79696040 Crouzet USA, 79696040 Datasheet - Page 76

no-image

79696040

Manufacturer Part Number
79696040
Description
TEMP PROBE ACCES, 6 MM DIA SLIDING CONNECTOR
Manufacturer
Crouzet USA
Series
-r
Datasheet

Specifications of 79696040

Accessory Type
Connector
Lead Free Status / Rohs Status
Lead free / RoHS Compliant
For Use With/related Products
Millenium 3
XB26
76
Part numbers
Type
XA03
XA04W
Accessories
Type
M3 SOFT
General characteristics
See General characteristics for the XA04 analogue extension on page 36, except for the adapted characteristics below:
Certifications
Conformity with the EMC directive
Inputs
Operating range
Input impedance
Maximum non destructive current/voltage
Resolution
Value of LSB
Input type
Conversion time
Sampling time
Accuracy at 25°C ambient temperature
Accuracy at 55°C ambient temperature
Temperature compensation
Isolation between analogue channel and
power supply
Dedicated isolated 24 V DC output for 4-20
mA sensor
Cable length
Protection against polarity inversions
Millenium 3 Custom
XA03: 3 Pt 100 temperature inputs in the same casing
XA04W: Mix of inputs in the same casing: Pt 100, pH,
ORP (Redox), Current (4 - 20 mA)
"Application-specific" examples:
- Regulation and measurement of (XA03)
- pH and Redox sensors for treating water in
swimming pools and fountains (XA04W)
Extensions compatible with any Millenium 3 Custom
expandable logic controller
For Pt100 probes, see page 54.
For pH and ORP probes, see page 78. The probes are
directly connected to the XA04W extension
"Application-specific" analogue extensions for XD10/XB10 and XD26/
Input
3 Pt 100 (-25
1 Pt 100 (0-50°C), 1 pH (0-14), 1 ORP
(0-1000mV), 1 current (4-20 mA)
Description
Multilingual programming software containing specific library functions (CD-ROM)
+125°C)
In accordance with
89/336/EEC:
EN (IEC) 61131-2
(Zone B)
EN (IEC) 61000-6-2,
EN (IEC) 61000-6-3,
Pt 100 (IP, IQ, IR)
-25°C, + 125°C
-
-
10 bits
0.15°C
Pt 100 probe
IEC 751
3-wire
Module cycle time
<1s
± 1°C
± 1°C
-
None
-
10 m max. with
shielded cable
88970800
UL, CSA,
-
EN (IEC) 61000-6-4
Supply
Via the 24 V
Via the 24 V
sensor)
88972805
UL & CSA
In accordance with
89/336/EEC:
EN (IEC) 61000-6-1
EN (IEC) 61000-6-3
Pt 100 (IP)
0-50°C
-
-
12 bits
0.012°C
Pt 100 probe
IEC 751
3-wire
Module cycle time
4s
± 0.8°C
± 0.8°C
-
None
-
3 m max. with
shielded cable
-
(1 dedicated output 24 V
XA03
controller
controller
pH (IQ)
0 - 14
> 10
-
12 bits
0.0034 pH
pH probe
Module cycle time
4s
± 0.05 pH
± 0.05 pH
No
Drift of 0.03 pH
from15 to 25°C
Drift of 0.15 pH
from 0 to 50°C
Isolated
-
3 m max. with
shielded cable
-
12
Ω
± 5% 0.6 W to supply the 4-20 mA
ORP (IR)
0 - 1000 mV
> 10
-
12 bits
0.24 mV
ORP probe
Module cycle time
4s
± 5 mV
± 5 mV
-
Isolated
-
3 m max. with
shielded cable
-
12
Ω
www.millenium3.crouzet.com
XA04W
4-20 mA (IS)
0 - 20 mA
10 Ω
30 mA
12 bits
4.9 μA
Common mode
Module cycle time
4s
± 0.1 mA
± 0.1 mA
-
Isolated
24 V
3 m max. with
shielded cable
Yes
Code
88970111
Code
88970800
88972805