K32-DICN Omron, K32-DICN Datasheet

no-image

K32-DICN

Manufacturer Part Number
K32-DICN
Description
CABLE SPECIAL EVENT INPUT
Manufacturer
Omron
Datasheet

Specifications of K32-DICN

Lead Free Status / RoHS Status
Lead free / RoHS Compliant
Other names
K32DICN
New High-speed, High-precision Temperature
Indicator
• Capable of high-speed sampling at 50 times per second (20 ms).
• High-resolution of 0.01°C with platinum-resistance thermometer
• Temperature input shift is easily set using two points.
• Series expanded to include DeviceNet models.
• UL certification approval (Certification Mark License).
• CE Marking conformance by third party assessment body.
• Water-resistant enclosure conforms to NEMA 4X (equivalent to IP66).
Model Number Structure
■ Model Number Legend
Base Units and Optional Boards can be ordered individually or as sets.
Base Units
1. Input Sensor Code
5. Supply Voltage
Optional Boards
Sensor Power Supply/Output Boards
Relay/Transistor Output Boards
Event Input Boards
Note: 1. CPA can be combined with relay outputs only.
Accessories (Sold Separately)
K32-DICN: Special Cable (for event inputs, with 8-pin connector)
K32-BCD: Special BCD Output Cable
K3HB-H@
K33-@
K34-@
K35-@
Pt100 input. Thermocouple sensor inputs also support a resolution of
0.1°C for all ranges.
Temperature Indicator
K3HB-H
TA: Temperature input
100-240 VAC: 100 to 240 VAC
24 VAC/VDC: 24 VAC/VDC
Display
2. Only one of the following can be used by each Digital Indicator: RS-232C/RS-485 communications, a linear output, or DeviceNet communications.
0
Thermocouple input/Platinum-resistance thermometer input
Refer to Safety Precautions for All Digital Panel
Meters.
2
3
4
1
Input
compensation
value 1
Input
compensation
input value 1
5
After
compensation
Before
compensation
Input
compensation
value 2
Input
compensation
input value 2
Input
Base Units with Optional Boards
2. Sensor Power Supply/Output Type Code
3. Relay/Transistor Output Type Code
4. Event Input Type Code
Rubber Packing
Note: Rubber packing is provided with the Controller.
K3HB-H@-@@@
K32-P1
None:
CPA:
L1A:
L2A:
A:
FLK1A: Communications (RS-232C) + Sensor power supply
FLK3A: Communications (RS-485) + Sensor power supply
None:
C1:
C2:
T1:
T2:
BCD∗: BCD output + transistor output (NPN open collector: HH/H/PASS/
DRT:
∗ A Special BCD Output Cable (sold separately) is required.
None:
1:
2:
3:
4:
None
Relay output (PASS: SPDT) + Sensor power supply
(12 VDC +/−10%, 80 mA) (See note 1.)
Linear current output (0 to 20 or 4 to 20 mA DC) + Sensor power
supply
(12 VDC +/−10%, 80 mA) (See note 2.)
Linear voltage output (0 to 5, 1 to 5, or 0 to 10 VDC) + Sensor power
supply (12 VDC +/−10%, 80 mA) (See note 2.)
Sensor power supply (12 VDC +/−10%, 80 mA)
(12 VDC +/−10%, 80 mA) (See note 2.)
(12 VDC +/−10%, 80 mA) (See note 2.)
None
Relay contact (H/L: SPDT each)
Relay contact (HH/H/LL/L: SPST-NO each)
Transistor (NPN open collector: HH/H/PASS/L/LL)
Transistor (PNP open collector: HH/H/PASS/L/LL)
L/LL)
DeviceNet (See note 2.)
None
5 inputs (M3 terminal blocks), NPN open collector
8 inputs (10-pin MIL connector), NPN open collector
5 inputs (M3 terminal blocks), PNP open collector
8 inputs (10-pin MIL connector), PNP open collector
1 2 3 4
Model
5
CSM_K3HB-H_DS_E_4_1
1

K32-DICN Summary of contents

Page 1

... Note: 1. CPA can be combined with relay outputs only. 2. Only one of the following can be used by each Digital Indicator: RS-232C/RS-485 communications, a linear output, or DeviceNet communications. Accessories (Sold Separately) K32-DICN: Special Cable (for event inputs, with 8-pin connector) K32-BCD: Special BCD Output Cable Input ...

Page 2

... Note power supply models require a control power supply capacity of approximately 1 A per Unit when power is turned ON. Particular attention is required when using two or more DC power supply models. The OMRON S8VS-series DC Power Supply Unit is recommended. 2. PNP input types are also available. 3. For K3HB-series DeviceNet models, use only the DeviceNet Connector included with the product. The crimp terminals provided are for Thin Ca- bles. Load: 500 Ω ...

