53230A-400 AGILENT TECHNOLOGIES, 53230A-400 Datasheet - Page 15

Frequency Counter With GPIB Port, 12, 0.2 Ppm

53230A-400

Manufacturer Part Number
53230A-400
Description
Frequency Counter With GPIB Port, 12, 0.2 Ppm
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 53230A-400

Bandwidth
350MHz
Counter Measuring Functions
Frequency, Period,Totalize, Ratio, Time Interval, Rise/Fall, Width, Burst Width, Duty Cycle, Phase, PRI, PRF
Lead Free Status / RoHS Status
na
Timebase
Timebase Uncertainty = ( Aging + Temperature + Calibration Uncertainty )
1. All Timebase Aging Errors apply only after an initial 30-days of continuous powered operation and for a constant altitude ±100 m.
2. Only use the Factory Calibration error values for the period before your first re-calibration. Factory Calibration uncertainty includes
3. Warm-up error applies when the instrument is powered on in a stable operating environment.
4. Retrace error may occur whenever the instrument line-power is removed or whenever the instrument is battery operated and the
Front/rear view of 53230A
Aging
Temperature (typ)
Calibration uncertainty
Supplemental characteristics (typ)
After the first 1-year of operation, use ½ x (30-day and 1-year) aging rates shown.
the instrument settability error, the factory calibration source uncertainty, and additional timebase uncertainty due to factory
calibration before the required initial 30-days of powered operation. Settability defines the resolution increments you can reach is in
steps of 0.1 ppb (0.01 ppb on Option 010).
When moved between different operating environments add the Temperature error during the initial 30-minutes of powered operation
battery fully discharges. Retrace error is the residual timebase shift that remains 72-hours after powering-on an instrument that has
experienced a full power-cycle of the timebase. Additional frequency shift errors may occur for instrument exposure to severe
impact shocks >50 g.
24-hour, T
30-day, T
1-year, T
0 ºC to 55 ºC relative to 25 °C
T
Initial factory calibration
5-min. warm-up error
72-hour retrace error
Allan deviation t = 1s
CAL
1
± 5 °C
(spec)
CAL
CAL
CAL
Timebase
±5 °C
±5 °C
±1 °C
4
3
2
(typ)
± 0.2 ppm (typ)
± 0.5 ppm
± 0.5 ppm
Standard
< 50 ppb
± 1 ppm
± 1 ppm
± 1 ppm
TCXO
1 ppb
15
Ultra-High Stability OCXO
± 0.3 ppb (typ)
Option 010
± 0.5 ppb
± 10 ppb
± 50 ppb
± 50 ppb
± 10 ppb
0.01 ppb
± 5 ppb
< 2 ppb