1147A AGILENT TECHNOLOGIES, 1147A Datasheet - Page 7

Current Oscilloscope Probe, AC/DC, With AutoProbe Interface

1147A

Manufacturer Part Number
1147A
Description
Current Oscilloscope Probe, AC/DC, With AutoProbe Interface
Manufacturer
AGILENT TECHNOLOGIES
Datasheet

Specifications of 1147A

Bandwidth
50MHz
For Use With
Agilent Technologies 6000 & DSO5000 Series Oscilloscopes
Lead Free Status / RoHS Status
na
Infi niiMax Active Probes and Accessories
Agilent 1130A Infi niiMax high-performance active probe system
• 1.5 GHz InfiniiMax probe system
• InfiniiMax probe amplifier supports
• Unrivaled InfiniiMax probing
• Compatible with 5000, 6000, and
The 1.5-GHz 1130A InfiniiMax probe
amplifier is a perfect complement to
the InfiniiVision 1 GHz models. Its
1.5-GHz bandwidth, extremely low
input capacitance (0.32 pF), high
common mode rejection, and the
patented resistor probe tip technology
provide ultra low loading of the
DUT and superior signal fidelity.
Agilent’s innovative InfiniiMax 1130A
differential probe is the easiest-to-
use, and highest performance probing
system available for high-speed
digital design, and represents a
new industry standard for accuracy,
flexibility, and reliability. Designers
can achieve 1-GHz system bandwidth
in conjunction with 1-GHz 6000 and
7000 Series oscilloscopes even when
manually browsing with the probe
or making hands-off measurements.
Optional solder-in probe heads and
solder-in sockets, as well as browser
configuration provide full bandwidth
at the probe tip.
both differential- and single-ended
measurements for a more cost-
effective solution
accessories support browsing,
solder-in, and socketed use models
at the maximum performance
available
7000 Series oscilloscopes (except
for 6000 Series 100-MHz models)
Agilent 1130A Infi niiMax probe offers you the highest performance available for
measuring differential and single-ended signals
Operating characteristics
Probe bandwidth (-3 dB)
Rise and fall time
(10 to 90%)
System bandwidth (-3 dB)
Input capacitance
Input resistance
Input dynamic range
Input common mode range ±6.75 Vdc to 100 Hz; ±1.25 V > 100 Hz
Maximum signal slew rate
DC attenuation
Offset range
Maximum input voltage
ESD tolerance
Maximum number of
probes supported by
3000 X/5000/6000/7000
Series
7
> 1.5 GHz
233 ps
1130A with MSO/DSO610xA, and MSO/DSO7104A/B: 1 GHz
Cm = 0.1 pF Cm is between tips
Cg = 0.34 pF Cg is ground for each tip
Cdiff = 0.27 pF Differential mode capacitance = Cm + Cg/2
Cse = 0.44 pF Single-ended mode capacitance = Cm + Cg
Differential mode resistance = 50 kΩ ± 1%
Single-ended mode resistance = 25 kΩ ± 1%
±2.5 V
18 V/ns when probing a single-ended signal
30 V/ns when probing a differential signal
10:1 ± 3% before calibration on oscilloscope
10:1 ± 1% after calibration on oscilloscope
±12.0 V when probing single-ended
30 Vpeak, CAT I
> 8 kV from 100 pF, 300 Ω HBM
2