GSM15-12 SL Power ( Ault / Condor ), GSM15-12 Datasheet

no-image

GSM15-12

Manufacturer Part Number
GSM15-12
Description
Power Supply, Switcher; 15 W; Single Output; 12 V; 1.25 A (Typ.); 90 to 264 VAC
Manufacturer
SL Power ( Ault / Condor )
Type
Medicalr
Datasheet

Specifications of GSM15-12

Brand/series
GSM15 Series
Configuration
Open-Frame
Current, Input
Maximum input current at 90 Vac, 60 Hz with full rated output load not to exceed 0.6 A.
Current, Inrush
The inrush at 240 Vac, averaged over the first ac half-cycle under cold start conditions will not exceed 37 A.
Current, Leakage
128 uA (Nom.)(Class I)
Current, Output
1.25 A (Typ.)
Dimensions
3.00 in. L x 2.10 in. W x 0.92 in. H
Mounting Type
Panel
Noise, Output
0.5% rms, 1% Pk-Pk, 20 MHz Bandwidth
Number Of Outputs
1
Operation
Switching
Output
12 VDC @ 1.25 A
Output Type
Single Output
Overvoltage Protection
Built in OVP on all models. Approximately 120-140% of output voltage.
Power Rating
15 W
Power, Output
15 W
Primary Type
AC-DC
Regulation, Total
2
Standards
UL Listed 2601-1, IEC60601-1, CSA Certified, EN60601-1 and CE Marked to LVD
Voltage, Input
90 to 264 VAC
Voltage, Output
12 VDC
• Industry’s smallest 15 W medically approved switcher
• Compact size (3.00” x 2.10” x 0.92”)
• Wide-range ac input: 90-264 Vac
• Less than 75 μA leakage current @ 120 Vac
• Approved to UL2601-1, EN60601-1
• EMI to FCC, CISPR 11 Class B
• Overvoltage protection standard
• RoHS compliant models available (G suffix)
SPECIFICATIONS
Ac Input
90-264 Vac, 47-63 Hz single phase. Class I or class II grounding.
Input Current
Maximum input current at 90 Vac, 60 Hz with full rated output load not to
exceed 0.6 A.
Input Protection
Internal ac fuse provided on all units. Designed to blow only if a catastrophic
failure occurs in the unit -- Fuse does not blow on unsustained overload or
short circuit.
Inrush Current
Inrush is limited by internal thermistors. The inrush at 240 Vac, averaged over
the first ac half-cycle under cold start conditions will not exceed 37 A.
Efficiency
69-85% depending on model.
Overload Protection
Fully protected against short circuit and output overload. Short circuit protec-
tion is cycling type power limit. Factory set to begin power limiting at 23 W.
Overvoltage Protection
Built in OVP on all models. Approximately 120-140% of output voltage.
Output Noise
0.5% rms, 1% Pk-Pk, 20 MHz Bandwidth, differential mode. Measured with
noise probe directly across output terminals of the power supply.
Transient Response
Main Output - 500 µs max. response time for return to within 0.5% of final
value for a 50% load step change, ∆i/ ∆t< 0.2 A/μs. Maximum voltage devia-
tion is 3.5%.
Hold-up Time
10 ms minimum from loss of ac input voltage at full load, nominal line (120
Vac).
GLOBAL PERFORMANCE SWITCHERS
FEATURES:
All specifications are typical at nominal input, full load at 25
marked to LVD
O
C unless otherwise stated
15 Watt Global Performance Medical Switcher
Temperature Coefficient
0.03% / °C typical.
EMI/EM Compliance
All models include built-in EMI filtering to meet the following EMC requirements of
IEC601-1-2.
Performance Requirement
Conducted Emissions
Surge Discharge
RF Field Susceptibility
Fast Transients/Bursts
Surge Susceptibility
Conducted RF Susceptibility
Voltage Sags & Surges
Medical Safety Approvals
All models are Certified to be in compliance with the applicable requirements of
UL2601-1, IEC60601-1, CSA-C22.2 No. 601-1, EN60601-1.
Leakage Current
The maximum leakage current for GSM15 series will be as follows;
132Vac/60Hz UL2601-1 test method
264Vac/50Hz IEC60601-1 test method
Class II
Class I
Class I
Class II
GND
GND
EMC Standard
EN55011, Class B; FCC Class B
EN61000-4-2, Level 3
EN61000-4-3, Level 3
EN61000-4-4, Level 3
EN61000-4-5, Level 3
EN61000-4-6
EN61000-4-11
GSM15 Medical
Connection
Connection
D C P o w e r S u p p l i e s
Normal
Normal
128 μA
75 μA
39 μA
66 μA
Typical Margin
2 dB Class II Gnd
6 dB Class I Gnd
2 kV
2 V
500 V
500 V
25%
5%
Single Fault
Single Fault
105 μA
180 μA
54 μA
94 μA

Related parts for GSM15-12

GSM15-12 Summary of contents

Page 1

... EN61000-4-11 Medical Safety Approvals All models are Certified compliance with the applicable requirements of UL2601-1, IEC60601-1, CSA-C22.2 No. 601-1, EN60601-1. Leakage Current The maximum leakage current for GSM15 series will be as follows; 132Vac/60Hz UL2601-1 test method GND Class I Class II 264Vac/50Hz IEC60601-1 test method ...

Page 2

... Medical Voltage Model Output Min. GSM15-5 5 GSM15- GSM15- GSM15- GSM15- Notes: A. Rating with unrestricted convection cooling. B. Peak Power for 60 sec. 10% duty cycle or continuous rating with 150 LFM of airflow. C. Output voltages preset at factory, not user adjustable. D. Add “G” suffix to model nuber for RoHS compliant model. ...