Description
Oscilloscope, 4-CHannel, 1 GHz
Manufacturer
Agilent Technologies Test Equipment
Type
Digitalr
Datasheet

Specifications of DSO7104B

Bandwidth
1000 MHz
Brand/series
DSO7000 Series
Configuration
Bench/Portable
Display Type
Color TFT LCD
Memory
8 Mpts
Number Of Channels
4
Sample Rate
4 GS/s
Agilent Technologies
Infi niiVision 7000B Series Oscilloscopes
Data Sheet
Engineered for the best signal visibility

DSO7104B Summary of contents

Page 1

... Agilent Technologies Infi niiVision 7000B Series Oscilloscopes Data Sheet Engineered for the best signal visibility ...

Page 2

... MHz 2 GSa/s MSO7032B MSO7034B DSO7052B DSO7054B 500 MHz 4 GSa/s MSO7052B MSO7054B DSO7104B 1 GHz 4 GSa/s MSO7104B Choose from fourteen Infi niiVision 7000B Series models. Agilent provides an easy 5-minute DSO-to-MSO upgrade kit for previously purchased 7000 Series DSOs. 2 Scope Digital Memory channels ...

Page 3

... What gives the Infi niiVision 7000B Series the best signal visibility? 1. Biggest display Oscilloscopes are visual tools and larger, high-resolution screens make the product better. Bigger displays have become increasingly important as general purpose scopes need more space to display digital waveforms and serial decode traces in addition to traditional analog waveforms ...

Page 4

... MegaZoom III sample rates, allowing you to quickly zoom in on areas of interest. Sample rate and memory depth go hand-in-hand. Deep memory in oscilloscopes sustains a high sample rate over longer time spans. Capture a mix of analog or digital signals. Compare multiple cycles of digital signals with slower analog signals – ...

Page 5

Other useful features Dedicated Front Panel Key Simply press the dedicated front panel Search Key and enter a variety of specifi c search criteria High resolution mode. Offers bits of vertical resolution. This is accomplished by serially ...

Page 6

Why does a fast update rate matter? While bandwidth, sample rate and memory depth are key criteria for deciding which scope to purchase, an equally important characteristic is update rate. What is update rate? Update rate is how many waveforms ...

Page 7

How update rate affects signal visibility Capturing random and infrequent events on an oscilloscope is all about statistical probabilities. The key to improving the probability of capturing a signal anomaly is to minimize dead-time and take more pictures of the ...

Page 8

... Agilent’s mask test option (Option LMT or N5455A) for Infi niiVision Series oscilloscopes pro- vides a fast and easy way to test your signals to specifi ed standards, and uncover unexpected signal anomalies, such as glitches. Mask testing on other oscilloscopes is based on software- intensive processing technology, which tends to be slow. ...

Page 9

Software applications Trigger on and decode CAN serial packets. Time-correlated display of the FlexRay physical layer signal with protocol decoding. On-screen serial decode of an SPI packet Time-correlated display of the MIL-STD 1553 physical layer signal with protocol decoding. Debug ...

Page 10

... For more information: Option SEC Secure Environment Mode Option for Agilent 7000B Series Oscilloscopes Data Sheet Evaluation kit (N2918A) The evaluation kit includes a variety of signals that demonstrate MegaZoom III technology with its fast deep memory, superior waveform update rate, high defi ...

Page 11

... Included with all MSO models is a logic probe with 2x8 flying leads (includes 20 IC clips and five ground leads) For more comprehensive information, refer to the Agilent 5000, 6000 and 7000 Series Oscilloscopes Probes and Accessories Data Sheet and Selection Guide (Agilent publication numbers 5968-8153EN/ENUS and 5989-6162EN). ...

Page 12

... By specifying the interaction of system components, LXI enables fast and effi cient test system creation and reconfi guration. The 7000B Series oscilloscopes follow specifi ed LAN protocols and adhere to LXI requirements such as a built-in Web control server, IVI-COM driver, and easy- to-use SCPI commands. The standard Agilent I/O Library Suite makes it easy to confi ...

Page 13

... Agilent Infi niiVision Portfolio Agilent’s Infi niiVision lineup includes 5000, 6000, and 7000B Series oscilloscopes. These share a number of advanced hardware and software technology blocks. Use the following selection guide to determine which scope best matches your specifi c needs. Largest display, ...

