U2001H AGILENT TECHNOLOGIES, U2001H Datasheet - Page 7

USB POWER SENSOR, 6GHZ, -50DBM To +30DBM

U2001H

Manufacturer Part Number
U2001H
Description
USB POWER SENSOR, 6GHZ, -50DBM To +30DBM
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of U2001H

Frequency Measuring Range
10MHz To 6GHz
Amplitude Accuracy
± 4%
Standing Wave Ratio
1.15
Connector Type
N
External Height
36mm
External Width
46mm
External Depth
207mm
Weight
0.324kg
Lead Free Status / RoHS Status
na
Specifications
Specifications contained in this chap-
ter are valid ONLY after proper calibra-
tion of the power sensor and apply to
continuous wave (CW) signals unless
otherwise stated. The recommended
calibration interval for this product is
one year. Specifications apply over a
temperature range 0 to +55 °C unless
otherwise stated. Specifications
quoted over a temperature range
of 25 °C ± 10 °C apply to a relative
humidity of 15% to 75% and conform
to the standard environmental test
conditions. Specifications are valid
after a 30-minute warm-up period.
Supplemental characteristics, shown
in italics, are intended to provide use-
ful information with regard to apply-
ing the power sensors in that they
contain typical, but non-warranted
performance parameters. These
characteristics are shown in italics
or denoted as “typical”, “nominal” or
“approximate”.
Measurement speed
Normal: 20 readings/s
x2: 40 readings/s
Fast: 110 readings/s
Buffered (50 readings):
1000 readings/s
The U2000 Series USB sensors have
two measurement modes:
Average only mode (default mode):
optimized for wide dynamic range. In
this measurement mode, a trigger can
be controlled externally via TTL input.
Normal
age power measurement in a defined
time interval (time-gated measure-
ment) with reduced dynamic range.
A trigger can be derived from an RF
signal (internal trigger) or controlled
externally via TTL input (external
trigger).
1. The 1000 reading/s is the derived measurement speed from the first 50 readings in buffered mode. The maximum number of measurements that can
2. Not applicable for U2004A.
be obtained in one second is 250 readings in buffered mode.
2
mode: used for making aver-
1
Frequency and power ranges
Maximum SWR
Model
U2000A
U2001A
U2002A
U2004A
U2000B
U2001B
U2000H
U2001H
U2002H
Model
U2000A
U2001A
U2002A
U2004A
U2000B
U2001B
U2000H
U2001H
U2002H
Frequency range
10 MHz to 18 GHz
10 MHz to 6 GHz
50 MHz to 24 GHz
9 kHz to 6 GHz
10 MHz to 18 GHz
10 MHz to 6 GHz
10 MHz to 18 GHz
10 MHz to 6 GHz
50 MHz to 24 GHz
Frequency range
10 to 30 MHz
30 MHz to 2 GHz
2 to 14 GHz
14 to 16 GHz
16 to 18 GHz
10 to 30 MHz
30 MHz to 2 GHz
2 to 6 GHz
50 MHz to 2 GHz
2 to 14 GHz
14 to 16 GHz
16 to 18 GHz
18 to 24 GHz
9 kHz to 2 GHz
2 to 6 GHz
10 MHz to 2 GHz
2 to 12.4 GHz
12.4 to 18 GHz
10 MHz to 2 GHz
2 to 6 GHz
10 MHz to 8 GHz
8 to 12.4 GHz
12.4 to 18 GHz
10 MHz to 6 GHz
50 MHz to 8 GHz
8 to 12.4 GHz
12.4 to 18 GHz
18 to 24 GHz
7
Power range
–60 to +20 dBm
–60 to +20 dBm
–30 to +44 dBm
–50 to +30 dBm
–50 to +30 dBm
Maximum SWR
(25 °C ± 10 °C)
1.15
1.13
1.19
1.22
1.26
1.15
1.13
1.19
1.13
1.19
1.22
1.26
1.30
1.13
1.19
1.12
1.17
1.24
1.12
1.17
1.15
1.25
1.28
1.15
1.15
1.25
1.28
1.30
Maximum power
+25 dBm avg, 20 VDC
+33 dBm pk, < 10 µs
+25 dBm avg, 5 VDC
+33 dBm pk, < 10 µs
+45 dBm avg, 20 VDC
+47 dBm pk, 1 µs
+33 dBm avg, 20 VDC
+50 dBm pk, 1 µs
+33 dBm avg, 10 VDC
+50 dBm pk, 1 µs
Maximum SWR
(0 °C to 55 °C)
1.21
1.15
1.20
1.23
1.27
1.21
1.15
1.20
1.15
1.20
1.23
1.27
1.30
1.15
1.20
1.14
1.18
1.25
1.14
1.18
1.17
1.26
1.29
1.17
1.17
1.26
1.29
1.31