National Instruments • Tel: (800) 433-3488 • Fax: (512) 683-9300 • info@ni.com • ni.com
Specifications – PCMCIA-4050
DC Voltage
Accuracy numbers are for 5
errors, temperature variation, linearity, and noise.
Noise Rejection
NMRR ........................................................ 80 dB (10 Hz reading rate, 50/60 Hz
DC ECMRR........................................................ 140 dB (with a 1 k imbalance in LO lead)
AC ECMR (50/60 Hz) ........................................ 150 dB (with a 1 k imbalance in LO lead)
Input Characteristics
Input bias current....................................... 1 nA maximum
Input resistance ......................................... >1 G
DC Current
*Requires 10 A shunt, CSM-10A.
Accuracy numbers are for 5
errors, temperature variation, linearity, and noise.
Input Characteristics
200 mA Shunt
10 A Shunt
AC Voltage
Accuracy numbers are for 5
errors, temperature variation, linearity, and noise.
Noise Rejection
AC CMRR (DC to 60 Hz)..................................... >80 dB (with a 1 k imbalance in LO lead)
Input Characteristics
Input resistance ......................................... 1 M
Bandwidth.................................................. 20 Hz - 25 kHz
Additional AC Errors
Frequency dependent errors
1
Range
20.0000 mA
200.000 mA
10.0000 A*
Range
250.000 V
25.0000 V
2.00000 V
200.000 mV 0.0029% + 27 µV 0.014% + 250 µV 0.017% + 250 µV
20.0000 mV 0.0029% + 27 µV 0.014% + 250 µV 0.017% + 250 µV
250.000 V
25.0000 V
2.00000 V
200.000 mV 0.3% + 0.22 mV
20.0000 mV 0.4% + 100 µV
Range
Input Frequency
20 Hz - 50 Hz
50 Hz - 100 Hz
100 Hz - 5 kHz
5 Hz - 10 kHz
10 Hz - 25 kHz
2
Input protection .................................... Fuse F1 500 mA/250 V fast-fusing
Shunt resistor ....................................... 1
Burden voltage ..................................... <400 mV at 200 mA
Input protection .................................... Fuse F1 12.5 A/250 V fast-fusing
Shunt resistor ....................................... 10 m
Burden voltage ..................................... <300 mV at 10 A
Digit Multimeters
0.0032% + 4.9 mV 0.021% + 49 mV 0.024% + 49 mV 0.0017%/°C + 4800 µV/°C
0.0032% + 4.9 mV 0.021% + 49 mV 0.024% + 49 mV 0.0017%/°C + 4800 µV/°C
0.6% + 500 mV
0.3% + 30 mV
0.4% + 3 mV
0.0029% + 37 µV 0.014% + 260 µV 0.017% + 260 µV
0.02% + 4 mA
(25 °C ± 1 °C)
0.1% + 27 µA
0.1% + 27 µA
(25 °C ± 1 °C)
(25 °C ± 1 °C)
(Accuracy = % of reading ± µV)
(Accuracy = % of reading ± µA)
(Accuracy = % of reading ± mV)
24 Hour
24 Hour
24 Hour
Additional Error (% of reading)
1
1
1
2
0.035% + 26 mA
2
2
0.14% + 250 µA
0.14% + 250 µA
(25 °C ± 10 °C)
0.62% + 680 mV
0.32% + 210 mV
0.42% + 21 mV
0.32% + 1.2 mV
0.42% + 170 µV
(25 °C ± 10 °C)
digits and include the effects of full and zero-scale
digits and include the effects of full and zero scale
digits and include the effects of full and zero-scale
(25 °C ± 10 °C)
90 Day
90 Day
90 Day
2.5%
1.0%
0.0%
1.0%
2.5%
powerline frequency ±1%)
1 M
0.15% + 250 µA
0.15% + 250 µA
0.035% + 26 mA
(25 °C ± 10 °C)
0.62% + 680 mV 0.007%/°C + 20 mV/°C
0.32% + 210 mV 0.007%/°C + 20 mV/°C
0.42% + 21 mV
0.32% + 1.2 mV
0.42% + 170 µV
(25 °C ± 10 °C) (% of reading/°C ± mV/°C)
(25 °C ± 10 °C)
1 Year
1 Year
1 Year
(250 V, 25 V)
all ranges
(2 V, 200 mV, 20 mV ranges);
Temperature Coefficient
(% of reading/°C ± µA/°C)
Temperature Coefficient
Temperature Coefficient
(% of reading/°C + µV/°C)
0.019%/°C + 2 mV/°C
0.007%/°C + 0.110 mV/°C
0.019%/°C + 12 µV/°C
0.0035%/°C + 25 µA/°C
0.0035%/°C + 25 µA/°C
0.007%/°C + 2.5 mA/°C
0.0009%/°C + 25 µV/°C
0.0009%/°C + 25 µV/°C
0.0009%/°C + 25 µV/°C
AC Current
*Requires 10 A shunt, CSM-10A.