Page 14

... Agilent Infi niiVision 7000B Series oscilloscopes: Engineered for the best signal visibility 12.1” large display makes it easier to view analog, digital and serial signals. High-resolution color display with XGA resolution and 256 levels of intensity reveals subtle details that most scopes won’t show you. ...

Page 15

Dedicated front panel controls make it easy to access the most common scope controls, including vertical and horizontal scaling. Search and Navigate front panel controls make it easy to fi view specifi c signal activity. Quickly pan and ...

Page 16

Performance characteristics Acquisition: scope channels Sample rate Memory depth Standard Vertical resolution Peak detection Averaging High resolution mode Filter Acquisition: digital channels (7000B Series MSO or MSO-upgraded 7000B Series DSO) Sample rate Maximum input frequency Memory depth Standard Standard Vertical ...

Page 17

Performance characteristics Vertical system: scope channels Scope channels Bandwidth (–3 dB)* AC coupled Calculated rise time (=0.35/bandwidth) Single-shot bandwidth Range 1 Maximum input Offset range Dynamic range Input impedance Coupling BW limit Channel-to-channel isolation Standard probes al) Probe ID * ...

Page 18

Performance characteristics Vertical system: scope channels (continued) ESD tolerance Noise, RMS, input shorted DC vertical gain accuracy vertical offset accuracy 1 Single cursor accuracy 1 Dual cursor accuracy* * Denotes warranted specifications, all others are typical. Specifications are ...

Page 19

Performance characteristics Horizontal Range Resolution Time scale accuracy* Vernier Delay range Analog delta-t accuracy Logic delta-t accuracy Modes XY Reference positions Segmented memory re-arm time Trigger system Sources Modes Holdoff time Trigger jitter * Denotes warranted specification. Specifications are valid ...

Page 20

Performance characteristics Trigger system (continued) Selections Edge Pattern Pulse width TV Sequence CAN FlexRay LIN USB SPI RS-232/UART Duration MIL-STD 1553 20 (continued) Edge, pulse width, pattern, TV, duration, sequence, CAN, LIN, USB, I ...

Page 21

Performance characteristics Trigger system (continued) Nth edge burst Autoscale Scope channel triggering Range (internal) Sensitivity* Coupling Digital (D15 - D0) channel triggering (7000B Series MSO or MSO-upgraded 7000B Series DSO) Threshold range (user defined) Threshold accuracy Predefined thresholds External (EXT) ...

Page 22

Performance characteristics Display system Display Throughput of scope channels Resolution Controls Built-in help system Real-time clock Measurement features Automatic measurements Voltage (scope channels only) Time Counter Threshold definition Cursors Waveform math Measurement statistics Precision Mode 22 (continued) 12.1-inch (255 mm ...

Page 23

Performance characteristics FFT Points Source of FFT Window Noise floor Amplitude Frequency resolution Maximum frequency Storage Save/recall (non-volatile) Storage type and format I/O Standard ports Max transfer rate Supported printers via USB General characteristics Physical size (WxHxD) Weight Probe comp ...

Page 24

Performance characteristics General characteristics (continued) Trigger out 10 MHz ref in/out Kensington lock Power requirements Line voltage range Line frequency Power usage Environmental characteristics Ambient temperature Humidity Altitude Vibration Shock Pollution degree Typical operator noise Indoor use Other Measurement categories ...

Page 25

... MHz DSO7052B 500 MHz DSO7054B 500 MHz MSO7052B 500 MHz MSO7054B 500 MHz DSO7104B 1 GHz MSO7104B 1 GHz Accessories included: Model number Standard 3-year warranty Standard probe 100 MHz : 10074C (default), N2871A (optional – for added price) 350 MHz – 1GHz : 10073C (default), N2873A (optional – for added price) ...

Page 26

Ordering information (continued) Options Product Description DSO to MSO upgrade* N2741A for DSO701xB N2735A for DSO703xB N2736A for DSO705xB N2737A for DSO710xB SEC Secure Environment Mode - Provides compliance with National Industrial Security Program Operating Manual (NISPOM) Chapter 8 requirements ...

Page 27

... Agilent Technologies Oscilloscopes Multiple form factors from 20 MHz to >90 GHz | Industry leading specs | Powerful applications 27 ...

Page 28

... Get the latest information on the products and applications you select. Agilent Direct www.agilent.com/fi nd/agilentdirect Quickly choose and use your test equipment solutions with confi dence. Agilent Channel Partners www.agilent.com/find/channelpartners Get the best of both worlds: Agilent’s ...