NIPCI-6122 NI [National Instruments Corporation], NIPCI-6122 Datasheet - Page 5

no-image

NIPCI-6122

Manufacturer Part Number
NIPCI-6122
Description
S Series Multifunction DAQ-up to 16-Bit, up to 3 MS/s per Channel, up to 8 Analog Inputs
Manufacturer
NI [National Instruments Corporation]
Datasheet
Specifications
These specifications are typical at 25 °C unless otherwise noted.
Analog
Input Characteristics
Number of channels
ADC Resolution
Sampling rate
Input signal ranges (selectable by channel)....... ±10, ±5, ±2.5, ±1.25 V
Input coupling................................................... DC
Input FIFO size
Calibration
Calibration interval........................................... 2 years
1
2.5 MS/s with NI-DAQmx; 3 MS/s with additional download. Special conditions apply.
NI 6123 ...................................................... 8
NI 6122 ...................................................... 4
NI 6133 ...................................................... 8
NI 6132 ...................................................... 4
NI 6122, 6123 ............................................ 16 bits, 1 in 65,536
NI 6132, 6133 ............................................ 14 bits, 1 in 16,384
Maximum (NI 6133, 6132)......................... 3.0 MS/s per channel
Maximum (NI 6123, NI 6122).................... 500 kS/s per channel
Minimum.................................................... No minimum
NI 6123 ...................................................... 16 or 32 MS
NI 6122 ...................................................... 16 MS
NI 6133 ...................................................... 16 or 32 MS
NI 6132 ...................................................... 16 MS
up to 3 MS/s per Channel, up to 8 Analog Inputs
S Series Multifunction DAQ – up to 16-Bit,
1
Digital I/O
Number of channels......................................... 8 input/output
Compatibility .................................................... TTL/CMOS
Power-on state ................................................. Input (high-impedance)
Input buffer....................................................... 2,044 B
Output buffer .................................................... 2,044 B
Transfer rate (1 word = 8 bits) ......................... 10 Mwords/s
Timing I/O
Number of channels......................................... 2 up/down counter/timers, 1 frequency scaler
Resolution
Compatibility .................................................... TTL/CMOS
Base clocks available
Max source frequency...................................... 20 MHz
Triggers
Analog Trigger
Source............................................................... All analog input channels
Level ................................................................. ± Full-scale
Slope................................................................. Positive or negative (software selectable)
Resolution......................................................... 8 bits, 1 in 256
Hysteresis......................................................... Programmable
Digital Trigger
Compatibility .................................................... TTL
Response .......................................................... Rising or falling edge
For detailed specifications, please refer to the NI 612x and NI 613x
specifications document.
Counter/timers........................................... 24 bits
Frequency scaler........................................ 4 bits
Counter/timers........................................... 20 MHz, 100 kHz
Frequency scaler........................................ 10 MHz, 100 kHz
National Instruments • Tel: (800) 813 3693 • info@ni.com • ni.com
5