Specifications of MS4024F

Voltage - Output
24V
Number Of Outputs
1
Power (watts)
40W
Applications
Medical
Power Supply Type
Switching (Open Frame)
Voltage - Input
85 ~ 264VAC
Mounting Type
Chassis Mount
1st Output
24 VDC @ 1.67A
Size / Dimension
5" L x 3" W x 1.08" H (127mm x 76.2mm x 27.5mm)
Power (watts) - Rated
40W
Operating Temperature
0°C ~ 50°C
Efficiency
88%
Approvals
CB, CE, CSA, cTUVus, UL
Line Regulation
±5%
Load Regulation
±5%
Lead Free Status / RoHS Status
Lead free / RoHS Compliant
3rd Output
-
2nd Output
-
4th Output
-
Other names
EPS368
A
1-888-357-2280
saleselpac@iccus.com
www.iccus.com
+44.1383.432920
saleseurope@iccus.com
TECHNICAL SUPPORT
WORLDWIDE
EUROPE
Notes
**Ripple and noise measured with 20MHz bandwidth; 10µF tantalum capacitor in parallel with a 0.1µF ceramic capacitor.
Input Voltage
Input Frequency
Input Current
Inrush Current
Zero Load
Power Consumption
Earth Leakage Current
Input
*visit www.iccus.com for complete details
5-YEAR LIMITED WARRANTY*
85 – 264VAC
100 – 240VAC Nominal
47 – 63Hz
<1.0A rms
<37A @230VAC Cold Start
<0.5W
<50µA @ 132VAC @ 60Hz
<75µA @ 264VAC @ 60Hz
Medical Switch Mode Power Supply
Output Voltage
Total Regulation
Minimum Load
Start-Up Delay
Hold-Up Time
Ripple & Noise
Over Voltage Protection
Over Temperature Protection
Over Current Protection
Short Circuit Protection
Output
MS40 Series
High Efficiency
Wide Range AC Input
Fully Regulated DC Output
Lifetime Expectation >5 years
Hold-up Time >16ms at full load
Output Floating
Safety Approval - EN60601-1 Class I
RoHs Compliant
See Table
+/-5%
No minimum load required
<150ms
>16ms
<1% pk-pk **
130 – 135%
Active - Recoverable; plus
Passive - Non Recoverable
120 – 180%
Shutdown, auto-restart
(hiccup mode)
40 Watts

Related parts for MS4024F

MS4024F Summary of contents

Page 1

TECHNICAL SUPPORT WORLDWIDE 1-888-357-2280 saleselpac@iccus.com www.iccus.com EUROPE +44.1383.432920 saleseurope@iccus.com Input Input Voltage Input Frequency Input Current Inrush Current Zero Load Power Consumption Earth Leakage Current Notes *visit www.iccus.com for complete details **Ripple and noise measured with 20MHz bandwidth; 10µF tantalum ...

Page 2

... Model Number MS4009F MS4012F MS4015F MS4018F MS4024F Notes 1) Maximum peak load (48W) lasting 500ms with a maximum 10% duty cycle. 2) Includes initial setting, line regulation, load regulation, and thermal drift. 3) Typical at 115VAC General Avg Efficiency 87.3% @ 115VAC; Efficiency 86.9% @ 230VAC MTBF min. 200,000 hours demonstrated 3.0" ...