FLUKE-8846A NIST W/O DATA Fluke, FLUKE-8846A NIST W/O DATA Datasheet - Page 8

no-image

FLUKE-8846A NIST W/O DATA

Manufacturer Part Number
FLUKE-8846A NIST W/O DATA
Description
Digital Multimeters 6.5 BENCH DMM NIST CAL CERT
Manufacturer
Fluke

Specifications of FLUKE-8846A NIST W/O DATA

Lead Free Status / Rohs Status
Lead free / RoHS Compliant
Other names
1259800
400 mA
Accuracy (8845A)
Accuracy is given as ± (% measurement + % of range)
Additional Current Errors
AC Current
The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs from 1 % to 5 % of
range, add an additional error of 0.1 % of range.
Input Protection ............................................................Tool accessible 11 A/1000 V and 440 mA/1000 V fuses, limits of 400 mA continuous 550
Measurement Method..................................................ac-coupled true-rms, dc-coupled to the fuse and shunt (no blocking capacitor)
AC Filter Bandwidth
Crest Factor Error (applies to non-sinusoidal waveforms only)
Input Characteristics
100 µA
1 mA
10 mA
100 mA
400 mA
1 A
3 A
10 A
100 μA
1 mA
10 mA
100 mA
1 A
3 A
10 A
Range
[1]
[2]
[3]
[1]
[2]
[3]
[4]
Slow.............................................................................3 Hz to 10 kHz
Medium .......................................................................20 Hz to 10 kHz
Fast ..............................................................................200 Hz to 10 kHz
Maximum Crest Factor ...........................................5:1 at full scale
Additional Crest Factor Errors (<100 Hz)..........Crest factor 1-2, 0.05 % of full scale
Range
[2]
[1][2]
[3]
[2][3]
[2]
[3]
[1]
Part of 10 A range
Available at front panel connectors only
400 mA available in software version 2.0 or greater only. 400 mA continuously; 550 mA for 2 minutes on,
1 minute off.
[4]
8846A Only
Part of 10 A range
Available at front panel connectors only
400 mA available in software version 1.0.700.18 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1
minute off; maximum crest factor 3:1 at 400mA
[1]
Digits
[3]
100.0000 μA
1.000000 mA
10.00000 mA
100.0000 mA
400.000 mA
1.000000 A
3.00000 A
10.00000 A
Resolution
0.01 + 0.02
0.007 + 0.005
0.007 + 0.02
0.01 + 0.004
0.03 + 0.004
0.03 + 0.02
0.05 + 0.02
0.10 + 0.008
(23 ±1 °C)
24 Hour
4½ Digits
10 nA
100 nA
1 μA
10 µA
100 µA
100 µA
1 mA
1 mA
NPLC
0.02
100
0.2
10
1
0.04 + 0.025
0.030 + 0.005
0.03 + 0.02
0.03 + 0.005
0.04 + 0.005
0.04 + 0.02
0.08 + 0.02
0.12 + 0.008
(23 ±5 °C)
Resolution
90 Days
5½ Digits
1 nA
10 nA
100 nA
1 μA
10 µA
10 µA
100 µA
100 µA
mA for 2 minutes on, 1 minute off.
Crest factor 2-3, 0.2 % of full scale
Crest factor 3-4, 0.4 % of full scale
Crest factor 4-5, 0.5 % of full scale
mA, 100 mA, 400 mA, 3A and 10A
Additional NPLC Noise Error for 1
6½ Digits
100 pA
1 nA
10 nA
100 nA
1 μA
1 μA
10 μA
10 µA
0.011 % of range ±4 μA
0.04 % of range ±4 μA
0.05 + 0.025
0.05 + 0.005
0.05 + 0.02
0.05 + 0.005
0.05 + 0.005
0.05 + 0.02
0.10 + 0.02
0.15 + 0.008
0.001 % of range
(23 ±5 °C)
0 % of range
0 % of range
1 Year
Shunt Resistance
(Ohms)
0.01 Ω
0.01 Ω
0.01 Ω
100 Ω
100 Ω
1 Ω
1 Ω
1 Ω
Outside 18 to 28 °C
0.002 + 0.003
0.002 + 0.0005
0.002 + 0.002
0.002 + 0.0005
0.005 + 0.0005
0.005 + 0.001
0.005 + 0.002
0.005 + 0.0008
Coefficient/ °C
Temperature
Additional NPLC Noise Error
<0.015 V
<0.15 V
<0.025 V
<0.25 V
<0.50 V
<0.05 V
<0.05 V
<0.5 V
Burden Voltage
for 100 μA, 10 mA, 1A
0.11 % of range ±4 μA
0.28 % of range ±4 μA
0.01 % of range
0 % of range
0 % of range
8845A and 8846A
Fluke Corporation
7