tmc60 Tumbler Technologies + TRUMPower, tmc60 Datasheet

no-image

tmc60

Manufacturer Part Number
tmc60
Description
Open Frame Ac/dc Power Supply Selection Guide; Medically Approved To Ul 60601-1, Tuv En60601-1
Manufacturer
Tumbler Technologies + TRUMPower
Datasheet
TRUMPower
DESCRIPTION
The TMC60 series of compact, open PCB constructed, AC-DC
switching power supplies are capable of delivering 37.5-64
watts of continuous output power at convection cooling. They
operate at 90-264 VAC input voltage without the need of
voltage selection, and are suited for medical, information
technology and industrial applications. Approval to both EN
60601-1 and EN 60950-1 safety standards improves design-in
time and reduces end equipment compliance costs.
FEATURES
INPUT SPECIFICATIONS
Input voltage:
Input frequency:
Input current:
Earth leakage current:
OUTPUT SPECIFICATIONS
Output voltage/current:
Maximum output power:
Ripple and noise:
Overvoltage protection:
Overcurrent protection:
Temperature coefficient:
Transient response:
ENVIRONMENTAL SPECIFICATIONS
Operating temperature:
Storage temperature:
Relative humidity:
Derating:
M e d i c a l a n d I T E a p p r o v a l s
C o m p a c t s i z e 2 ” x 4 ” x 1 . 1 8 ”
S i n g l e , d u a l a n d t r i p l e o u t p u t s
F u l l r a n g e i n p u t 9 0 - 2 6 4 VA C
L o w e a r t h l e a k a g e c u r r e n t
L e v e l B e m i s s i o n s
RoHS compliant
Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com
90-264 VAC
47-63 Hz
1.3 A (rms) for 100 VAC
0.7 A (rms) for 240 VAC
150 µA max. @ 264 VAC, 60 Hz
See rating chart
See rating chart
2% peak to peak on 5.0 V model, 1% peak
to peak on other models.
Provided on output #1 only;
set at 112-132% of its nominal output
voltage
All outputs protected to short circuit
conditions
All outputs ±0.04% /
Maximum excursion of 4% or better on all
models, recovering to 1% of final value
within 500 µs after a 25% step load
change
-10
-40
5% to 95% non-condensing
Derate from 100% at +50
50% at +70
o
o
C to +70
C to +85
o
C
o
o
C
C
3350 Scott Blvd., Bldg. 13, Santa Clara, California 95054, USA
Tumbler Technologies + TRUMPower
o
C maximum
37.5-64 WATT MEDICAL & ITE SWITCHING POWER SUPPLIES
o
C linearly to
GENERAL SPECIFICATIONS
Switching frequency:
Efficiency:
Hold-up time:
Line regulation:
Inrush current:
Withstand voltage:
MTBF:
EMC Performance
EN55011 / EN55022:
FCC:
VCCI:
EN61000-3-2:
EN61000-3-3:
EN61000-4-2:
EN61000-4-3:
EN61000-4-4:
EN61000-4-5:
EN61000-4-6:
EN61000-4-8:
EN61000-4-11:
SAFETY STANDARD APPROVALS
TMC60 SERIES
UL 60601-1
CSA C22.2 No. 601.1
TÜV EN60601-1
62±5 KHz
80-88% typical except TMC60-T31-3 and
TMC60-T31-5 at 75% typical
12 msec minimum at 110 VAC
±0.5% maximum at full load
30 A @ 115 VAC, or 60 A @ 230 VAC,
at 25
4000 VAC from input to output
1500 VAC from input to ground
500 VAC from output to ground
400,000 hours at full load at 25
calculated per MIL-HDBK-217F
Class B conducted, class B radiated
Class B conducted, class B radiated
Class B conducted, class B radiated
Harmonic distortion, class A and D
Line flicker
ESD, ±8 KV air and ±6 KV contact
Radiated immunity, 3 V/m
Fast transient/burst, ±2 KV
Surge, ±1 KV diff., ±2 KV com.
Conducted immunity, 3 Vrms
Magnetic field immunity, 3 A/m
Voltage dips,
30% reduction for 500 ms
60% reduction for 100 ms
>95% reduction for 10 ms
o
C cold start
o
C ambient,

Related parts for tmc60

tmc60 Summary of contents

Page 1

... TRUMPower DESCRIPTION The TMC60 series of compact, open PCB constructed, AC-DC switching power supplies are capable of delivering 37.5-64 watts of continuous output power at convection cooling. They operate at 90-264 VAC input voltage without the need of voltage selection, and are suited for medical, information technology and industrial applications. Approval to both EN 60601-1 and EN 60950-1 safety standards improves design-in time and reduces end equipment compliance costs ...

Page 2

... PIN MODEL TMC60-S05 TMC60-S12 TMC60-S15 TMC60-S24 TMC60-S48 TMC60-D23 TMC60-D25 TMC60-T31 TMC60-T32 TMC60-T39 TMC60-T31-3 TMC60-T31-5 Phone: 408-988-6616 ● Fax: 408-988-6622 ● email: sales@trumpower.com ● Website: www.TRUMPower.com TMC60 MEDICAL & ITE SERIES Output #2 Min.-Max.Current Tol. Vnom Imin at forced air (2) (N/A) 2% (N/A) (N/A) 2% (N/A) (N/A) ...