Specifications of 10070D

Test Probe Type
Oscilloscope
Test Probe Functions
Frequency
For Use With
Infiniium 8000 Series Oscilloscopes
Lead Free Status / RoHS Status
na
Infiniium Performance Characteristics
Vertical: scope channels
Input channels
Analog bandwidth at 50 Ω (–3 dB)*
Calculated rise time
Input impedance*
Sensitivity
Input coupling
Hardware bandwidth limit
Vertical resolution
Channel-to-channel isolation
(any two channels with
equal V/div settings)
DC gain accuracy*
Maximum input voltage*
Offset range
Offset accuracy*
Dynamic range
DC voltage measurement accuracy*
1 MΩ
50 Ω
1 MΩ
50 Ω
Dual cursor
Single cursor
2
2
3
2,4
1
at 50 Ω
2,4
DSO8064A, MSO8064A, DSO8104A, MSO8104A
DSO8064A/DSO8104A: 4 analog
MSO8064A/MSO8104A: 4 analog + 16 digital
DSO8064A/MSO8064A: 600 MHz
DSO8104A/MSO8104A: 1 GHz
DSO8064A/MSO8064A: 583 ps
DSO8104A/MSO8104A: 350 ps
1 MΩ ± 1% (13 pF typical), 50 Ω ± 1.5%
1 mV/div to 5 V/div (1 MΩ)
1 mV/div to 1 V/div (50 Ω)
1 MΩ: AC, DC; 50 Ω:DC
20 MHz
8 bits, ≥12 bits with averaging
DC to 50 MHz: 50 dB
>50 MHz to 500 MHz: 40 dB
>500 MHz to 1 GHz: 30 dB
± 1.25% of full scale at full resolution channel scale
150 V RMS or DC, CAT I
± 250 V (DC + AC) in AC coupling
5 Vrms, CAT I
Vertical sensitivity
1 mV to <10 mV/div
10 mV to <20 mV/div
20 mV to <100 mV/div
100 mV to <1 V/div
1 V to 5 V/div
1 mV to <5 mV/div
5 mV to <200 mV/div
200 mV to 1 V/div
± (1.25% of channel offset +2% of full scale +1 mV)
± 8 div from center screen (1 MΩ)
± 12 div from center screen (50 Ω)
± [(DC gain accuracy)+(resolution)]
± [(DC gain accuracy)+(offset accuracy)+(resolution/2)]
Example for single cursor accuracy for 70 mV signal, 10 mV/div, 0 offset: Accuracy =
± [1.25% (80 mV) + (1.25% (0) + 2% (80 mV) + 1 mV ) + (0.4%/2) (80 mV)] = ±3.8 mV
Available offset
± 2 V
± 5 V
± 10 V
± 20 V
± 100 V
± 2 V
± 5 V
± 20 V
19