Specifications of FLUKE-196BM

Bandwidth
100MHz
Channels
2
Display Type
LCD - Monochrome
Features
DMM, RS232/USB Interface, Save, Recall, Record, Trend
Probe Type
10:1 Voltage Probe (1 Gray, 1 Red)
Sampling Rate (per Second)
1GS/s
Input Impedance
1M
Rise Time (typ)
3.5nS
Voltage - Input (max)
CAT II 1000V , CAT III 600V
Voltage - Supply
7.2V Battery, 17.8VDC Adapter
Includes
Battery Charger/Adapter, Case, Clips, Medical Acc Kit, Probe Set, Software with Medical Ext, Test Leads, USB/RS232 Cable
Lead Free Status / Rohs Status
Lead free / RoHS non-compliant
Power
Line power: Country-specific line voltage
adapter/battery charger included
Battery power: Rechargeable NiMH (installed)
Battery operating time: 4 hours
Battery charging time: 4 hours
Battery power saving functions: Auto power down
with adjustable power down time. On-screen battery
status indicator.
Safety
Compliance:
EN61010-1-2001, Pollution degree 2;
UL 3111-1, with approval;
CAN/CSA C22.2, No. 61010-1-04, with approval;
ANSI/ISA-82.02.01
Input voltage ratings
Maximum probe voltage: 1000 V CAT II/600 V CAT III
(Maximum voltage between 10:1 probe tip [VPS200]
and reference lead)
Floating voltage: 1000 V CAT II/600 V CAT III
(Maximum voltage between earth ground and any
terminal [signal input or shielding])
Independently isolated inputs:
1000 V CAT II/600 V CAT III
(Maximum voltage between any terminal of one input
or probe [VPS200] and any other terminal of another
input or probe [VPS200])
Maximum voltage on BNC input directly
(input A or B): 300 V CAT III
Maximum voltage on meter input:
1,000 V CAT II/600 V CAT III
Environmental
Operating temperature: 0 °C to +50 °C
Storage temperature: -20 °C to +60 °C
Oscilloscope mode
Vertical deflection
Number of inputs: 2
Input coupling: AC, DC with ground level indicator
Input sensitivity: 5 mV to 500 V/div (with the
included VPS40 (Fluke 125, 124) and STL120 shielded
test leads measure up to 600 Vrms, CAT III)
Input voltage: 600 V CAT III. See “general specifica-
tions” for more detailed information.
Vertical resolution: 8 bit
Accuracy: ± (1 % of reading + 0.05 x range/div)
Input impedance: 1 MW ± 1 % // 225 pF with
STL120 shielded test leads; 1 MW ± 1 % // 20 pF
± 3 pF with BB120; 5 MW ± 1 % // 15.5 pF with
VPS40, 10:1 Voltage probe
4 Fluke Corporation
Bandwidth and risetime
Bandwidth (risetime)
Instrument risetime (input directly)
with VPS40 probes
input A and B directly
with STL120 Shielded Test Leads
Technical specifications 120 Series
ScopeMeter 190C/190B Series, 120 Series and the NEW ScopeMeter 125
Fluke 125, 124 Fluke 123
12.5 MHz
40 MHz
40 MHz
8.75 ns
12.5 MHz
20 MHz
20 MHz
17.5 ns
Humidity:
10 °C to 30 °C: 95% RH non condensing
30 °C to 40 °C: 75% RH non condensing
40 °C to 50 °C: 45% RH non condensing
Maximum operating altitude: 3,000 m (10,000 feet)
Maximum storage altitude: 12 km (40,000 feet)
Electro-Magnetic-Compatibility (EMC): EN 61326-1
for emission and immunity
Optically-isolated PC/printer interface
PC communication: Transfer instrument settings,
screen images, waveform data and waveform
references, compatible with FlukeView
for Windows
To printer: Supports HP Laserjet
LQ, Seiko DPU-414 and Postscript printers via optional
PAC 91
Warranty
Three-years, parts and labor on mainframe instrument
One-year on accessories
Accessories
Horizontal
Max. sample rate (both channels simultaneously):
Fluke 125, 124: 2.5 GS/s for repetitive signals;
25 MS/s for single shot
Fluke 123: 1.25 GS/s for repetitive signals; 25 MS/s
for single shot
Number of digitizers: 2
Time base range: 10 ns/div to 1 min/div; 20 ns/div
to 1 min/div (Fluke 123)
Maximum record length: 512 Min-Max points
per input
Accuracy: ± (0.1 % of reading + 1 pixel)
Glitch detect: 40 ns
Display and acquisition
Display modes: Input A, input A and B, envelope,
smooth
Acquisition modes: Normal (including glitch capture),
single shot, roll
Trigger and delay
Source: Input A, input B, external via optional ITP120
Modes: Automatic Connect-and-View, Free Run, Edge,
Single Shot, Video, Video Line
Standard accessories
Rechargeable battery pack (installed)
Line voltage adapter/battery charger
Voltage probes and accessories
Multimeter test leads
User manual
®
via optional PM9080.
®
Fluke 199C, Fluke 196C, Fluke 199B,
Fluke 196B, Fluke 192B
BP190
BC190
10:1 voltage probe (VPS200, 1 red + 1 grey)
including hook clip, ground lead with mini alliga-
tor clip, ground lead with hook clip, 4 mm add-on
probe tip, ground lead to 4 mm banana plug
TL75 Hard Point test lead set (1red, 1 black)
10 language versions on CD-ROM, “Getting Started”
booklet included with instrument
, DeskJet, Epson FX/
®