215-4 NEWPORT ELECTRONICS, 215-4 Datasheet - Page 3

Voltage Meter

215-4

Manufacturer Part Number
215-4
Description
Voltage Meter
Manufacturer
NEWPORT ELECTRONICS
Datasheet

Specifications of 215-4

Meter Function
DC Voltmeter
Meter Range
± 200mV To ± 200V
Digit Height
14.2mm
Power Consumption
1W
Operating Temperature Range
0°C To +55°C
Meter Signal Input
Voltage
Meter Output
Voltage
Lead Free Status / RoHS Status
na
Unpacking & Inspection
This instrument has no power-on switch. An external switch or circuit-breaker shall be included in the
building installation as a disconnecting device. It shall be marked to indicate this function, and it shall be
in close proximity to the equipment within easy reach of the operator. The switch or circuit-breaker shall
not interrupt the Protective Conductor (Earth wire), and it shall meet the relevant requirements of
IEC 947–1 and IEC 947-3 (International Electrotechnical Commission). The switch shall not be
incorporated in the mains supply cord.
Furthermore, to provide protection against excessive energy being drawn from the mains supply in case of
a fault in the equipment, an overcurrent protection device shall be installed.
EMC Considerations
Unpack the instrument and inspect for obvious shipping damage. Do not attempt to operate the
unit if damage is found.
This instrument is a panel mount device protected in accordance with Class I of EN 61010
(115/230 AC power connections). Installation of this instrument should be done by Qualified
personnel. In order to ensure safe operation, the following instructions should be followed.
This device is marked with the international Caution symbol.
manual before installing or commissioning this device as it contains important information
relating to Safety and EMC (Electromagnetic Compatibility).
The Protective Conductor must be connected for safety reasons. Check that the power
cable has the proper Earth wire, and it is properly connected. It is not safe to operate this
unit without the Protective Conductor Terminal connected.
Do not exceed voltage rating on the label located on the top of the instrument housing.
Always disconnect power before changing signal and power connections.
Do not use this instrument on a work bench without its case for safety reasons.
Do not operate this instrument in flammable or explosive atmospheres.
Do not expose this instrument to rain or moisture.
Whenever EMC is an issue, always use shielded cables.
Never run signal and power wires in the same conduit.
Use signal wire connections with twisted-pair cables.
Install Ferrite Bead(s) on signal wires close to the instrument if EMC problems persist.
SAFETY CONSIDERATIONS
i
It is important to read this

Related parts for 215-4