6240 AEMC, 6240 Datasheet
6240
Specifications of 6240
Available stocks
6240 Summary of contents
Page 1
... Micro-ohMMeter Model 6240 Selectable test current and resistance ranges 0.25% accuracy and 1µΩ resolution Simple one button “press-to-start” operation Software temperature correction for accurate measurements Automatically takes multiple measurements by simply moving the test probes from point to point Front panel buttons for polarity reversal, data storage, ...
Page 2
... Resistance measurements are automatically calculated from 1µΩ to 400Ω with resolutions down to 1µΩ. The Model 6240 has selectable test currents of 10A, 1A, 100mA and 10mA and includes a polarity reversal function. The Model 6240 is designed to conduct tests on both resistive and inductive material ...
Page 3
... Yes Yes DataView ® is included with the Model 6240. cAtALoG No. 400Ω 0.25% ± 2cts 100mΩ 10mA ...
Page 4
AC line recharging receptacle Kelvin terminals * Function buttons allow direct access to data storage and retrieval, polarity reversal, auto/manual measurement and displayed values. Functional DisplaY Power save indicator Battery charge indicator Caution indicator Polarity indicator Overheating indicator Reading hold ...
Page 5
... Customize views, templates and reports to your exact needs Create and store a library of configurations that can be used with the Model 6240 as needed Display test results Print reports using standard included templates or custom templates you design Statistical presentation of measurements. ...
Page 6
... Fax (508) 698-2118 Chauvin Arnoux , Inc. ® d.b.a. AEMC Instruments ® 200 Foxborough Blvd. Foxborough, MA 02035 For full product specifications, please visit our website at www.aemc.com sales@aemc.com Chauvin Arnoux One Source ™ 950.BR-6240_0210 Printed in the USA ...