C10083CA HAMAMATSU [Hamamatsu Corporation], C10083CA Datasheet - Page 2

no-image

C10083CA

Manufacturer Part Number
C10083CA
Description
Mini-spectrometer TM series High sensitivity type (integrated with back-thinned type CCD image sensor)
Manufacturer
HAMAMATSU [Hamamatsu Corporation]
Datasheet
2
Mini-spectrometer TM series
TM series mini-spectrometers use a transmission holographic
grating made of quartz and precision optical components
arranged on a rugged optical base, making it possible to
deliver high throughput and highly accurate optical charac-
teristics.
Spectral resolution vs. wavelength
Optical component layout
Light spectrally separated by a grating is focused according to light wavelength on predetermined image sensor posi-
tions, and high-order light is also input onto positions at integer-multiples of wavelengths. In these mini-spectrometers an
optical filter is attached to the image sensor to cut off high-order light, but this also causes a drop in the image sensor
output at the following wavelengths.
C10082CA/10082CAH: Near 340 nm and 500 nm, C10083CA/C10083CAH: Near 500 nm and 700 nm
Types not using a high-order light cut-off filter are also available. Please specify by adding “-01” to the type number when
ordering. (Example: C10082CA-01)
TRANSMISSION GRATING
8.0
7.0
6.0
5.0
4.0
3.0
2.0
1.0
0.0
200
FOCUSING MIRROR
300
C10083CA (Slit width 70 µm, Optical NA 0.22)
C10082CA (Slit width 70 µm, Optical NA 0.22)
C10083CAH (Slit width 10 µm, Optical NA 0.11)
C10082CAH (Slit width 10 µm, Optical NA 0.11)
400
WAVELENGTH (nm)
500
600
700
ENTRANCE SLIT
800
C10082CA, C10082CAH, C10083CA, C10083CAH
COLLIMATING LENS
IMAGE SENSOR
900 1000
KACCB0159EA
KACCC0287EA
Light to be measured is guided into the entrance port of TG
series through an optical fiber and the spectrum measured
with the built-in image sensor is output through the USB port
to a PC for data acquisition. There are no moving parts inside
the unit so stable measurements are obtained at all times. An
optical fiber that guides light input from external sources al-
lows a flexible measurement setup.
HIGH POWER UV-VIS FIBER LIGHT SOURCE
Connection example (transmission light measurement)
(Deuterium lamp and Halogen lamp)
L10290
FIBER
AC ADAPTER
(Accessories)
QUARTZ CELL
(For holding liquid sample)
MINI-SPECTROMETER
PC
USB CABLE
KACCC0317EA

Related parts for C10083CA