Specifications of U3402A

Dmm Type
Bench
Voltage Measuring Range Dc
120mV To 1000V
Voltage Measuring Range Ac
120mV To 750V
Current Measuring Range Dc
12mA To 12A
Current Measuring Range Ac
12mA To 12A
Lead Free Status / RoHS Status
na
* U3402A only
** U3401A only
Features
measurement speeds*
Up to 120,000 counts resolution
Up to 0.012% basic DCV accuracy
11 basic measurements and up to
six built-in math functions
Dual display on bright VFD
Selectable resolutions for variable
Kensington lock slot security
Agilent U3400 Series 4½ and 5½ Digit
Digital Multimeters
Simply right for your basic needs,
without compromising quality
The U3400 Series digital multimeters
(DMMs) come with functions vital for your
everyday measurements: DC, AC and AC+
DC voltage and current, 2- and 4-wire
resistance, frequency, continuity and
diode tests.
Math functions dBm, Min/Max, Relative,
Compare, Hold and Percentage** are also
available to ease your measurement
analysis.
Not only are the U3400 Series built robust
for lasting reliability over the long haul,
these DMMs also ensure dependable
measurements with up to 0.012% basic
DCV accuracy.
Data Sheet
Basic + Good = Elegant Simplicity
Efficient testing with dual display
and selectable resolutions*
Dual display on the U3400 Series lets you
view two parameters simultaneously as you
measure, and further enhances your
troubleshooting tasks. For example, the
ability to see both AC voltage and
frequency simultaneously would help you
to measure the frequency response of
amplifier circuits more efficiently and
effectively. Refer to Page 3 for more typical
dual display combinations and applications.
Up to three measurement speeds are
available on the U3402A: Slow, Medium and
Fast. This means greater flexibility in
catering to different testing needs: a faster
measurement speed at lower resolution or
a higher resolution at slower measurement
speed.
Physical security
Your instruments may be at risk of theft or
misplacement whenever you leave them
unattended on the bench. With the U3400
Series’ rear Kensington lock slot, you can
secure your DMM and be assured that it is
where you expect it to be for your continued
testing the next day.

U3402A Summary of contents

Page 1

... Refer to Page 3 for more typical dual display combinations and applications three measurement speeds are available on the U3402A: Slow, Medium and Fast. This means greater flexibility in catering to different testing needs: a faster measurement speed at lower resolution or a higher resolution at slower measurement speed ...

Page 2

... Take a closer look Input terminals and current fuse Figure 1 Front panel of the U3402A. The front panel of the U3401A is similar with slight differences in the position of certain functions. Please refer to the U3401A’s User’s and Service Guide U3401-90001 for details. AC power connector Figure 2 Rear panel of the U3401A/U3402A ...

Page 3

... Adjustment of AC motor control Power supply AC ripple and DC current measurement Current dissipation measurement RF frequency measurement Frequency response checking 3 Available in U3402A, U3401A U3402A, U3401A U3402A U3402A U3402A U3402A U3402A U3402A U3402A, U3401A U3402A, U3401A U3402A, U3401A U3402A, U3401A U3401A U3402A, U3401A U3402A, U3401A U3402A ...

Page 4

U3400 Series specifications DC voltage DCV resolution, full-scale reading and accuracy [± reading + count)] U3401A 4.5-digit DMM Range Resolution 500. µV 5.0000 V 100 µV 50.000 500. 1000.0 V ...

Page 5

... V 100 mV [1] Accuracy specified at input >5% of full scale. [2] Input voltage <200 V RMS. [3] For 5 kHz to 10 kHz, accuracy is 0.7% + 15. [4] In VAC 750 V range, 1000.0 V RMS is readable with audio warning. U3402A 5.5-digit DMM Rate Range Resolution Slow 120.000 mV 1 µV 1.20000 V 10 µ ...

Page 6

... V 100 mV [1] Accuracy specified at input >5% of full scale. [2] Input voltage <200 V RMS. [3] In VAC 750 V range, 1000.0 V RMS is readable with audio warning. [4] For 5 kHz to 10 kHz, accuracy is 0.7% + 25. U3402A 5.5-digit DMM Rate Range Resolution Slow 120.000 mV 1 µV 1.20000 V 10 µ ...

Page 7

... A 100 µA 10.000 [1] Typical at full-scale reading and voltage across the input terminals. [ range, > ADC is readable for 20 seconds maximum with audio warning. U3402A 5.5-digit DMM Rate Range Slow 12.0000 mA 120.000 mA 1200.00 mA 12.0000 A Medium 40.000 mA 120.00 mA 1200 ...

