I200 Fluke Electronics, I200 Datasheet - Page 2

no-image

I200

Manufacturer Part Number
I200
Description
CURRENT CLAMP AC 200A
Manufacturer
Fluke Electronics
Type
AC Clampr
Datasheets

Specifications of I200

Style
Pincer
Ac Current Range
500mA ~ 200A
Bandwidth
10kHz
Cable Length
59.055" (1500.00mm)
Max Conductor Size
0.787" (20.00mm)
Termination
Banana Plug, Double, Sheathed
Ratings
CAT III 600V
Lead Free Status / RoHS Status
Lead free by exemption / RoHS compliant by exemption
Other names
664755
Using the Current Clamp
To use the Current Clamp, follow these instructions:
1.
2.
3.
4.
5.
6.
7.
Example with multimeters for i200:
Current Clamp sensitivity = 1 mA/A. Multimeter displays 168 mA.
Example with multimeters for i200s:
Current Clamp set to 10 mV/A. Multimeter displays 1.85V.
Actual
sensitivit
Actual
sensitivit
Connect the i200/i200s Current Clamp to the desired input on
the measuring instrument.
i200: See Figure 2.
i200s: See Figure 3. When you are using a multimeter, use the
Dual Banana to BNC Adapter (PM9081) to connect the Current
Clamp to the input.
i200s: On the Current Clamp, select the least sensitive range
(10 mV/A).
i200s: Select the appropriate clamp sensitivity on your
ScopeMeter test tool or oscilloscope.
Position the Current Clamp perpendicular to and centered
around the conductor.
Make sure that the arrow marked on the clamp jaw points
toward the load for phase measurements or away from the
load (toward the source) for neutral measurements. (See
Figure 4.)
Observe the current value and waveform on the instrument’s
display.
i200s: If desired, select a lower range on the Current Clamp
and set the corresponding sensitivity (mV/A setting) on the
ScopeMeter test tool or oscilloscope.
display
display
current
current
y
y
Current
Current
value
value
=
=
Clamp
Clamp
=
=
168
10
1
1.85V
mA/A
mV/A
mA
=
=
168A
1850
10
mV/A
mV
=
185A
Measurement at
Measurement at
If the sensitivity setting (mV/A) of the ScopeMeter test
tool or oscilloscope does not correspond with the
setting of the Current Clamp, the ScopeMeter test tool
or oscilloscope may display a much lower current
than the actual value. This may result in a false and
misleading reading and, as a consequence, incorrect
measures to be taken.
selector
conductor
conductor
Range
a neutral
a phase
Figure 3. Measurement Setup for i200s
Figure 2. Measurement Setup for i200
Figure 4. Orientation of the Current Clamp
Warning
V
COM
V
ST8591_190
Measurement Considerations
Observe the following guidelines for positioning the Current Clamp
Jaws:
Observe the following guidelines when making measurements:
Maintenance
Before each use, inspect the clamp. Look for cracks or missing
portions of the clamp housing and output cable insulating cover and
for loose or weakened components. Pay particular attention to the
insulation surrounding the clamp jaws. Do not use a damaged
clamp. If a clamp is damaged, tape it shut to prevent unintended
operation. A damaged clamp under warranty will be promptly
repaired or replaced (at Fluke's discretion) and returned at no
charge.
Cleaning and Storage
Periodically wipe the case with a damp cloth and detergent; do not
use abrasives or solvents. Open the jaws and wipe the magnetic
pole pieces with a lightly oiled cloth. Do not allow rust or corrosion
to form on the magnetic core ends.
If your Current Clamp does not work
If the Current Clamp does not perform properly, use the following
steps to help isolate the problem:
Center the conductor inside the clamp jaws.
Make sure the clamp is perpendicular to the conductor.
Make sure that the arrow marked on the jaw of the Current
Clamp points toward the correct direction.
Avoid taking measurements close to other current-carrying
conductors.
On the i200s Current Clamp, select the most appropriate range
for the current being measured to get the best accuracy.
Inspect the jaw mating surface for cleanliness. If any foreign
material is present, the jaws will not close properly and errors
will result.
Verify that the function selection and range on the Multimeter,
ScopeMeter test tool or oscilloscope are correct and adjusted
to the sensitivity of the Current Clamp.
LIMITED WARRANTY & LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and
workmanship for one year from the date of purchase. This warranty
does not cover fuses, disposable batteries or damage from
accident, neglect, misuse or abnormal conditions of operation or
handling. Resellers are not authorized to extend any other warranty
on Fluke’s behalf. To obtain service during the warranty period,
send your defective product to the nearest Fluke Authorized Service
Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER
WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT
LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY
CAUSE OR THEORY.
Since some states or countries do not allow the exclusion or
limitation of an implied warranty or of incidental or consequential
damages, this limitation of liability may not apply to you.
or call Fluke using any of the phone numbers listed below:
Fluke Corporation
P.O. Box 9090
Everett WA
98206-9090, USA
+1-888-993-5853 in U.S.A. and Canada
To locate an authorized service center,
+1-425-446-5500 from other countries
SERVICE CENTERS
visit us on the World Wide Web:
+31-40-267-5200 in Europe
http://www.fluke.com
Fluke Industrial B.V.
P.O. Box 90
7600 AB Almelo
The Netherlands