C7860

Manufacturer Part Number
C7860
Description
Deuterium Lamp Related Products
Manufacturer
Hamamatsu Photonic Systems
Datasheet
DEUTERIUM LAMP POWER SUPPLY C7860/M7628 SERIES
Dimensional Outline (Unit: mm)
Precautions When Using Deuterium Lamps
1. High voltage is used to operate the lamp. Use extreme caution to prevent electric shocks.
2. The guarantee period will be one year after shipment. The area of guarantee is limited to replacement of the faulty lamp. Faults resulting
C7860
M7628
DEUTERIUM
These product conforms to SAFETY/ EN61010-1 of the European Union.
The M7628 is recognized upon UL 1950 (File No.191652).
from natural disasters and incorrect usage will also be excluded from guarantee.
DISABLE
POWER
INPUT
LAMP
DISABLE INPUT CIRCUIT
DISABLE FUNCTION
POSITION
(1) ANODE
(2) APERTURE
(3) HEATER
(4) GROUND (GND)
(5) +24 VDC ±10 %
(6) +24 RETURN (GND)
(7) +5 VDC (+)
(8) +5 RETURN (-)
7
8
FRONT VIEW
DISABLE = +5 V BETWEEN PINS 7 & 8 POLARITY AS INDICATED, LOAD = 2 mA
ENABLE = 0 V BETWEEN PINS 7 & 8 OR PINS 7 & 8 OPEN
1.8 k
OPTO COUPLER
113
#6-32 SCREW TERMINALS, 8 PLACES
SIDE VIEW
TOP VIEW
200
118
102
7.7
REAR VIEW
COMPONENT
MAXIMUM
36.2 MAX.
HEIGHT
28.5
1.5
4-M3
6.2
TLSOA0061EA
TLSOA0057EA