Engineering
Scaling to
8
Display
Units
8.3.4 ENTERING ZERO OFFSET NUMBERS
You have two (2) ways to enter a zero offset. The first and easiest is
using the TARE function. This is accomplished by connecting a
momentary contact to the rear P2 connector at P2-1 and P2-4.
Each time this momentary contact is activated, the display will
automatically display zero.
The second method for entering a zero offset number is Reading
Offset (“RDG OF”).
The RDG OF (Reading Offset) menu item should be used if the
meter shows a nonzero reading with zero input. The offset value
zeroes the display by cancelling out the nonzero reading.
If your meter displays a positive reading at zero input, you must
enter a negative offset value. If your meter displays a negative
reading at zero input, you must enter a positive offset value.
If you are using an active decimal point (RDG.2=0), your offset
value will be the negative of the display reading at zero point.
If you are using the more common independent decimal point (the
factory setting, RDG.2=1), follow these instructions to convert the
display reading to the appropriate offset value:
1. Note the display reading at zero input, ignoring the decimal
2. Shift the count value to the left side of the decimal point.
3. If the count value is positive, make it negative by replacing the
Example 1: Your meter displays 000.003 when the input is zero.
The count value is 000003. Shift this value to the left side of the
decimal point: 003.000. Change the leading zero to a minus sign:
-03.000. This is the “converted offset value” you will use for
configuring RDG OF.
Example 2: Your meter displays -00.003 when the input is zero. The
count value is -00003. Shift this value to the left side of the
decimal point: -03.000. Change the leading minus sign to a zero:
003.000. This is the “converted offset value” you will use for
configuring RDG OF.
point. This reading represents the count value - the number of
whole counts that need to be offset.
leading digit (the left-most digit) with a minus sign. If the count
value is negative, make it positive by replacing the negative sign
with a zero.
8-10
Related parts for INFS-0210-DC1
Image
Part Number
Description
Manufacturer
Datasheet
Request
R
Part Number:
Description:
Strain Gauge Meter
Manufacturer:
NEWPORT ELECTRONICS
Datasheet:
Part Number:
Description:
Strain Gauge Meter
Manufacturer:
NEWPORT ELECTRONICS
Datasheet:
Part Number:
Description:
Strain Gauge Meter
Manufacturer:
NEWPORT ELECTRONICS
Datasheet:
Part Number:
Description:
DISPLAYS / ELECTRONIC,INFB ULTRA HIGH PERFORMANCE METER,DISPLAYS / ELECTRONIC,PHOTOMOS RELAY,METERS,INF-B ULTRA HIGH PERFORMANCE METER ,NEWPORT ELECTRONICS
Manufacturer:
NEWPORT ELECTRONICS
Part Number:
Description:
DISPLAYS / ELECTRONIC,INFB ULTRA HIGH PERFORMANCE METER,DISPLAYS / ELECTRONIC,PHOTOMOS RELAY,METERS,INF-B ULTRA HIGH PERFORMANCE METER ,NEWPORT ELECTRONICS
Manufacturer:
NEWPORT ELECTRONICS
Part Number:
Description:
DISPLAYS / ELECTRONIC,INFB ULTRA HIGH PERFORMANCE METER,DISPLAYS / ELECTRONIC,PHOTOMOS RELAY,METERS,INF-B ULTRA HIGH PERFORMANCE METER ,NEWPORT ELECTRONICS
Manufacturer:
NEWPORT ELECTRONICS
Part Number:
Description:
DISPLAYS / ELECTRONIC,INFB ULTRA HIGH PERFORMANCE METER,DISPLAYS / ELECTRONIC,PHOTOMOS RELAY,METERS,INF-B ULTRA HIGH PERFORMANCE METER ,NEWPORT ELECTRONICS
Manufacturer:
NEWPORT ELECTRONICS
Part Number:
Description:
NEWPORT ULTRA-MINIATURE TEMPERATURE TRANSMITTERS
Manufacturer:
NEWPORT ELECTRONICS
Part Number:
Description:
Isolation Amplifier
Manufacturer:
NEWPORT ELECTRONICS
Datasheet:
Part Number:
Description:
Signal Conditioning, Output Isolator
Manufacturer:
NEWPORT ELECTRONICS
Datasheet:
Part Number:
Description:
Signal Conditioning, Output Isolator
Manufacturer:
NEWPORT ELECTRONICS
Part Number:
Description:
Signal Conditioning, Output Isolator
Manufacturer:
NEWPORT ELECTRONICS
Datasheet: