FLUKE-789 Fluke Electronics, FLUKE-789 Datasheet - Page 49

PROCESSMETER DMM W/4-20MA SOURCE

FLUKE-789

Manufacturer Part Number
FLUKE-789
Description
PROCESSMETER DMM W/4-20MA SOURCE
Manufacturer
Fluke Electronics
Series
ProcessMeter™r
Type
Digital (DMM)r
Datasheets

Specifications of FLUKE-789

Includes
Battery, Clips, Test Leads
Style
Handheld
Display Digits
4.5
Display Type
LCD, Dual
Display Count
20000
Function
Voltage, Current, Resistance, Frequency
Functions, Extra
Continuity, Diode Test
Features
Auto Off, Backlight, Hold, Min/Max/Ave
Ranging
Auto/Manual
Response
True RMS
Lead Free Status / RoHS Status
Lead free / RoHS compliant by exemption
Other names
1633984
614-1075
789
AC Current Measurement
DC Current Measurement
AC conversion: true rms
Maximum crest factor: 3 (between 50 and 60 Hz)
For non-sinusoidal waveforms, add ( 2 % reading + 2 % f.s.) typical
Overload protection 440 mA, 1000 V fast-blow fuse
Note: 440 mA continuous, 1 A 30 seconds maximum
Specifications are valid from 5 % to 100 % of amplitude range.
Note: 440 mA continuous, 1 A 30 seconds maximum
Overload protection: 440 mA, 1000 V fast-blow fuse
45 Hz to 2 kHz
1.000 A (Note)
30.000 mA
1.000 A (Note)
Range
Range
Resolution
Resolution
0.001 A
0.001 mA
0.001 A
Accuracy, (% of Reading + Counts)
Accuracy, (% of Reading + Counts)
1 % + 2
0.05 % + 2
0.2 % + 2
Typical Burden
Typical Burden
1.5 V/A
14 mV/mA
1.5 V/A
ProcessMeter
Voltage
Voltage
Specifications
41