FLUKE-88-5 Fluke Electronics, FLUKE-88-5 Datasheet - Page 82

no-image

FLUKE-88-5

Manufacturer Part Number
FLUKE-88-5
Description
MULTIMETER AUTOMOTIVE 88V
Manufacturer
Fluke Electronics
Series
8xr
Type
Digital (DMM)r
Datasheets

Specifications of FLUKE-88-5

Includes
Battery, Hanger, Temperature Probe, Test Leads
Style
Handheld
Display Digits
4.5
Display Type
LCD, Bar Graph
Display Count
20000
Function
Voltage, Current, Resistance, Capacitance, Temperature, Frequency
Functions, Extra
Continuity, Diode Test
Features
Auto Off, Backlight, Hold, Min/Max/Ave
Ranging
Auto/Manual
Response
Average
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
2550519
Users Manual
1.
72
600 mV dc
600 mV ac
6 V
60 V
600 V
1000 V
Maximum input for specified accuracy = 10X range or 1000 V.
Input Range
1
70 mV (to 400 Hz)
150 mV
0.3 V
3 V
30 V
100 V
Table 18. Frequency Counter Sensitivity and Trigger Levels
5 Hz - 20 kHz
Minimum Sensitivity
70 mV (to 400 Hz)
150 mV
0.7 V
7 V ( 140 kHz)
70 V ( 14.0 kHz)
200 V ( 1.4 kHz)
0.5 Hz - 200 kHz
40 mV
1.7 V
4 V
40 V
100 V
Approximate Trigger Level
(DC Voltage Function)