N9010A-503 Agilent Technologies Test Equipment, N9010A-503 Datasheet - Page 13

no-image

N9010A-503

Manufacturer Part Number
N9010A-503
Description
EXA SIGNAL ANALYZER, 3 GHZ
Manufacturer
Agilent Technologies Test Equipment
Type
Networkr
Datasheet

Specifications of N9010A-503

Frequency
3 GHz
Lead Free Status / Rohs Status
RoHS Exempt Product
Power Suite Measurement Specifications
Channel power
Amplitude accuracy, W-CDMA or IS95
(20 to 30 °C, attenuation = 10 dB)
Occupied bandwidth
Frequency accuracy
Adjacent channel power
Accuracy, W-CDMA (ACLR) (at specifi c
mixer levels and ACLR ranges)
Dynamic range (typical)
Offset channel pairs measured
ACP speed (fast method). Data
measurement and transfer time
Multiple number of carriers measured
Power statistics CCDF
Histogram resolution
MS
BTS
Without noise correction
With noise correction
±0.94 dB (±0.30 dB 95th percentile)
± [span/1000] nominal
Adjacent
±0.22 dB
±1.07 dB
–68 dB
–73 dB
1 to 6
14 ms nominal (σ = 0.2 dB)
Up to 12
0.01 dB
Alternate
±0.34 dB
±1.00 dB
–74 dB
–76 dB
13