12391 NIST CAL CERT Simpson, 12391 NIST CAL CERT Datasheet - Page 24

no-image

12391 NIST CAL CERT

Manufacturer Part Number
12391 NIST CAL CERT
Description
Analog Multimeters 260-8P VOM NIST CAL CERT W/O DATA
Manufacturer
Simpson
Datasheet

Specifications of 12391 NIST CAL CERT

Lead Free Status / Rohs Status
Lead free / RoHS Compliant
4.9.3
d.
e.
f.
This range is only intended for measurements in low voltage circuits (under 25V
DC) such as the primary power in vehicles or their accessories.
a.
b.
c.
NOTE: The function switch has no effect
d.
e.
The 10A range is not internally fused in the 260-8P.
When using the 10A range, never disconnect a test lead from a jack or from the
circuit while the circuit is energized. Doing so may damage the test jacks or leads
and the arcing may be hazardous to the operator. Turn off circuit power and wait
for the meter reading to drop to zero.
The polarity reversing feature of the function switch does not operate on the 0-
10A range. If polarity is found to be incorrect, turn off circuit power, wait for the
meter indication to reach zero and then interchange test lead connections to the
circuit.
Open the grounded side of the circuit in which the current is being measured.
Connect the VOM in series with the circuit. Connect the red test lead to the
positive side and the black test lead to the negative side.
Turn the power on. Read the current in milliamperes on the black DC scale.
For the 1 mA range, use the 0-10 figures and divide by 10. For the 10 mA
range, use the 0-10 figures directly. For the 100 mA range, use the 0-10
figures and multiply by 10. For the 500 mA range, use the 0-50 figures and
multiply by 10.
Turn the power off and disconnect the test leads.
Plug the black test lead into the -10A
jack and the red test lead into the +10A
jack.
Set the range switch at 10 AMPS (dual
position with 10 mA).
Open the grounded side of the circuit
in which the current is being mea-
sured. Connect the VOM in series with
the circuit, connecting the red test lead
to the positive side and the black test
lead to the negative side.
on polarity for the 10 AMPS range.
Turn the power on. Read the current
directly on the black DC scale. Use the
0-10 figures to read directly in am-
peres.
Turn the power off and disconnect the
test leads.
Direct Current Measurement 0-10A Range
!
!
24
Figure 4-12. Jacks and Switch
Positions for Measuring Direct
Current
+ D.C.
- D.C.
COMMON
-10 A.
-
+
500 MA.
50 A.
AMPS.
ALL TERMINALS 1 V MAX
260
100MA.
10MA.
1MA.
AMPS.
+
50
250 MV.
+10A
AMPS.