Accuracy numbers are for 5
errors, temperature variation, linearity, and noise.
Input Characteristics
200 mA Shunt
10 A Shunt
Resistance
Accuracy numbers are for 5
errors, temperature variation, linearity and noise.
Measurement modes ................................ 2-wire
Diode Testing
Accuracy numbers are for 5
errors, temperature variation, linearity, and noise.
Test current................................................ 100 µA
General Specifications
Settling time .............................................. Affected by source impedance and
Warm-up time ............................................ 30 minutes for measurements
Altitude....................................................... Up to 2,000 m; at higher altitudes the
Working voltage ......................................... 250 V maximum between either input
Power Requirement
+5 VDC ...................................................... 45 mA in operational mode
Environment
Operating temperature .............................. 0 to 55 °C
Storage temperature.................................. -20 to 70 °C
Relative humidity ....................................... 10 to 90%, noncondensing
Certifications and Compliances
CE Mark Compliance
The product meets applicable EU directives for CE Mark compliance as follows:
Safety compliance IEC 1010-1 Certified, Designed for UL 3111, Low-Voltage
Directive EN 61010-1, Installation Category II, Pollution Degree 2, Double
Insulated, Indoor use.
Range
20.0000 mA 0.35% + 110 µA
200.000 mA 0.45% + 0.22 mA
10.0000 A*
Range
Extended ohm 0.1% + 6 k
Range
2 V
(> 2 M )
2.00000 M
200.000 k
20.0000 k
2.00000 k
200.000
Input protection .................................... Fuse F1 500 mA/250 V fast-fusing
Shunt resistor ....................................... 1
Burden voltage ..................................... <400 mV at 200 mA
Input protection .................................... Fuse F1 12.5 A/250 V fast-fusing
Shunt resistor ....................................... 10 m
Burden voltage ..................................... <300 mV at 10 A
0.3% + 22 mA
0.006% + 0.5
0.006% + 0.4
0.006% + 0.4
0.006% + 60 µV
0.012% + 55
0.012% + 37
(25 °C ± 1 °C)
(25 °C ± 1 °C)
(Accuracy = % of reading ± mA)
(Accuracy = % of reading ± )
(25 °C ± 1 °C)
24 Hour
24 Hour
24 Hour
(Accuracy = % of reading ± µV)
1
1
1
2
2
2
0.37% + 170 µA
0.47% + 1.2 mA
0.32% + 120 mA 0.32% + 120 mA 0.026%/°C + 11 mA/°C
0.024% + 400 µV 0.027% + 400 µV
digits and include the effects of full and zero scale
0.077% + 370
0.077% + 350
digits and include the effects of full and zero-scale
digits and include the effects of full and zero-scale
(25 °C ± 10 °C)
(25 °C ± 10 °C)
(25 °C ± 10 °C)
0.024% + 4
0.024% + 4
0.024% + 4
0.1% + 60 k
90 Day
90 Day
90 Day
input signal changes
accurate within specifications
installation category must be derated
terminal and earth ground
0.37% + 170 µA 0.019% /°C + 0.120 mA/°C
0.47% + 1.2 mA 0.007%/°C + 12 µA/°C
(25 °C ± 10 °C)
0.080% + 20
(25 °C ± 10 °C)
(25 °C ± 10 °C)
0.080% + 2
0.027% + 4
0.027% + 4
0.027% + 4
0.1% + 60 k
1 Year
1 Year
1 Year
Temperature Coefficient
(% of reading/°C + mA/°C)
Temperature Coefficient
Temperature Coefficient
(% of reading/°C ± µV/°C)
(% of reading/°C ± /°C)
0.0020%/°C + 0.40 /°C
0.0020%/°C + 0.40 /°C
0.0020%/°C + 0.40 /°C
0.0072%/°C + 6 k /°C
0.0072%/°C + 35 /°C
0.0072%/°C + 35 /°C
0.002%/°C + 40 µV/°C