N9923AU-308 AGILENT TECHNOLOGIES, N9923AU-308 Datasheet - Page 9

no-image

N9923AU-308

Manufacturer Part Number
N9923AU-308
Description
Vector Voltmeter Software Upgrade Key
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of N9923AU-308

Features
Adds Vector Voltmeter Capability
For Use With
N9923A FieldFox Vector Network Analyzer
Lead Free Status / RoHS Status
na
Vector voltmeter used for cable trimming
Power meter display
9
9
Broadband calibration
FieldFox allows you to make broadband
calibrations, which means the instrument
is calibrated over the maximum frequency
range. After a broadband calibration, you can
change the frequency range or number of
points without recalibrating the instrument.
The calibration is interpolated, and accuracy is
maintained.
User calibration kit support
FieldFox supports many Agilent standard cali-
bration kits. In addition to CalReady
and QuickCal, FieldFox also provides a
comprehensive calibration utility. To obtain
the most accurate measurement, users
need to use cal kits that match their device
connector types. FieldFox allows users to
defi ne their own mechanical calibration kits.
Power meter
FieldFox can connect with the Agilent U2000
Series USB power sensors to make RF/
microwave power measurements up to 24 GHz.
FieldFox provides true average power
measurements with a high dynamic range
from –60 dBm to +44 dBm (sensor dependent).
The sensor has an internal zeroing function
and the sensor does not require external
calibration.
Electrical delay and port extension
For in-fi xture measurements, use FieldFox’s
port extension or electrical delay capability to
easily extend the reference plane to the device
interface for accurate measurements. You can
use the electrical delay capability to measure
deviation from linear phase by removing the
linear portion of the phase delay.
FieldFox