2
2.3.6 Thermocouple Input (OS532, OS533, OS534)
°F
Using the Handheld Infrared Thermometer
The thermometer accepts thermocouple input. It displays
thermocouple temperature and the target temperature (via
infrared) simultaneously. This function provides an accurate
method of determining an unknown emissivity.
• To Determine an unknown target emissivity
LCK
1.
2.
3.
4.
5.
6.
Connect a contact thermocouple probe (Type K) to the
thermometer as shown in Figure 1-3.
Measure the object temperature using the thermocouple
probe.
Aim at the object and measure the temperature via infrared.
Press and hold the
mode (E) appears.
Set the emissivity by pressing the
the temperature reading matches the thermocouple
temperature measurement.
The thermometer now displays the correct object
emissivity.
°F
"TC--- is” is displayed when the thermocouple
input is open or out of range (0 to 1600°F).
key until the Emissivity Display
or
keys until