Page 3

Characteristics −19,999 to 99,999 Display range Accuracy Thermocouple input: (±0. ±1°C, whichever is larger) ± 1 digit max. (See note.) Platinum resistance thermometer input: (±0. ±0.8°C, whichever is larger) ± 1 digit max. Sampling period ...

Page 4

Input Ranges Platinum-resistance Thermometer/Thermocouple Input type Platinum- resistance thermometer Name Pt100 K Connected terminals Tem- 2300 pera- ture 1800 1300.0 range 1300 (°C) 850.0 900 800 700 500.0 600 400 150.00 200 100 0 −100 −20.0 ...

Page 5

Common Specifications ■ Event Input Ratings Input type S-TMR, HOLD, RESET, ZERO, BANK1, BANK2, Contact ON: 1 kΩ max., OFF: 100 kΩ min. No-contact ON residual voltage max. OFF leakage current: 0.1 mA max. Load current ...

Page 6

DeviceNet Communications Communications protocol Supported communi- Remote I/O communications cations I/O allocations Message communications Connection methods Baud rate Communications media Communications distance Communications power supply Allowable voltage fluctuation range Current consumption Maximum number of nodes Maximum number of slaves Error ...

Page 7

Power Supply Derating Curve for Sensor (Reference Value) With 12 V Max. current (mA) 140 1 120 100 −20 − Ambient temperature (°C) Note: 1. The above values ...

Page 8

... Refer to the following User's Manual for application precautions and other information required when using the Digital Indicator: K3HB-S/-X/-V/-H Digital Indicator User's Manual (Cat. No. N128) The manual can be downloaded from the following site in PDF format: OMRON Industrial Web http://www.fa.omron.co.jp Continuous Data Output REQ ...

Page 9

K3HB-H 9 ...

Page 10

... BCD (NPN Open Collector): BCD Applicable Connector (Sold separately) HDR-E50MAG1 (HONDA TSUSHIN KOGYO CO., LTD.) Special Cable (Sold separately) K32-BCD (OMRON) (HDR-E50MAG1 with 0.3-m cable) The BCD COMMON is shared. The pins indicated in the above diagram as blank (white) boxes have been removed. * Only one of the following can be used for each Digital Indicator: communications, BCD, or DeviceNet. ...

Page 11

... TIMING 1 3: HOLD 5: ZERO 7: BANK4 9 9: BANK1 • Applicable Connector (Sold separately) XG4M-1030 (OMRON) • Special Cable (Sold separately) K32-DICN (OMRON) (XG4M-1030 with 3 m cable) * Excluding the K3HB-H BCD Output Cable Model K32-BCD K3HB end 38 mm Cover: HDR-E50LPA5 (manufactured by Honda Tsushin Co., Ltd.) Connector: HDR-E50MAG1 (manufactured by Honda Tsushin Co ...

Page 12

Main Functions Measurement Timing Hold Normal • Continuously performs measurement and always outputs based on comparative results. Input H comparative set value Measurement value H output Peak Hold/Bottom Hold • Measures the maximum (or minimum) value in a specified ...

Page 13

Input Compensation/Display Zero-trimming Compensates for mild fluctuations in input signals due to factors such as sensor temperature drift, based on OK (PASS) data at mea- surement. (This function can be used with sampling hold, peak hold, or bottom hold.) ...

Page 14

Output Comparative Output Pattern The output pattern for comparative outputs can be selected. In addi- tion to high/low comparison with set values, output based on level changes is also possible. (Use the type of output pattern appropriate for the ...

Page 15

... Rubber Packing (Sold Separately) K32-P1 If the rubber packing is lost or damaged, it can be ordered using the following model number: K32-P1. (Depending on the operating environment, deterioration, contraction, or hardening of the rubber packing may occur and so, in order to ensure the level of waterproofing specified in NEMA4, periodic replacement is recommended ...

Page 16

... Please read and understand this catalog before purchasing the products. Please consult your OMRON representative if you have any questions or comments. WARRANTY OMRON's exclusive warranty is that the products are free from defects in materials and workmanship for a period of one year (or other period if specified) from date of sale by OMRON. OMRON MAKES NO WARRANTY OR REPRESENTATION, EXPRESS OR IMPLIED, REGARDING NON-INFRINGEMENT, MERCHANTABILITY, OR FITNESS FOR PARTICULAR PURPOSE OF THE PRODUCTS ...