Page 8

... Accuracy specified at input >5% of full scale and >1 A for 10 A range unless otherwise stated. [2] Typical at full-scale reading and voltage across the input terminals. [3] Input current >35 µA RMS. [4] Input current <3 A RMS. [ range, > ADC is readable for 20 seconds maximum with audio warning. U3402A 5.5-digit DMM Rate Range Resolution Slow 12.0000 mA 0.1 µ ...

Page 9

... Accuracy specified at input >5% of full scale and >1 A for 10 A range unless otherwise stated. [2] Typical at full-scale reading and voltage across the input terminals. [3] Input current >35 µA RMS. [ range, > ADC is readable for 20 seconds maximum with audio warning. U3402A 5.5-digit DMM Rate Range Resolution Slow 12 ...

Page 10

... W 500.00 kW 100 W 5.0000 MW 50.000 [1] To reduce noise interference that may be induced by the test leads, we recommend using a shielded test cable for measuring resistances above 500 kW. [2] Relative (REL) operation is used. U3402A 5.5-digit DMM Rate Range [1] 120.000 W Slow 1.20000 kW 12.0000 kW 120.000 kW 1.20000 MW 12 ...

Page 11

... Hz to 100 kHz 500 RMS 5 V 0.25 V RMS 50 V 2.5 V RMS 500 RMS 750 RMS U3402A 5.5-digit DMM Range Measurement range 1200. 1200 Hz 12.0000 kHz kHz 120.000 kHz 100 Hz to 120 kHz 1.0000 MHz 1 kHz to 1 MHz [1] Specified accuracy at input >5% of full scale. ...

Page 12

... In VDC 1000 V range, 1200 V is readable. In VAC 750 V range, 1000 V is readable. [3] Reading is displayed in dB when relative (REL) operation is used. [4] Input voltage <200 V RMS. [5] For input voltage within the frequency range kHz. U3402A 5.5-digit DMM Rate Voltage range [1][2] Slow 120 ...

Page 13

... Approximately 1.0 second when the displayed reading reaches 99.9% DC value of the tested input signal at the same range AC-coupled true RMS: measures the AC component with up to 400 VDC bias on any range Maximum 3:0 at full scale U3401A parallel with <100 pF U3402A ± parallel with <120 pF U3401A: 1000 V RMS/1400 V PEAK U3402A: 750 V RMS/1200 V PEAK 2x10 7 ...

Page 14

... U3401A: 1000 V RMS on all ranges U3402A: 750 V RMS on all ranges Approximately 1.5 seconds (2.5 seconds for U3402A) when the displayed reading reaches 99.9% AC RMS value of the tested input signal at the same range DC coupled to the fuse and current shunt, AC coupled true rms measurement (measures the AC ...

Page 15

... Maximum 3:0 at full scale 10% of range to full-scale input on all ranges; auto or manual range selection 0.1 second or 1 period of the input signal, whichever is longer U3401A parallel with <100 pF U3402A ± parallel with <120 pF U3401A: 1000 V RMS/1400 V PEAK U3402A: 750 V RMS/1100 V PEAK 7 2x10 ...

Page 16

... Medium: 0.01 dB for all ranges Fast: 0.1 dB for all ranges 110 W, 124 W, 125 W, 135 W, 150 W, 250 W, 300 W, [2] [2] [2] [2] 500 W, 600 W, 800 W, 900 W, 1000 W, 1200 W, 8000 W U3401A: dBm, Relative, Min/Max, Compare, Hold, Percentage U3402A: dBm, Relative, Min/Max, Compare, Hold RS-232 (for calibration use only) 16 ...

Page 17

... RS-232 (for calibration use only) Dimensions (W × H × D) 255 mm × 105 mm × 305 mm (with bumpers) 215 282 mm (without bumpers) Weight 3.44 kg (with bumpers) Warranty One year for U3401A/U3402A Three months for standard shipped accessories Calibration cycle One year Warm-up time At least 30 minutes ...

Page 18

Ordering Information Each U3400 Series includes these standard shipped accessories: Quick Start Guide Product Reference CD Certificate of Calibration 34138A Test Lead Kit AC Power Cord Accessories 34138A Test Lead Kit 34133A Precision Electronic Test Leads U1161A Extended Test Lead ...

Page 19

... Elegant Simplicity Some of the best equations express large ideas with elegant simplicity. The same is true with the Agilent U3401A and U3402A dual-display DMMs. When you need basic capabilities with good performance at 4½ or 5½ digits, respectively, these affordable benchtop units make the grade in educational, electronics and communications applications ...