LCR400 AIM-TTI INSTRUMENTS, LCR400 Datasheet - Page 3

LCR Meter 0.1% Basic Accuracy

LCR400

Manufacturer Part Number
LCR400
Description
LCR Meter 0.1% Basic Accuracy
Manufacturer
AIM-TTI INSTRUMENTS
Datasheet

Specifications of LCR400

Inductance Measuring Range
0.001H To 9900H
Parameters Measured:
Parameter Selection:
Measurement Modes:
Range Hold:
Zero Function:
Measurement Freq.:
Displayed Functions:
Resistance:
Inductance:
Capacitance:
D:
Q:
Accuracy Limits:
R
L
C
(R accuracies apply for Q<0.1. L accuracies apply for Q>10.
C accuracies apply for D<0.1 and after Null).
Limits Set-up:
Binning:
Component Connection: 4-terminal connection via internal fixture for
Internal Fixture:
Bias Voltage:
Protection:
Display Type:
Annunciators:
Displayed Functions:
Interface Type:
Interface Function:
LCR400 low-cost precision LCR bridge
MEASUREMENT FUNCTIONS
MEASURING RANGES
LIMITS COMPARATOR
COMPONENT CONNECTION
DISPLAY
DIGITAL INTERFACE
Designed and manufactured in Europe
0.1% ± 1 digit
0.5% ± 1 digit
0.1% ± 1 digit
0.5% ± 1 digit
0.1% ± 1 digit
0.5% ± 1 digit
2% 1 digit
2% ± 1 digit
2% ± 1 digit
Tel: +44 (0)1480 412451 Fax: +44 (0)1480 450409
Glebe Road, Huntingdon. Cambs. PE29 7DR U.K.
Email: sales@tti-test.com Web: www.tti-test.com
Thurlby Thandar Instruments Ltd.
500pF-20000µF
0.8mH-2500H
0.2mH-9900H
10nF-1000µF
2nF-5000µF
0.1 -20M
4mH-500H
0.4 -5M
2 -1M
100Hz
R, L, C, D, & Q.
Manual or automatic selection of R, L or C.
Series or parallel equivalent circuit.
Prevents autoranging when changing
components.
Nulls out up to 100pF at test fixture.
Selectable as 100Hz/120Hz, 1kHz, 10kHz.
R+Q, L+Q, C+D, C+R.
0.1m to 990M
0.001H to 9900H
0.001pF to 99000µF
0.001 to 999
0.001 to 999
Multiple Upper and Lower limits can be set
from keyboard or from RS-232 interface
Up to 8 Pass bins can be defined plus Fail on
minor parameter and general Fail.
both radial and axial components.
External 4-terminal connection via adaptor.
Sprung jaws for vertical lead insertion. Plug-in
sliding pillars for horizontal lead insertion.
Switchable 2V polarising voltage. External
bias up to 50V can be applied.
Can withstand charged capacitors up to 50V
and up to 1 Joule of total energy.
Dual 5-digit 0.56” LEDs.
LED annunciators for all functions and mea-
surement units.
R+Q, L+Q, C+D, C+R, PASS/FAIL, Bin No.
RS-232 via 9 pin D connector, 9600 baud.
Full command and readback capability.
200pF-500µF
50pF-2000µF
20µH-1000H
0.1 -10M
0.4mH-50H
80µH-250H
1nF-100µF
2 -500k
0.4 -2M
1kHz
0.4 -200k
0.1 -500k
100pF-10µF
20pF-50µF
5pF-200µF
2µH-100H
2 -50k
40µH-5H
8µH-25H
10kHz
Thurlby Thandar Instruments Ltd. operates a policy of continuous development and reserves
the right to alter specifications without prior notice.
Keyboard:
Non-volatile memory:
Input Voltage:
Temperature Range:
Safety:
EMC:
Size:
Weight:
SMD Tweezers:
BNC Adaptor:
Kelvin Clip Leads:
LCR PC-Link:
GENERAL
OPTIONAL ACCESSORIES
Full numeric keyboard.
Up to 9 complete set-ups.
230V or 115V ±14%, 50/60Hz, internally ad-
justable, 25VA max. Installation Category ll.
+5
-40
Complies with EN61010-1.
Complies with EN61326.
360 x 240 x 95 mm including feet.
2.9kg.
Four terminal tweezers for measuring surface
mount components. Connects directly into
standard test fixture.
Plugs into the standard component fixture to
convert it into a set of four BNC connectors.
Supplied complete with four BNC to BNC
leads.
High quality Kelvin Clip test lead set terminat-
ing in BNC plugs (requires BNC adaptor).
Windows based software which uses the
RS232 interface to provide results logging and
to store/restore instrument set-ups.
o
o
C to 40
C to 70
o
C operating 20 - 80% RH,
o
C storage.
82100-1050 Iss. X