REFERENCE CONDITIONS
Ambient Temperature: 73 F (23 C)
Relative Humidity: 45 to 55%
Measurement Magnitude Frequency: 45-65 Hz (voltage measurements)
Line Voltage Waveform: Sine, deviation between effective and rectified value < .5%
Battery Voltage: 9V
Operating Position: Horizontal
ELECTROMAGNETIC COMPATIBILITY EMC
Interference Emission:
EN 50081-1: 1992
Interference Immunity:
EN 50082-1: 1992
ELECTRICAL SAFETY
Designed to IEC 1010-1
Protection Class: II
Test Voltage:
Overvoltage Classification: II
Pollution Degree: 2
Protection: Housing IP52
AMBIENT CONDITIONS
Operating Temp. Range: 32 F (0 C) to 104 F (40 C)
Storage Temp. Range: -4 F (-20 C) to 140 F (60 C) (without batteries)
Relative Humidity: Max. 45-55%, condensation must be avoided
MECHANICAL DESIGN
Dimensions: W 6.50” (165mm) x
D 4.33” (110mm) x H 4.92” (125mm)
Weight: 3.5 lbs. (1.6kg) (with batteries)
POWER SUPPLY
Standard or Storage Battery: 6 ea. 1.5V “D” size batteries
Working Range: 7V to 10V
Battery Service Life: Number of possible measurements with 1 set of “D” size
batteries (with lamp in test prod switched off): at least 3,000 measurements of an
insulation resistance of 1M (Nominal Voltage of1000 V, 5 sec. Measurement duration,
switched off for 25 sec. etc.)
501 Insulation Tester
5.55Kv
.5V
4 F (2 C)

Related parts for 00043