M252A GOSSEN METRAWATT, M252A Datasheet
M252A
Specifications of M252A
M252A Summary of contents
Page 1
... Special instrument for the performance of measurements at current transformer circuits Applications The instruments of the E- and high resolution series are extremely rugged, reliable digital multimeters with housings made of impact resistant ABS. With a resolution of 60,000 counts and different measur- ing functions, they’ve been developed for professional use. ...
Page 2
... Adherence to the specification is confirmed by means of the included DKD calibration certificate, which is valid worldwide (recognized by EA and ILAC. If any type of DMM is due to recalibra- tion (recommended intervals years) the multimeters can inexpensively recalibrated at the factory or at any calibration labo- ratory. Applicable Regulations and Standards ...
Page 3
... Input sensitivity, sinusoidal signal, 10% to 100% of the measuring range 9) Plus sensor deviation 10) Off-time > 30 min and T *) without ZERO max. ± 15 digit Key for U > 100 measuring range counts, rdg. = measured value (reading) ⏐E ⏐E XTRA TECH SPECIAL TRMS Digital Multimeters ±(... % rdg. + ... d) ±(... % rdg. + ... ...
Page 4
... METRAHIT⏐E XTRA TRMS Digital Multimeters Internal Clock Time format DD.MM.YYYY hh:mm:ss Resolution 0.1 s ±1 min. per month Accuracy Temperature Influence 50 ppm/K Influencing Quantities and Influence Error Influencing Sphere of Measured Quantity / Quantity Influence Measuring Range V V 600 Ω ... 6 MΩ ...
Page 5
... For voltage Intermittent signal at above 1000 V For current Intermittent signal at above 10 A continuous signal at above 16 A GMC-I Gossen-Metrawatt GmbH ⏐E METRAHIT⏐E XTRA TRMS Digital Multimeters Fuse for METRAHIT E Fuse Electrical Safety . Per IEC 61010-1:2001/VDE 0411-1:2002 METRAHIT E , METRAHIT E XTRA Safety class ...
Page 6
... BASE METRAwin ® METRAwin ment data logging program for recording, visualizing and documen- ting measured values of multimeters of the METRAHIT e-series. Communication between the PC and the measuring instrument(s) is established via available interface adapters. Depending upon device type, one or several of the following ope- . ...
Page 7
... BASE For METRA HIT⏐ multimeters (with/without protective rub- ber holster) and METRAport M250A XTRA F836 Ever-Ready Case For multimeter (without protective rubber holster) and accessories M253A TECH M252A SPECIAL M251A BASE Z216C GTZ3240000R0001 GTZ3409000R0001 Designation Z102A Imitation leather without protective GTZ3408000R0001 rubber holster for METRA HIT⏐ ...
Page 8
... WZ11A AC clamp current transformer 1 ... 200 A~ Z3511 AC clamp current transformer 4 ... 500 A~ Z3512 AC clamp current transformer 0.5 ... 1000 A~ Z3514 AC clamp current transformer 1 ... 2000 A ~ Shunt Resistors for Multimeters without Current Measuring Function NW300 Plug-in shunt resistor, 0 ... 300 mA mA encapsulated NW3A Plug-in shunt resistor, 0 ... 3 A encapsulated ● ...